Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolu- tion, and Solvent Use

Applications | 2007 | Agilent TechnologiesInstrumentation
HPLC
Industries
Homeland Security
Manufacturer
Agilent Technologies

Summary

Importance of the Topic


Accurate separation and quantification of explosive residues are critical in environmental monitoring, regulatory compliance, and forensic investigations. High-performance liquid chromatography (HPLC) methods that balance speed, resolution, and solvent usage support faster decision-making and cost-effective laboratory workflows.

Study Objectives and Overview


This study evaluates ZORBAX Extend-C18 column configurations under EPA method 8330 conditions for 14 nitroaromatic and nitramine explosives. The aim is to demonstrate how simple column substitutions and minor method adjustments can optimize separation performance, analysis time, and organic solvent consumption without compromising selectivity.

Methodology


An isocratic HPLC method was developed with the following parameters:
  • Mobile phase: 5 mM ammonium formate in water (A) and methanol (B)
  • Gradient: Isocratic at 75%A/25%B
  • Column temperature: Optimized around 40–41 °C to resolve critical analyte pairs
  • Column formats tested: 4.6 × 100 mm, 1.8 µm; 4.6 × 100 mm, 3.5 µm; 3.0 × 100 mm, 3.5 µm
  • Sample: EPA 8330 Mix A and Mix B combined and diluted in methanol/water

Instrumentation


  • Agilent 1200 Rapid Resolution LC system:
    • G1312B binary pump SL
    • G1376C automatic liquid sampler
    • G1316B thermally controlled column compartment
    • G1365C multiwavelength detector at 254 nm (2 µL cell)
  • ZORBAX Extend-C18 columns:
    • Rapid Resolution HT (4.6 × 100 mm, 1.8 µm)
    • Rapid Resolution (4.6 × 100 mm, 3.5 µm)
    • Solvent Saver Plus (3.0 × 100 mm, 3.5 µm)

Key Results and Discussion


Temperature adjustments of just 1 °C improved resolution of critical peak pairs (Rs7,6 from 1.19 to 1.50). The 1.8 µm, 4.6 × 100 mm column achieved baseline resolution for all 14 compounds (Rs > 1.5) in 26 min with 44.2 mL solvent. Substituting a 3.5 µm, 4.6 × 100 mm column reduced run time to 16 min (Rs ≈ 1.3 for critical pairs) and cut pressure by 44%. Switching to a 3.0 mm id column maintained a 16 min runtime while halving solvent consumption to 17.6 mL. These results illustrate how stationary phase particle size and column diameter directly influence separation efficiency, throughput, and solvent usage.

Benefits and Practical Applications


  • Flexible method customization via simple column swaps
  • High resolution minimizes peak coelution and obviates secondary confirmation
  • Short analysis times increase laboratory throughput
  • Reduced solvent usage lowers operating costs and environmental footprint

Future Trends and Applications


Emerging UHPLC systems with sub-2 µm particles and robust high-pressure capabilities will further accelerate separations. Green chromatography initiatives will drive smaller-diameter columns and alternative solvent systems. Integration with high-resolution mass spectrometry and automated method scouting will simplify method development for complex matrices.

Conclusion


ZORBAX Extend-C18 columns provide a versatile platform for EPA 8330 explosive residue analysis. By selecting particle size and column dimensions, analysts can tailor HPLC methods to achieve the desired balance of resolution, speed, and solvent economy, streamlining workflows in environmental and forensic laboratories.

Reference


  1. EPA Method 8330, Nitroaromatics and Nitramines by High Performance Liquid Chromatography (HPLC), Revision 0, September 1994.
  2. R. Kinghorn, C. Milner, J. Zweigenbaum, Analysis of Trace Residues of Explosive Materials by Time-of-Flight LC/MS, Agilent 5989-2449EN (2005).
  3. L. R. Snyder, J. J. Kirkland, Introduction to Modern Liquid Chromatography, 2nd ed., pp. 38–42, 1979.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
New Technology for Environmental Assays – Improve Resolution, Run More Samples at Lower Cost
New Technology for Environmental Assays – Improve Resolution, Run More Samples at Lower Cost Pittcon 2008 Group/Presentation Title Agilent Restricted Month ##, 200X Introduction – What are Some Issues with Typical Environmental HPLC Assays? Lots of analytes need to be…
Key words
title, titlepresentation, presentationrestricted, restrictedgroup, groupmau, mauagilent, agilentbenzo, benzomin, minwater, wateracetonitrile, acetonitrilepyrene, pyrenefluoranthene, fluoranthenemobile, mobilephase, phaseresolution
Acclaim® Explosives Columns - Manual
Acclaim® Explosives Columns - Manual
2011|Thermo Fisher Scientific|Manuals
for Acclaim® Explosives Columns Acclaim Explosives Product Manual Page 1 of 1 PRODUCT MANUAL FOR ACCLAIM® EXPLOSIVES COLUMNS Acclaim Explosives E1, 5 µm, Analytical Column, 4.6 x 250 mm (P/N 064305) Acclaim Explosives E2, 5 µm, Analytical Column, 4.6 x…
Key words
explosives, explosivesacclaim, acclaimmanual, manualproduct, productminutes, minutesmau, mauphase, phasecolumns, columnsmobile, mobiledimensions, dimensionstetryl, tetrylhmx, hmxrdx, rdxcolumn, columnrecommended
Simultaneous UHPLC/MS Analyses of Explosive Compounds
Simultaneous UHPLC/MS Analyses of Explosive Compounds
2010|Thermo Fisher Scientific|Applications
Application Note: 51879 Simultaneous UHPLC/MS Analyses of Explosive Compounds Guifeng Jiang, Thermo Fisher Scientific, San Jose, CA, USA Introduction Key Words • MSQ Plus Mass Detector • Explosives • Library Spectra • Sensitivity • UHPLC Explosive compounds, which are recognized…
Key words
tatp, tatpexplosive, explosivemsq, msqpetn, petntnt, tnttetryl, tetrylapci, apcidetection, detectionrdx, rdxspectra, spectramass, masscompounds, compoundsseventeen, seventeenplus, plusuhplc
Transferring and optimizing the determination of explosives to fast and high resolution chromatography
Transferring and optimizing the determination of explosives to fast and high resolution chromatography Application Note Environmental Authors Abstract Detlef Wilhelm This Application Note describes the method for detection of nanogram levels of AnaTox GmbH & Co. KG explosive constituents in…
Key words
explosives, explosivesnitroglycerine, nitroglycerinenitropenta, nitropentatnt, tnttetryl, tetrylhmx, hmxagilent, agilentrdx, rdxordnance, ordnanceverify, verifymethod, methoddone, doneprecision, precisiongradient, gradientrrht
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike