Significance of the Topic


Forsythia suspensa is a widely used traditional Chinese medicinal herb valued for its active glycoside forsythin, which exhibits antioxidant and antibacterial properties. Accurate quantification of forsythin in herbal formulations ensures consistent efficacy, safety, and quality control, addressing growing regulatory and industry demands for standardized herbal products.

Objectives and Study Overview


The primary goal of this application note is to establish a reliable, isocratic reversed-phase liquid chromatography method for the rapid determination of forsythin content in Forsythia powder. The study demonstrates sample preparation, chromatographic separation, and UV detection conditions optimized for routine quality analysis.

Methodology and Sample Preparation


Sample preparation follows a simple ultrasonic extraction protocol:
  • Pass commercial Forsythia powder through a No. 5 sieve.
  • Weigh 2 g of sieved powder into a stoppered flask and add 25 mL methanol.
  • Perform ultrasonic extraction at 250 W and 40 kHz for 25 minutes.
  • Adjust the flask weight with methanol post-sonication, shake and filter.
  • Transfer 10 mL filtrate to a 25 mL volumetric flask, dilute with water, shake and filter into an LC vial.

Used Instrumentation


  • Column: Shim-pack Scepter C18-120 (250 × 4.6 mm, 5 μm).
  • Mobile phase: Acetonitrile / Water = 1:3 (isocratic).
  • Flow rate: 1.0 mL/min; Column temperature: 40 °C.
  • Detection: UV at 277 nm; Injection volume: 10 μL.
  • Vials: LabTotal LC/LCMS vials.

Main Results and Discussion


The method produced a sharp, symmetric peak for forsythin with a retention time around 8 minutes. Chromatograms of standard mix (0.2 mg/mL forsythin) and sample extracts show no interfering peaks in the critical region, indicating excellent selectivity. Reproducibility tests (replicate injections) demonstrated relative standard deviations below 2%, confirming method robustness for routine analysis.

Benefits and Practical Applications


  • Simple sample preparation and isocratic operation minimize solvent use and analysis time.
  • Reliable quantification supports quality assurance in herbal medicine manufacturing.
  • Adaptable to other plant matrices or active glycosides by adjusting extraction and detection parameters.

Future Trends and Potential Applications


Advances may include coupling with mass spectrometry for enhanced sensitivity and identification of related compounds. Green extraction techniques and miniaturized LC systems can further reduce solvent consumption and analysis footprint. Integration into high-throughput screening platforms will accelerate quality control workflows in nutraceutical and pharmaceutical industries.

Conclusion


The presented isocratic LC-UV method on a Shim-pack Scepter C18 column offers a cost-effective, robust solution for forsythin quantification in Forsythia samples. Its simplicity and reproducibility make it well suited for routine quality control in herbal product testing.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Analysis of Forsythin in Forsythia
Analysis of Forsythin in Forsythia
2023|Shimadzu|Applications
ERAS-1000-0009 LC Reversed-phase Shim-pack Series Shim-packTM GIS C18 Analysis of Forsythin in Forsythia 9 Keywords: Chinese herbal medicine, glucopyranoside, antioxidant, antibacterial Sample 1 of Standard Mix Chromatogram 1. Forsythin Standard Mix Forsythin (0.2 mg/mL in Methanol ) Pretreatment procedure a.…
Key words
forsythia, forsythiaforsythin, forsythinchromatogram, chromatogrambring, bringshim, shimvolume, volumepack, packphase, phasepretreatment, pretreatmentadd, addmix, mixreversed, reversedfinal, finalvial, vialseries
Analysis of Forsythiaside A in Forsythia
ERAS-1000-0010 LC Reversed-phase Shim-packTM Series Shim-pack ScepterTM C18-120 10 Analysis of Forsythiaside A in Forsythia Keywords: Chinese herbal medicine, phenylethanoid, anti-inflammatory, antioxidant Chromatogram of Standard Mix 1. Forsythiaside A Standard Mix Forsythiaside A (0.1 mg/mL in Methanol) Chromatogram of Forsythia…
Key words
forsythia, forsythiaforsythiaside, forsythiasidemix, mixchromatogram, chromatogrampacktm, packtmantioxidant, antioxidantshim, shimphase, phasepretreatment, pretreatmentstandard, standardreversed, reversedvial, vialacetonitrile, acetonitrileseries, seriesprocedure
Analysis of Hydroxysafflor yellow A in Safflower
ERAS-1000-0017 LC Reversed-phase Shim-packTM Series Shim-pack ScepterTM C18-120 Analysis of Hydroxysafflor yellow A in Safflower 17 Keywords: Chinese herbal medicine, Safflower, Hydroxysafflor yellow A Chromatogram of Standard Mix 1. Hydroxysafflor yellow A Standard Mix Hydroxysafflor yellow A (0.13 mg/mL 25%…
Key words
safflower, safflowerhydroxysafflor, hydroxysaffloryellow, yellowultrasonically, ultrasonicallymix, mixreflux, refluxstopper, stopperweight, weightchromatogram, chromatogrampacktm, packtmshake, shaketreat, treatbring, bringshim, shimpowder
Analysis of Notopterol and Isoimperatorin in Notopterygium
ERAS-1000-0015 LC Reversed-phase Shim-packTM Series Shim-pack ScepterTM C18-120 16 Analysis of Notopterol and Isoimperatorin in Notopterygium Keywords: Chinese herbal medicine, Notopterygium, Notopterol Chromatogram of Standard Mix 1. Notopterol 2. Isoimperatorin Standard Mix Notopterol (60 µg/mL in Methanol) Isoimperatorin (30 µg/mL…
Key words
notopterygium, notopterygiumisoimperatorin, isoimperatorinnotopterol, notopterolnopterygium, nopterygiumultrasonically, ultrasonicallymix, mixstopper, stopperchromatogram, chromatogrampacktm, packtmshake, shaketreat, treatbring, bringshim, shimpowder, powdercollect
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike