Assay of potassium bitartrate using ion chromatography

Applications | 2019 | Thermo Fisher ScientificInstrumentation
Ion chromatography
Industries
Pharma & Biopharma
Manufacturer
Thermo Fisher Scientific

Summary

Importance of the Topic


Potassium bitartrate is a key pharmaceutical ingredient used as a laxative and effervescent agent. Reliable and rapid quantitation of its content is crucial to ensure compliance with compendial standards and product quality. The conventional USP titrimetric assay is time consuming and relies on manual handling of reagents.

Objective and Study Overview


This study aimed to develop a selective ion chromatography assay to replace the titrimetric monograph method of the USP for quantitation of potassium bitartrate. The proposed method reduces analysis time, simplifies operation, and improves reproducibility.

Methodology and Instrumentation


The method employed an RFIC system with the following configuration:
  • Eluent generation via KOH cartridge and trap column
  • IonPac AG20 guard column and AS20 analytical column
  • Suppressed conductivity detection using a dynamically regenerated suppressor
  • Chromeleon software for compliant method control and reporting
Sample preparation involved dissolving the substance in deionized water to yield stock and working solutions at defined concentrations. Method validation followed USP general chapters for compendial methods.

Results and Discussion


The assay achieved baseline separation of tartrate from seven common anions within 5 minutes. Calibration over 0.2 to 25 mg/L yielded r2 of 0.9999. LOD and LOQ were 9.13 and 30.4 µg/L respectively. Accuracy studies showed recoveries between 97 and 102 and precision RSD values below 0.72. Robustness tests under ±10 variances in flow rate, eluent concentration, and column temperature confirmed method stability.

Benefits and Practical Applications


  • Rapid throughput with 5 min runtime per injection
  • Reduced reagent use due to onboard eluent generation
  • Enhanced reproducibility and traceability using RFIC automation
  • Direct replacement for USP titration assay in quality control

Future Trends and Opportunities


Automation and reagent free eluent generation will become more prevalent in pharmaceutical analysis. Expanding this approach to other drug substances and integrating hyphenated detectors may further improve specificity. Data connectivity and advanced software controls will support regulatory compliance and real time monitoring.

Conclusion


A validated ion chromatography assay offers a fast, accurate, and environmentally friendly alternative to the traditional titrimetric method for potassium bitartrate. Its ease of use and high performance make it well suited for routine quality control in pharmaceutical laboratories.

References


  • USP General Chapter 1225 Validation of Compendial Methods
  • USP General Chapter 621 Chromatography
  • ICH Q2A and Q2B Guidelines on analytical validation

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Determination of ammonia impurity in potassium bitartrate using ion chromatography
APPLICATION NOTE 73482 Determination of ammonia impurity in potassium bitartrate using ion chromatography Authors: Beibei Huang and Jeffrey Rohrer Thermo Fisher Scientific, Sunnyvale, CA Keywords: Dionex IonPac CS16 column, suppressed conductivity detection, pharmaceutical, USP Monograph, drug substance, RFIC system Goal…
Key words
bitartrate, bitartrateammonia, ammoniapotassium, potassiumusp, uspmonograph, monographimpurity, impuritytest, testcolorimetric, colorimetricconcentration, concentrationsodium, sodiumdifference, differenceeluent, eluentmodernize, modernizemethod, methodresolution
Assay of Nitrite and Determination of Nitrate Impurity in Sodium Nitrite Using a Reagent-Free Ion Chromatography System
Jingli Hu and Jeffrey Rohrer Thermo Fisher Scientific, Sunnyvale, CA, USA Appli cat i on N ote 1 1 4 8 Assay of Nitrite and Determination of Nitrate Impurity in Sodium Nitrite Using a Reagent-Free Ion Chromatography System Key Words…
Key words
nitrite, nitritesodium, sodiumnitrate, nitratediff, diffusp, uspave, avemethod, methodmonograph, monographconcentration, concentrationassay, assayrevision, revisionsolution, solutionrecovery, recoverycalculate, calculateeluent
Determination of oxalate in cromolyn sodium
Determination of oxalate in cromolyn sodium
2020|Thermo Fisher Scientific|Applications
APPLICATION NOTE 73345 Determination of oxalate in cromolyn sodium Authors: Sachin Patil and Jeffrey Rohrer, Thermo Fisher Scientific, Sunnyvale, CA, USA Keywords: RFIC, Dionex IonPac AS20 column, USP monograph, suppressed conductivity, pharmaceuticals, drug substance, ion chromatography Goal To develop an…
Key words
oxalate, oxalatecromolyn, cromolynsuppressed, suppressedassay, assaysodium, sodiumconductivity, conductivityeluent, eluentusp, uspmonograph, monographsulfate, sulfatedionex, dionexdescribes, describesarea, areapeak, peakexisting
Ion Chromatography Assay for Chloride and Sulfate in Adenosine
Ion Chromatography Assay for Chloride and Sulfate in Adenosine
2016|Thermo Fisher Scientific|Applications
Lipika Basumallick and Jeffrey Rohrer Thermo Fisher Scientific, Sunnyvale, CA, USA Key Words High-Pressure Ion Chromatography, Suppressed Conductivity, Fast Separation, Pharmaceuticals, Drug Products, Dionex IonPac AS18-4 µm Column Introduction Adenosine (Figure 1) is a nucleoside composed of adenine attached to…
Key words
adenosine, adenosinesulfate, sulfatechloride, chlorideasym, asymmix, mixusp, uspstandards, standardsrobustness, robustnessmonograph, monographdionex, dionexassay, assayassaying, assayingeluent, eluentturbidity, turbidityunaffected
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike