Determination of Total Phosphorus Using Two-Dimensional Ion Chromatography

Applications | 2010 | Thermo Fisher ScientificInstrumentation
Ion chromatography, 2D-LC
Industries
Environmental
Manufacturer
Thermo Fisher Scientific

Summary

Significance of the Topic


Total phosphorus is a key nutrient and pollutant indicator in environmental and industrial waters. Accurate quantification at low microgram-per-liter levels is mandated by regulatory agencies (for example, U.S. EPA methods 365.1–365.5) to support water-quality monitoring, wastewater management, and agricultural runoff assessment. Conventional colorimetric assays rely on toxic antimony reagents and suffer from matrix interferences, driving the need for a confirmatory, reagent-free technique with high sensitivity and robustness.

Objectives and Study Overview


This application note describes a simple two-dimensional ion chromatography (2D-IC) approach for determining total phosphorus in persulfate-digested water samples. The aim is to isolate orthophosphate from a large excess of sulfate and other anions, enhance signal-to-noise ratio, and achieve a detection limit near 10 µg/L using a single analytical column and heart-cut concentrator setup.

Methodology and Instrumentation


Samples undergo oxidative digestion with persulfate to convert all phosphorus species to orthophosphate. The analytical sequence comprises two chromatographic passes and a trap-and-reinject heart-cut:
  • First Pass: Injection (250 µL) onto an IonPac AS11-HC column under isocratic 20 mM KOH (1.0 mL/min) to elute bulk matrix ions while retaining phosphate.
  • Heart-Cut: Between 11 and 15 minutes, valve switching directs the phosphate-rich fraction through a suppressor (ASRS 300) to a TAC-ULP1 concentrator column, where anions are trapped in a narrow band.
  • Second Pass: The concentrator is returned inline with the separator; a KOH gradient (20 mM to 40 mM back to 20 mM) elutes the trapped anions to the AS11-HC column for final separation and suppressed conductivity detection.
Instrument configuration is based on a Dionex ICS-3000 RFIC-EG system—with SP pump, EG module, ASRS suppressor, DC detector, and 10-port valve—with an alternative ICS-2100 platform also applicable.

Main Findings and Discussion


On a full-scale chromatogram, the phosphate peak (~1 ppm of total anion content) is obscured on the trailing edge of a dominant sulfate peak. After trap-and-reinject, the phosphate peak becomes sharper (peak width halved) and approximately three times taller, boosting signal-to-noise ratio by a factor of three. A linear calibration from 10 to 30 µg/L orthophosphate demonstrates reliable quantitation at regulatory levels.

Benefits and Practical Applications


  • Enhanced sensitivity and detectability of phosphate in complex matrices without toxic reagents.
  • Improved peak shape and quantitation through two-dimensional separation on a single column.
  • Reduced instrument complexity and cost by leveraging an auxiliary valve and concentrator rather than a second analytical column or pump.
  • Reagent-free suppressed conductivity detection provides a greener, confirmatory alternative to colorimetry.

Future Trends and Potential Applications


Advances may include miniaturized concentrator formats, integration with high-resolution mass spectrometry for speciation, fully automated sample prep and valve control, and extension of the 2D-IC approach to other challenging analytes in food, soil extracts, and industrial process streams. Development of higher-capacity suppressors and faster gradient programs will further lower detection limits and cycle times.

Conclusion


The described two-dimensional IC method provides a robust, sensitive, and cost-effective strategy for total phosphorus determination in digested water samples. By combining heart-cut trapping with gradient elution on a single AS11-HC column, the approach overcomes matrix suppression, eliminates toxic reagents, and simplifies instrumentation while achieving regulatory detection requirements.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Overcoming Challenging Matrices in Ion Chromatography
Overcoming Challenging Matrices in Ion Chromatography
2014|Thermo Fisher Scientific|Presentations
Overcoming Challenging Matrices in Ion Chromatography Presented by: Kirk Chassaniol Thermo Fisher Scientific – NA IC Tech Support Gulf Coast Conference 2014 October 14, 2014 The world leader in serving science 1 Column Selection: Dionex IonPac Columns • Selecting the…
Key words
dionex, dionexonguard, onguardinguard, inguardcartridge, cartridgesample, sampleadjoining, adjoiningelimination, eliminationanions, anionsmatrix, matrixhrp, hrpexchange, exchangeatc, atcminutes, minutescartridges, cartridgesthermo
Thermo Scientific Dionex IonPac AS11 and AS11-HC Anion-Exchange Column
Thermo Scientific Dionex IonPac AS11 and AS11-HC Anion-Exchange Column
2012|Thermo Fisher Scientific|Brochures and specifications
chromatography Product Specifications Separate a wide variety of inorganic anions and organic acids using the Dionex IonPac™ AS11 and AS11-HC Anion-Exchange Columns with potassium hydroxide gradients. Thermo Scientific Dionex IonPac AS11 and AS11-HC Anion-Exchange Column Column: 16 21 Eluent Source:…
Key words
anions, anionsautosuppression, autosuppressioninorganic, inorganicacids, acidsorganic, organicdionex, dionexanion, anionhydroxide, hydroxideminutes, minutesrecycle, recyclemode, modeunk, unkexternal, externalexchange, exchangecapacity
Determination of trace anions in organic solvents using matrix elimination and preconcentration
APPLICATION UPDATE 163 Determination of trace anions in organic solvents using matrix elimination and preconcentration Authors Terri Christison and Jeff Rohrer Thermo Fisher Scientific Sunnyvale, CA Keywords Ion chromatography, IC, anion contamination, corrosion, acetate, chloride, formate, isopropanol, lactate, matrix elimination,…
Key words
dionex, dionexanion, anionkoh, kohsample, sampleperistaltic, peristalticpump, pumpegc, egcanions, anionsrinse, rinseautosampler, autosamplerconcentrator, concentratordeionized, deionizedwaste, wasteionpac, ionpacthermo
Environmental Water Applications Notebook
Environmental Water Applications Notebook
2012|Thermo Fisher Scientific|Guides
Environmental Water Applications Notebook Anions • Cations • Bromate • Haloacetics Acids • Disinfection Byproducts Table of Contents Introduction to Environmental Water Analysis.......................................................................................................................... 4 Analysis of Anions..................................................................................................................................................................... 7 Monitoring Inorganic Anions and Cations During Desalination............................................................................................... 8 Determination of Total Phosphorus…
Key words
anions, anionsbromate, bromatewater, waterbromide, bromidedrinking, drinkingpostcolumn, postcolumninorganic, inorganicconductivity, conductivitydetermination, determinationcations, cationsreagent, reagentsuppressed, suppressedeluent, eluenthydroxide, hydroxidedisinfection
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike