Characterization of Poly(n-butyl methacrylate)

Applications | 2023 | Agilent TechnologiesInstrumentation
Consumables, LC columns, GPC/SEC
Industries
Pharma & Biopharma
Manufacturer
Agilent Technologies

Summary

Importance of the topic


Poly(n-butyl methacrylate) (PBMA) is a soft, flexible polymer with a low glass transition temperature around 20 °C. It finds broad use in surface coatings, biomedical cements and controlled drug-release systems. Accurate determination of its molecular weight distribution is crucial for ensuring product performance, reproducibility and regulatory compliance in coatings, medical devices and pharmaceutical formulations.

Objectives and study overview


This application note describes the use of size-exclusion chromatography (SEC/GPC) to characterize PBMA samples over a wide molar mass range. The goals are:
  • To establish optimal chromatographic conditions for narrow and broad dispersity PBMA fractions.
  • To demonstrate calibration and data-processing workflows using a PBMA standard kit.
  • To provide practical recommendations for sample preparation, column selection and analysis parameters.

Methodology and instrumentation


All analyses were performed on a GPC1200 system equipped with a differential refractive index detector. Sample separation used PSS SDV columns and tetrahydrofuran as the mobile phase. Key parameters:
  • Mobile phase: tetrahydrofuran
  • Stationary phase: PSS SDV columns
  • Flow rate: 1.00 mL/min
  • Column temperature: 25 °C
  • Detection: refractive index (GPC1200)
  • Calibration: PBMA calibration kit
  • Data processing: PSS WinGPC software
  • Injection volume: 20 µL

Recommended sample concentrations vary with molar mass and dispersity:
  • Narrow dispersity (PDI ≤ 1.5): 2 g/L for M 100 Da–10 000 Da; 1–2 g/L for M 10 000 Da–1 000 000 Da; ≤ 0.5 g/L for M > 1 000 000 Da.
  • Broad dispersity (PDI > 1.5): 3–5 g/L across all molar masses.

Column sets suited to different molecular-weight ranges:
  • Low M: P/N 201-0001 (set of 3) or sda083003lis
  • Medium M: P/N 201-0002 (set of 2) or sda083005lim
  • High M: P/N 201-0003 (set of 3) or sda083005lxl
  • Ultrahigh M: P/N 202-0001 (set of 3)

Main results and discussion


Chromatograms overlayed narrow-PDI PBMA fractions with distinct elution volumes proportional to molar mass. The plot demonstrates baseline separation of standards from ~3 670 Da up to ~772 000 Da, confirming column resolution and calibration accuracy. Peak positions shift to higher elution volumes for lower molar masses, with an internal standard ensuring retention time consistency. Data show linear log–log correlation between molar mass and elution volume across the calibration range.

Benefits and practical applications of the method


SEC/GPC characterization of PBMA using this approach offers:
  • Reliable molar mass distribution profiles for QA/QC in polymer manufacturing.
  • Rapid method setup with defined sample concentrations and column choices.
  • Versatility for analyzing narrow and broad dispersity samples in coatings, biomaterials or drug formulations.

Future trends and potential applications


Advances likely include integration of multi-detector systems (light scattering, viscometry) for absolute molar mass determination, hyphenation with online fraction collectors for process control and adaptation to green solvents. Machine-learning-driven data analysis may further enhance throughput and insight into polymer architecture.

Conclusion


The presented GPC method provides a robust, reproducible workflow for comprehensive PBMA characterization. Defined calibration, sample preparation and column strategies ensure accurate molar mass distribution assessment, supporting diverse industrial and biomedical applications.

Reference


Agilent Technologies (formerly PSS Polymer Standards Service). Application Note 10300: Characterization of Poly(n-butyl methacrylate), August 1, 2023.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Molar mass determination of Poly(Butyl methacrylate-Styrene) Copolymer
Errata Notice This document contains references to PSS or Polymer Standards Service. Please note that PSS is now Agilent. This document will be republished as an Agilent document in the future. # 10003 - Column Application Note Molar mass determination…
Key words
pss, psssdv, sdvfax, faxmail, mailphone, phonemolar, molarpoly, polystyrene, styreneweb, webdocument, documentpolymer, polymerreadycal, readycalelugram, elugramrepublished, republishederrata
Characterization of Poly(t-butyl acrylate)
Characterization of Poly(t-butyl acrylate)
2023|Agilent Technologies|Applications
Errata Notice This document contains references to PSS or Polymer Standards Service. Please note that PSS is now Agilent. This document will be republished as an Agilent document in the future. # 10013 - Column Application Note Characterization of Poly(t-butyl…
Key words
pss, psssdv, sdvfax, faxmail, mailphone, phonepoly, polyweb, webdocument, documentpolymer, polymerreadycal, readycalelugram, elugramrepublished, republishederrata, erratarecommandations, recommandationspdi
Characterization of Poly(n-propyl methacrylate)
Characterization of Poly(n-propyl methacrylate)
2023|Agilent Technologies|Applications
Errata Notice This document contains references to PSS or Polymer Standards Service. Please note that PSS is now Agilent. This document will be republished as an Agilent document in the future. # 10055 - Column Application Note Characterization of Poly(n-propyl…
Key words
pss, psssdv, sdvfax, faxmail, mailphone, phonemethacrylate, methacrylatepoly, polyweb, webdocument, documentpolymer, polymerreadycal, readycalelugram, elugramrepublished, republishederrata, erratarecommandations
Characterization of Poly(ethyl methacrylate)
Characterization of Poly(ethyl methacrylate)
2023|Agilent Technologies|Applications
Errata Notice This document contains references to PSS or Polymer Standards Service. Please note that PSS is now Agilent. This document will be republished as an Agilent document in the future. # 10042 - Column Application Note Characterization of Poly(ethyl…
Key words
pss, psssdv, sdvfax, faxmail, mailphone, phonemethacrylate, methacrylatepoly, polyweb, webdocument, documentpolymer, polymerreadycal, readycalelugram, elugramrepublished, republishederrata, erratarecommandations
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike