Characterization of deuterated Poly(styrene)
Applications | 2023 | Agilent TechnologiesInstrumentation
Deuterated poly(styrene) plays a key role as a contrast agent in neutron scattering experiments, enabling detailed investigations of polymer structure and dynamics. Accurate characterization of its molar mass and distribution is essential for reproducibility, data interpretation, and quality control in both academic and industrial research.
This application note describes a size-exclusion chromatography (SEC) method optimized for deuterated poly(styrene). The goal is to determine molar mass distribution and assess method suitability across a wide range of molecular weights, guiding users in selecting appropriate chromatographic conditions.
The separation was performed in tetrahydrofuran (THF) at 23 °C with a flow rate of 1.00 mL/min. Samples were injected at 20 µL. Calibration employed a high-molar-mass poly(styrene) kit. Data processing and molar mass calculations were conducted using PSS WinGPC software. Sample concentration recommendations vary by polydispersity index (PDI) and molecular weight:
Elugrams recorded on PSS SDV columns demonstrate clear baseline separation of deuterated poly(styrene) fractions. The resulting molar mass distributions align with expected values, confirming the accuracy of calibration. Slight shifts in retention times due to deuteration were observed but do not compromise resolution. The method covers oligomeric to ultrahigh-molecular-weight fractions in a single analysis.
Advances may include coupling SEC with multi-angle light scattering (MALS) or nuclear magnetic resonance (NMR) detectors for absolute molar mass determination, exploring alternative green solvents, and microfluidic SEC platforms to minimize solvent use and sample requirements.
The described SEC method employing THF, PSS SDV columns, and RI detection provides a robust, reproducible approach to characterize deuterated poly(styrene) over a wide molecular weight spectrum. It supports critical applications in neutron scattering and polymer research.
Consumables, LC columns, GPC/SEC
IndustriesEnergy & Chemicals
ManufacturerAgilent Technologies
Summary
Importance of the Topic
Deuterated poly(styrene) plays a key role as a contrast agent in neutron scattering experiments, enabling detailed investigations of polymer structure and dynamics. Accurate characterization of its molar mass and distribution is essential for reproducibility, data interpretation, and quality control in both academic and industrial research.
Objectives and Overview
This application note describes a size-exclusion chromatography (SEC) method optimized for deuterated poly(styrene). The goal is to determine molar mass distribution and assess method suitability across a wide range of molecular weights, guiding users in selecting appropriate chromatographic conditions.
Methodology
The separation was performed in tetrahydrofuran (THF) at 23 °C with a flow rate of 1.00 mL/min. Samples were injected at 20 µL. Calibration employed a high-molar-mass poly(styrene) kit. Data processing and molar mass calculations were conducted using PSS WinGPC software. Sample concentration recommendations vary by polydispersity index (PDI) and molecular weight:
- Narrow PDI (Mw/Mn ≤ 1.5): 2 g/L for 100–10 000 Da, 1–2 g/L for 10 000–1 000 000 Da, ≤ 0.5 g/L for > 1 000 000 Da.
- Broad PDI (> 1.5): 3–5 g/L across all molar masses.
Used Instrumentation
- Chromatography Columns: PSS SDV series, configured in sets or single linear columns for low to ultrahigh molecular weights.
- Detector: Shodex RI-71 refractive index detector.
- Calibration Standards: Poly(styrene) high-molar-mass kit from PSS (now Agilent).
- Software: PSS WinGPC for data acquisition and processing.
Main Results and Discussion
Elugrams recorded on PSS SDV columns demonstrate clear baseline separation of deuterated poly(styrene) fractions. The resulting molar mass distributions align with expected values, confirming the accuracy of calibration. Slight shifts in retention times due to deuteration were observed but do not compromise resolution. The method covers oligomeric to ultrahigh-molecular-weight fractions in a single analysis.
Benefits and Practical Applications
- Reliable molar mass and distribution data for deuterated poly(styrene), critical for neutron scattering sample preparation.
- Broad applicability across research, quality assurance, and batch consistency monitoring.
- Flexible column configurations tailored to specific molecular weight ranges.
Future Trends and Opportunities
Advances may include coupling SEC with multi-angle light scattering (MALS) or nuclear magnetic resonance (NMR) detectors for absolute molar mass determination, exploring alternative green solvents, and microfluidic SEC platforms to minimize solvent use and sample requirements.
Conclusion
The described SEC method employing THF, PSS SDV columns, and RI detection provides a robust, reproducible approach to characterize deuterated poly(styrene) over a wide molecular weight spectrum. It supports critical applications in neutron scattering and polymer research.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Characterization of Poly(butadiene-styrene) Copolymer
2023|Agilent Technologies|Applications
Errata Notice This document contains references to PSS or Polymer Standards Service. Please note that PSS is now Agilent. This document will be republished as an Agilent document in the future. # 10097 - Column Application Note Characterization of Poly(butadiene-styrene)…
Key words
pss, psssdv, sdvpoly, polybutadiene, butadienestyrene, styrenefax, faxmail, mailphone, phoneweb, webdocument, documentpolymer, polymerreadycal, readycalrubbers, rubberselugram, elugramrepublished
Characterization of brominated Poly(styrene)
2023|Agilent Technologies|Applications
Errata Notice This document contains references to PSS or Polymer Standards Service. Please note that PSS is now Agilent. This document will be republished as an Agilent document in the future. # 10024 - Column Application Note Characterization of brominated…
Key words
pss, psssdv, sdvfax, faxmail, mailphone, phonepoly, polystyrene, styreneweb, webdocument, documentpolymer, polymerelugram, elugramrepublished, republishederrata, erratarecommandations, recommandationspdi
Characterization of Poly(styrene)
2023|Agilent Technologies|Applications
Errata Notice This document contains references to PSS or Polymer Standards Service. Please note that PSS is now Agilent. This document will be republished as an Agilent document in the future. # 10025 - Column Application Note Characterization of Poly(styrene)…
Key words
pss, psssdv, sdvfax, faxmail, mailphone, phonepoly, polystyrene, styreneweb, webdocument, documentpolymer, polymerreadycal, readycalelugram, elugramrepublished, republishederrata, erratarecommandations
Molar mass determination of Poly(Butyl methacrylate-Styrene) Copolymer
2023|Agilent Technologies|Applications
Errata Notice This document contains references to PSS or Polymer Standards Service. Please note that PSS is now Agilent. This document will be republished as an Agilent document in the future. # 10003 - Column Application Note Molar mass determination…
Key words
pss, psssdv, sdvfax, faxmail, mailphone, phonemolar, molarpoly, polystyrene, styreneweb, webdocument, documentpolymer, polymerreadycal, readycalelugram, elugramrepublished, republishederrata