Peptide and protein analysis by capillary HPLC – Optimization of chromatographic and instrument parameters

Brochures and specifications | 2007 | Agilent TechnologiesInstrumentation
HPLC
Industries
Proteomics
Manufacturer
Agilent Technologies

Summary

Importance of the topic


Understanding and quantifying peptides and proteins is critical in biochemical research, drug development, and quality control. Capillary reversed-phase HPLC offers high sensitivity and resolution for low-volume biological samples, enabling detailed peptide mapping, identification, and confirmation of biomolecules. Optimizing chromatographic and instrumental parameters ensures reproducible retention times (<0.5% RSD) and peak areas (<2% RSD), enhancing confidence in routine analyses and complementing mass spectrometric detection.

Objectives and Overview of the Study


This application note demonstrates systematic optimization of chromatographic conditions (column chemistry, particle size, pore diameter, mobile phase composition, pH, temperature, and gradient) alongside instrumental settings (pump flow calibration, injector delay volume, detector cell volume, and temperature control) using the Agilent 1100 and 1200 Series capillary HPLC systems. The goal is to achieve robust, precise peptide and protein separations suitable for UV or MS detection.

Methodology and Instrumentation


The study employed an Agilent 1100 capillary LC system featuring:
  • Micro vacuum degasser (1 mL) and high-pressure capillary pump with electronic flow control (1–1300 µL/min)
  • Micro well-plate autosampler (0.01–8 µL) with automatic delay volume reduction
  • Thermostatted column compartment (10–80 °C)
  • Diode array detector with 500 nL flow cell and multi-wavelength acquisition

Columns tested included various Agilent Zorbax capillary phases (C18, SB-C18, XDB-C18, C8) at 80 Å and 300 Å pore size, plus Hypersil ODS (120 Å). Mobile phases comprised water/ACN mixes with 0.05%–0.1% TFA or 0.1% formic acid, and gradients ranged from steep (0.5–1.1% ACN per minute) to shallow (0.25% ACN/min). Temperature, pH, and solvent modifier concentrations were systematically varied to assess their impact.

Main Results and Discussion


Column chemistry: 300 Å reversed-phase columns (Extend C18) provided shorter retention and improved peak shapes for peptides (<2 kDa) and proteins (>10 kDa). Polar ODS phases resulted in peak tailing and extended retention. C8 phases offered faster elution but lower selectivity.

Mobile phase modifier: TFA (0.05%–0.1%) yielded optimal UV peak shape; lower TFA (0.01%) improved resolution of closely eluting peptides and reduced ion suppression for MS. Formic acid (0.1%) is recommended for LC–MS, though minor peak broadening and shifts in retention order may occur.

Gradient and temperature: Shallow gradients (0.25% ACN/min) enhanced resolution in peptide maps, at the cost of longer run times. Raising column temperature from 30 °C to 60 °C improved separation of early-eluting peptides and reduced analysis time by 10 minutes.

Instrument parameters: Activating automatic delay volume reduction halved system dead volume. Optimized pump flow calibration and solvent-specific compressibility settings yielded retention time RSD <0.3% and area RSD <2%. Detector multi-wavelength monitoring (206, 280, 400 nm) facilitates complementary peptide and protein detection.

Benefits and Practical Applications


  • High precision and reproducibility in peptide/protein retention times and peak areas.
  • Flexible method development by screening a small panel of capillary columns under identical conditions.
  • Compatibility with both UV and MS detectors, enabling orthogonal confirmation of analyte identity.
  • Reduced sample consumption and sensitivity for low-abundance peptides or protein digests.

Future Trends and Applications


  • Integration of capillary HPLC with high-resolution mass spectrometry for comprehensive proteomic profiling.
  • Development of novel stationary phases (e.g., core–shell or mixed-mode) to further enhance selectivity.
  • Automated method scouting platforms to accelerate the column and gradient optimization process.
  • Application in single-cell proteomics and biopharmaceutical QA/QC for detailed characterization of complex samples.

Conclusion


Systematic optimization of chromatographic parameters (column chemistry, mobile phase, gradient, temperature, pH) together with careful instrument setup (pump calibration, injector delay, detector cell volume) on capillary HPLC platforms enables robust, high-resolution peptide and protein separations. Achievable retention time RSDs below 0.5% and area RSDs below 2% support reliable routine analyses. These optimized methods are broadly applicable in proteomics, biopharmaceutical development, and QA/QC workflows.

References


  • Agilent Technologies. Chromatography and Spectroscopy Supplies Reference Guide, publication 5988-4785EN, 2002.
  • Rivier J, McClintock R. Reversed-Phase HPLC of Insulins from Different Species. J Chromatogr. 1983;268:112–119.
  • Geng X, Regnier FE. Retention Model for Proteins in Reversed-Phase Liquid Chromatography. J Chromatogr. 1984;296:15–30.
  • Agilent Technologies. Performance Characteristics of the 1100 Series Capillary HPLC System, technical note 5988-7511EN, 2002.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Promix - Enter a New Era in Biomolecule Analysis with Promix Columns
Promix TM Enter a New Era in Biomolecule Analysis with PromixTM Columns “Unsurpassed Selectivity and Peak Capacity for Peptides and Proteins” TM Applications: Promix Technology •Proteomics •Peptide/Protein Analysis •Peptide/Protein Purification Features & Benefits: • 2-D HPLC with a Single Column…
Key words
promix, promixpromixtm, promixtmarg, argval, valile, ilehumalog, humaloginsulin, insulinhis, hiscollected, collectedtyr, tyrcolumn, columnfraction, fractionlantus, lantusselectivity, selectivitypeptides
Performance characteristics of the Agilent 1100 Series capillary LC system using diode-array UV and MS for detection
Performance characteristics of the Agilent 1100 Series capillary LC system using diode-array UV and MS for detection Technical Note Abstract Capillary LC is the method of choice, when sample volumes are limited or/and higher sensitivity is essential. A LC capillary/MS…
Key words
delay, delayazithromycine, azithromycineinjection, injectionvolume, volumeflow, flowbypass, bypassvalve, valveabsorbance, absorbancestep, stepmixing, mixingcapillary, capillaryprecision, precisioncarry, carryheight, heighttimes
PepMap C18 Columns for Reversed-Phase Chromatography of Peptides
PepMap C18™ Columns for Reversed-Phase Chromatography of Peptides Operating Instructions 1 Table 3 Product Description PepMap C18™ columns are high-resolution columns designed for reversed-phase chromatography of peptides and similar biomolecules. The column packing consists of 5 µ spherical silica particles…
Key words
perseptive, perseptivetest, testcolumn, columnbiosystems, biosystemsreversed, reversedpreparing, preparingmmd, mmdtyr, tyrgly, glyphe, phegradient, gradientphase, phaseeluent, eluentpeptide, peptidegrip
PepMap C18 Columns for Reversed-Phase Chromatography of Peptides
PepMap C18™ Columns for Reversed-Phase Chromatography of Peptides Operating Instructions 1 Table 3 Product Description PepMap C18™ columns are high-resolution columns designed for reversed-phase chromatography of peptides and similar biomolecules. The column packing consists of 5 µ spherical silica particles…
Key words
biosystems, biosystemstest, testcolumn, columnreversed, reversedpreparing, preparingmmd, mmdtyr, tyrgly, glyphe, phegradient, gradientphase, phasepeptide, peptideeluent, eluentgrip, gripdrop
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike