High Throughput HPLC Analysis of Barbiturates
Applications | 2006 | Agilent TechnologiesInstrumentation
Barbiturates remain significant in forensic toxicology and clinical settings due to their diverse pharmacokinetics and potential for misuse. Rapid, robust analytical methods are essential for timely identification and quantification in biological samples.
This application note describes the transformation of a United States Pharmacopeia (USP) phenobarbital HPLC method into a high-throughput workflow. Four barbiturates (phenobarbital, butalbital, allobarbital, hexobarbital) and an internal standard were selected to demonstrate method scalability using Rapid Resolution (RR) and Rapid Resolution High Throughput (RRHT) columns.
Starting from a conventional 4.6×250 mm, 5 µm Eclipse XDB-C18 column (32 min runtime at 1 mL/min), shorter columns were tested:
Mobile phase: 60% acetate buffer (pH 4.5) / 40% methanol. Detector: UV at 254 nm (2 µL flow cell). Instrument: Agilent 1200 Series HPLC.
Replacement of larger particles with smaller, uniformly manufactured particles preserved selectivity (α factors constant across columns) while improving efficiency and peak sharpness. Van Deemter analysis showed that 1.8 µm particles maintain high efficiency at elevated flow rates, enabling a reduction in analysis time without sacrificing resolution. Doubling the flow rate on the RRHT column shortened runtime to ~3.5 min with acceptable backpressure in an appropriate UHPLC system.
Direct column substitution allows laboratories to upgrade existing HPLC protocols to high-throughput modes with minimal method redevelopment. Faster analysis enhances sample throughput, reduces solvent usage, and supports rapid decision-making in emergency toxicology and routine quality control.
Opportunities for further acceleration include utilizing elevated column temperatures, exploring alternative stationary phases, and adopting even shorter columns. Next-generation systems with higher pressure capabilities will broaden the range of amenable analytes and complex matrices.
RR and RRHT column technologies provide a straightforward approach for converting conventional HPLC assays into ultra-fast methods. This strategy maintains chromatographic integrity while delivering substantial gains in productivity and throughput for forensic and clinical laboratories.
HPLC
IndustriesForensics
ManufacturerAgilent Technologies
Summary
Importance of the Topic
Barbiturates remain significant in forensic toxicology and clinical settings due to their diverse pharmacokinetics and potential for misuse. Rapid, robust analytical methods are essential for timely identification and quantification in biological samples.
Objectives and Study Overview
This application note describes the transformation of a United States Pharmacopeia (USP) phenobarbital HPLC method into a high-throughput workflow. Four barbiturates (phenobarbital, butalbital, allobarbital, hexobarbital) and an internal standard were selected to demonstrate method scalability using Rapid Resolution (RR) and Rapid Resolution High Throughput (RRHT) columns.
Methodology and Instrumentation
Starting from a conventional 4.6×250 mm, 5 µm Eclipse XDB-C18 column (32 min runtime at 1 mL/min), shorter columns were tested:
- 4.6×100 mm, 3.5 µm RR column: reduced runtime to ~13 min at similar resolution.
- 4.6×50 mm, 1.8 µm RRHT column: further reduced runtime to ~7 min with maintained selectivity.
Mobile phase: 60% acetate buffer (pH 4.5) / 40% methanol. Detector: UV at 254 nm (2 µL flow cell). Instrument: Agilent 1200 Series HPLC.
Main Results and Discussion
Replacement of larger particles with smaller, uniformly manufactured particles preserved selectivity (α factors constant across columns) while improving efficiency and peak sharpness. Van Deemter analysis showed that 1.8 µm particles maintain high efficiency at elevated flow rates, enabling a reduction in analysis time without sacrificing resolution. Doubling the flow rate on the RRHT column shortened runtime to ~3.5 min with acceptable backpressure in an appropriate UHPLC system.
Benefits and Practical Applications
Direct column substitution allows laboratories to upgrade existing HPLC protocols to high-throughput modes with minimal method redevelopment. Faster analysis enhances sample throughput, reduces solvent usage, and supports rapid decision-making in emergency toxicology and routine quality control.
Future Trends and Applications
Opportunities for further acceleration include utilizing elevated column temperatures, exploring alternative stationary phases, and adopting even shorter columns. Next-generation systems with higher pressure capabilities will broaden the range of amenable analytes and complex matrices.
Conclusion
RR and RRHT column technologies provide a straightforward approach for converting conventional HPLC assays into ultra-fast methods. This strategy maintains chromatographic integrity while delivering substantial gains in productivity and throughput for forensic and clinical laboratories.
References
- Elmhurst College. Barbiturate Chemistry. Available at: http://www.elmhurst.edu/~chm/vchembook/6673barbit.html
- United States Pharmacopeia 27–National Formulary 22. U.S. Pharmacopeial Convention; 2004.
- Henderson JW Jr. Plug & Play Fast and Ultra-Fast Separations Using 3.5 µm Rapid Resolution and 1.8 µm RRHT Columns. Agilent Technologies; publication 5989-2908EN.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Improve Peak Shape and Productivity in HPLC Analysis of Pharmaceutical Compounds with Eclipse Plus C8 Columns
2006|Agilent Technologies|Applications
Improve Peak Shape and Productivity in HPLC Analysis of Pharmaceutical Compounds with Eclipse Plus C8 Columns Application Pharmaceuticals Authors John W. Henderson Jr., Nona Martone, and Cliff Woodward Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808-1610 USA Abstract Seven…
Key words
mau, maubore, borenarrow, narrowrapid, rapidsaver, saverparaben, parabenmin, minresolution, resolutionrrht, rrhtbarbitals, barbitalscolumns, columnsplus, pluscolumn, columneclipse, eclipseparabens
Agilent 1260 Infinity Binary LC Optimization Guide
2010|Agilent Technologies|Guides
Agilent 1260 Infinity Binary LC Optimization Guide Authors: Angelika Gratzfeld-Huesgen, Michael Frank, Christian Gotenfels Table of Contents Introduction 5 1. Theory of using smaller particles in liquid chromatography 6 1.1. Benefits of columns packed with sub-2-micron particles 12 Frictional heating…
Key words
application, applicationcolumn, columnresolution, resolutionrrht, rrhtrapid, rapiddelay, delaysheet, sheetdata, datavolume, volumecolumns, columnsagilent, agilentdatasheet, datasheetzorbax, zorbaxfast, fastthreaded
Maximizing Resolution or Minimizing Analysis Time? Comparing and Evaluating Column Choices for High Speed and High Resolution LC
2008|Agilent Technologies|Posters
Maximizing Resolution or Minimizing Analysis Time? Comparing and Evaluating Column Choices for High Speed and High Resolution LC Maureen Joseph, Brian Bidlingmeyer, William Long and William Barber Agilent Technologies, 2850 Centerville Rd., Wilmington, DE 19808, USA s Abstract Eclipse Plus…
Key words
mau, mauparticle, particleresolution, resolutionmin, minterm, termchoices, choicesdecreases, decreasesselectivity, selectivitysize, sizephase, phaserrht, rrhtdeemter, deemterbonded, bondedestradiol, estradiolincreasing
HPLC Separation Fundamentals - LC Columns and Consumables
2007|Agilent Technologies|Presentations
HPLC Separation Fundamentals LC Columns and Consumables Rita Steed Application Engineer Agilent Technologies, Inc December 11, 2007 Separation fundamentals Agilent Restricted December 11, 2007 Presentation Outline ¾ Major HPLC modes ¾ Key Equations • Resolution • van Deemter ¾ Common…
Key words
fundamentals, fundamentalsrestricted, restrictedseparation, separationagilent, agilentresolution, resolutionphase, phasepressure, pressureparticle, particleterms, termssize, sizemobile, mobilehplc, hplcrrht, rrhtchromatography, chromatographycolumn