Analysis of Cationic Lipids Used as Transfection Agents for siRNA with Charged Aerosol Detection
Applications | 2011 | Thermo Fisher ScientificInstrumentation
Small interfering RNA (siRNA) therapies rely on efficient delivery of short RNA segments into cells to modulate gene expression in conditions such as cancer, viral infections, and age-related diseases. Cationic lipid carriers form positively charged complexes that facilitate membrane penetration, but impurities and degradation products can impact transfection efficiency and safety. A robust analytical method to quantify both active lipid components and trace impurities is critical for quality control in research and the development of siRNA therapeutics.
This study describes the development and validation of a high-performance liquid chromatography method coupled with charged aerosol detection (HPLC-CAD) for five widely used cationic lipids: DC-Chol, DOTAP, DDAB, DOTMA, and DOPE. Goals include determining limits of detection and quantitation, establishing calibration curves, assessing purity relative to manufacturer certificates, and demonstrating application to commercial lipid formulations and stability testing.
The HPLC system comprised a Thermo Scientific Dionex UltiMate 3000 RSLC fitted with a Fused-Core C8 column (2.7 µm, 150 × 4.6 mm) maintained at 45 °C. Mobile phase A was water/methanol/trifluoroacetic acid (600:400:1) and phase B was alcohol/THF/trifluoroacetic acid (750:250:1). A gradient from 45% B to 75% B over 20 min was applied at 0.5 mL/min. Injection volume was 2 µL. Detection used the ESA Corona® ultra charged aerosol detector with high-voltage filter, nebulizer temperature 15 °C, and sample temperature 15 °C. Standard solutions (3–2000 ng on column) were prepared in methanol/chloroform.
Calibration curves fitted with second-order polynomials yielded r2 > 0.9995 across 3–2000 ng. Limits of detection ranged from 2 to 5 ng on column, and quantitation limits from 7 to 15 ng. Precision exhibited <5% RSD for amounts ≥12.5 ng. Purity assays on 2000 ng injections revealed DC-Chol at 94.9 % mass purity, DOTMA at 100 %, DDAB at 98.8 %, and DOTAP/DOPE at 100 %, consistent with supplier specifications. Application to three commercial transfection formulations resolved major lipids and minor components. A forced-degradation study showed a new degradant peak after eight days at ambient temperature.
As regulatory guidelines for siRNA therapeutics evolve, methods like HPLC-CAD will play a key role in impurity profiling and stability testing. Future extensions may include high-throughput platforms, coupling with orthogonal detectors for structural confirmation, and adaptation to broader classes of lipid nanoparticles, peptides, and polymeric carriers.
The developed HPLC-CAD assay offers sensitive, precise, and reproducible measurement of cationic lipid carriers and their impurities. Its superior dynamic range and ease of use make it a valuable tool for quality control in siRNA delivery research and product development.
HPLC
IndustriesClinical Research, Lipidomics
ManufacturerThermo Fisher Scientific
Summary
Significance of the Topic
Small interfering RNA (siRNA) therapies rely on efficient delivery of short RNA segments into cells to modulate gene expression in conditions such as cancer, viral infections, and age-related diseases. Cationic lipid carriers form positively charged complexes that facilitate membrane penetration, but impurities and degradation products can impact transfection efficiency and safety. A robust analytical method to quantify both active lipid components and trace impurities is critical for quality control in research and the development of siRNA therapeutics.
Objectives and Study Overview
This study describes the development and validation of a high-performance liquid chromatography method coupled with charged aerosol detection (HPLC-CAD) for five widely used cationic lipids: DC-Chol, DOTAP, DDAB, DOTMA, and DOPE. Goals include determining limits of detection and quantitation, establishing calibration curves, assessing purity relative to manufacturer certificates, and demonstrating application to commercial lipid formulations and stability testing.
Methodology and Instrumentation
The HPLC system comprised a Thermo Scientific Dionex UltiMate 3000 RSLC fitted with a Fused-Core C8 column (2.7 µm, 150 × 4.6 mm) maintained at 45 °C. Mobile phase A was water/methanol/trifluoroacetic acid (600:400:1) and phase B was alcohol/THF/trifluoroacetic acid (750:250:1). A gradient from 45% B to 75% B over 20 min was applied at 0.5 mL/min. Injection volume was 2 µL. Detection used the ESA Corona® ultra charged aerosol detector with high-voltage filter, nebulizer temperature 15 °C, and sample temperature 15 °C. Standard solutions (3–2000 ng on column) were prepared in methanol/chloroform.
Main Results and Discussion
Calibration curves fitted with second-order polynomials yielded r2 > 0.9995 across 3–2000 ng. Limits of detection ranged from 2 to 5 ng on column, and quantitation limits from 7 to 15 ng. Precision exhibited <5% RSD for amounts ≥12.5 ng. Purity assays on 2000 ng injections revealed DC-Chol at 94.9 % mass purity, DOTMA at 100 %, DDAB at 98.8 %, and DOTAP/DOPE at 100 %, consistent with supplier specifications. Application to three commercial transfection formulations resolved major lipids and minor components. A forced-degradation study showed a new degradant peak after eight days at ambient temperature.
Benefits and Practical Applications
- The HPLC-CAD method provides uniform, mass-based response independent of chromophores or light-scattering properties.
- Sensitivity is at least 20-fold higher than ELSD, enabling impurity detection to 0.1 %.
- Wide dynamic range permits simultaneous quantitation of main lipid and trace impurities in a single run.
- Routine operation is simpler and less costly than mass spectrometry, supporting QC workflows for siRNA carrier development.
Future Trends and Possibilities of Use
As regulatory guidelines for siRNA therapeutics evolve, methods like HPLC-CAD will play a key role in impurity profiling and stability testing. Future extensions may include high-throughput platforms, coupling with orthogonal detectors for structural confirmation, and adaptation to broader classes of lipid nanoparticles, peptides, and polymeric carriers.
Conclusion
The developed HPLC-CAD assay offers sensitive, precise, and reproducible measurement of cationic lipid carriers and their impurities. Its superior dynamic range and ease of use make it a valuable tool for quality control in siRNA delivery research and product development.
Reference
- Downward J. RNA Interference. Br Med J. 2004;328:1245–1248.
- Pitkänen L, Ruponen M, Nieminen J, Urtti A. Pharm Res. 2003;4:576–583.
- Narang A, Thoma L, Miller D, Mahato RC. Bioconjug Chem. 2005;16:156–168.
- Dumitriu S, ed. Polymeric Biomaterials. 2nd ed. 2002;813.
- Zhong Z, Ji Q, Zhang JA. J Pharm Biomed Anal. 2010;51(4):947–951.
- ESA, Dionex Corporation. Corona ultra for UHPLC, Response Consistency.
- Dionex Corp. Improving the Quantitation of Unknown Trace Impurity Analysis of Active Pharmaceutical Ingredients.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Quantifying impurities in cationic lipids raw materials with the inverse gradient method using LC-CAD-MS
2024|Thermo Fisher Scientific|Applications
Application note | 003384 Biopharma Quantifying impurities in cationic lipids raw materials with the inverse gradient method using LC-CAD-MS Authors Application benefits Sissi White , Sven Hackbusch , 1 2 • Enable the quantification of impurities even in the absence…
Key words
inverse, inversedotap, dotapimpurity, impuritygradient, gradientimpurities, impuritiesamount, amountcad, cadlipid, lipidcalculated, calculatedpiece, pieceretention, retentiondelta, deltapeak, peakdifference, differencecationic
Thermo Scientific Charged Aerosol Detectors
2018|Thermo Fisher Scientific|Brochures and specifications
Discover what you’re missing Thermo Scientific Charged Aerosol Detectors Hidden peaks revealed The analyte detection challenge No single liquid chromatography (LC) detector delivers ideal results. Often, one analyte responds more strongly than another, or may not respond at all. What…
Key words
aerosol, aerosolcharged, chargedcad, cadtenofovir, tenofovirinverse, inversevanquish, vanquishgradient, gradientdetector, detectorresponse, responseeluter, eluteremtricitabin, emtricitabinveo, veoacid, acidthermo, thermocorona
Charged Aerosol Detection 101
2017|Thermo Fisher Scientific|Presentations
Charged Aerosol Detection 101 Dr. Alexander Schwahn European Sales Support Expert for Biopharma Industry Thermo Fisher Scientific, Reinach, Switzerland The world leader in serving science Outline • • • • • • 2 Introduction to charged aerosol detection (CAD) How…
Key words
cad, cadcorona, coronaaerosol, aerosolcharged, chargedveo, veocounterions, counterionsrebaudioside, rebaudiosiderabaudioside, rabaudiosideresponse, responseelsd, elsddetection, detectionstevioside, steviosideparticle, particleelectrometer, electrometerfocusjet
Sensitive cationic lipids impurities analysis with quantitation by charged aerosol detection and simultaneous mass confirmation by MS
2025|Thermo Fisher Scientific|Posters
Pharma and biopharma Sensitive cationic lipids impurities analysis with quantitation by charged aerosol detection and simultaneous mass confirmation by MS Sylvia Grosse1, Sissi White2, Susanne Fabel1, Katherine Lovejoy1, Min Du2 Thermo Fisher Scientific, Germering, Germany, Thermo Fisher Scientific, Lexington, MA,…
Key words
impurity, impuritydotap, dotapcad, cadinverse, inverselipid, lipidcationic, cationicisq, isqrsd, rsdvanquish, vanquishaerosol, aerosolgradient, gradientconfirmation, confirmationpeak, peakscientifictm, scientifictmrepeatability