Quick Start Guide Dionex Inuvion Core IC system

Manuals | 2023 | Thermo Fisher ScientificInstrumentation
Ion chromatography
Industries
Other
Manufacturer
Thermo Fisher Scientific

Summary

Significance of the Topic


Ion chromatography (IC) is a cornerstone analytical technique for quantifying ionic species in environmental, pharmaceutical, industrial and quality-control laboratories. The performance of an IC system relies on proper system preparation and conditioning to achieve stable baselines, low detection limits and reproducible results. The Thermo Scientific™ Dionex™ Inuvion™ Core IC system offers modularity and automation to streamline routine analysis, reducing setup time and improving laboratory workflow.

Objectives and Overview


This Quick Start Guide aims to provide step-by-step instructions for bringing the Dionex Inuvion Core IC system into service efficiently. Key goals include:
  • Establishing correct tubing connections and priming protocols for pump and suppressor.
  • Hydrating and conditioning the suppressor for displacement chemical regeneration (DCR) mode.
  • Installing and equilibrating guard and analytical columns prior to sample analysis.
  • Monitoring baseline conductivity to verify system readiness.

Methodology and Instrumentation


The startup procedure encompasses two main phases: fluidic conditioning and column/suppressor preparation. Steps include:
  1. Pump priming with ultrapure water: connect eluent line to DI water reservoir, open priming knob, activate prime mode in Chromeleon™ CDS, purge until bubble-free flow is observed, then close the knob.
  2. System flush: run DI water at 1 mL/min for 60 min, turn off pump, degas eluent offline, then repeat priming with degassed eluent and flush for 30 min.
  3. Suppressor hydration (DCR mode): if suppressor is not pre-hydrated, push 3 mL DI water slowly through the eluent inlet, then 5 mL of 200 mN H₂SO₄ (anions) or 100 mM TBAOH (cations) through the regeneration inlet. Allow 20 min for full resin and membrane hydration.
  4. Reagent delivery: set up an optional regeneration pump or pressurized bottle for continuous regenerant supply to the suppressor, routing regen inlet and outlet correctly.
  5. Column installation and conditioning: mount guard column upstream of the analytical column, verify flow direction, flush new column to waste at standard flow rate for 30 min to remove storage preservatives, then connect to suppressor.
  6. System equilibration: initiate method-specified eluent at the target flow rate and allow baseline conductivity to stabilize below 25 µS/cm for anion methods or below 2 µS/cm for cation methods before starting sample sequences in Chromeleon CDS.

Used Instrumentation


The Inuvion Core IC system comprises modular components that can be configured according to application needs:
  • Digital pressure regulator (optional)
  • Conductivity detection cell
  • Tubing chases to waste and eluent organizer
  • Regen pump (optional)
  • Seal wash pump (optional)
  • High-pressure pump
  • Leak sensor
  • Injection valve
  • Column holder (guard and analytical columns)
  • Suppressor module
  • Eluent monitor (optional)

Key Results and Discussion


Following the prescribed startup workflow yields the following benefits:
  • Bubble-free fluid delivery and stable flow rates reduce noise and baseline drift.
  • Proper suppressor hydration ensures efficient ion suppression and minimal background conductivity.
  • Pre-conditioning of columns accelerates equilibration and extends column lifetime.
  • Modular design and optional components support automated regeneration and continuous monitoring, enhancing uptime.

Benefits and Practical Applications


The optimized startup protocol for the Inuvion Core IC system minimizes downtime and standardizes preparation steps across laboratories. Applications range from trace anion and cation determination in environmental waters to ion analysis in pharmaceuticals, food safety and industrial process control.

Future Trends and Directions


Advancements in IC technology are focusing on further automation, miniaturization and integration with advanced data-analysis platforms. Emerging trends include:
  • Automated reagent handling and suppressor regeneration to support unattended operation.
  • On-board diagnostics and predictive maintenance using AI-driven monitoring.
  • Compact, high-throughput configurations for decentralized testing and field deployment.

Conclusion


Implementing a structured quick-start protocol for the Dionex Inuvion Core IC system ensures rapid attainment of stable baselines and reliable chromatographic performance. Adherence to priming, suppressor conditioning and column equilibration steps standardizes workflows and enhances data quality across diverse ionic analysis applications.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Quick Start Guide Dionex Inuvion IC system
Quick Start Guide Dionex Inuvion IC system
2023|Thermo Fisher Scientific|Manuals
IC systems Quick Start Guide Dionex Inuvion IC system Use this quick start guide to get your Thermo Scientific™ Dionex™ Inuvion™ IC system up and running quickly and easily. IC tubing connections, priming and system rinse 1. Confirm the pump…
Key words
inuvion, inuvioneluent, eluentchase, chasedionex, dionexregen, regenseparator, separatorpump, pumptubing, tubingstart, startsuppressor, suppressorport, portflow, flowoptional, optionalquick, quickguide
Quick Start Guide Dionex Inuvion IC system with RFIC
Quick Start Guide Dionex Inuvion IC system with RFIC
2023|Thermo Fisher Scientific|Manuals
IC systems Quick Start Guide Dionex Inuvion IC system with RFIC Use this quick start guide to get your Thermo Scientific™ Dionex™ Inuvion™ IC system with RFIC up and running quickly and easily. IC tubing connections, priming and system rinse…
Key words
inuvion, inuvioneluent, eluentrfic, rficchase, chasedionex, dionexsuppressor, suppressorpump, pumptubing, tubingoptional, optionalstart, startbottle, bottleport, portquick, quickguide, guidefirmly
Aquion Quick Start Guide
Aquion Quick Start Guide
2021|Thermo Fisher Scientific|Manuals
Aquion Quick Start Guide Use this quick start guide to help get your Thermo Scientific™ Dionex™ Aquion™ IC system up and running quickly and easily. IC tubing connections, priming and system rinse 1. Verify that the pump is turned off.…
Key words
eluent, eluentsuppressor, suppressoraquion, aquionout, outcolumn, columnpump, pumpoptional, optionalvalve, valvehead, headdionex, dionexwaste, wastexducer, xducerstart, startheater, heaterdampener
Combustion ion chromatography with a Dionex Integrion HPIC System
Combustion ion chromatography with a Dionex Integrion HPIC System
2017|Thermo Fisher Scientific|Technical notes
TECHNICAL NOTE Combustion ion chromatography with a Dionex Integrion HPIC System Authors Daniel Khor,1 Jeff Rohrer,1 Adelon Austin,2 Mark Manahan2 1 Thermo Fisher Scientific, Sunnyvale, CA, USA; 2COSA Xentaur, TX, USA Keywords AQF-2100H, pyrolysis Goal To provide instructions to set…
Key words
integrion, integriondionex, dionexhpic, hpiceluent, eluentcolumn, columnsystem, systemconnect, connectport, portvalve, valvecombustion, combustionviper, viperwaste, wastesuppressor, suppressorpeek, peekmitsubishi
Other projects
GCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike