Simple and Selective Bioanalytical Plasma Sample Extraction Using Oasis® MCX SPE for High Analyte Recovery

Applications | 2024 | WatersInstrumentation
Sample Preparation, Consumables
Industries
Clinical Research
Manufacturer
Waters

Summary

Importance of the Topic


Extracting analytes from biological fluids with high recovery and minimal matrix interference is critical in bioanalytical workflows. Tyrosine kinase inhibitors are widely used in leukemia treatment and require precise quantification in plasma for pharmacokinetic and therapeutic drug monitoring. Mixed-mode ion-exchange SPE streamlines sample preparation, improving accuracy and reproducibility.

Study Objectives and Overview


This study aimed to evaluate a generic protocol using Waters Oasis MCX SPE cartridges to extract multiple tyrosine kinase inhibitors from plasma samples. Key goals included:
  • Assessing recovery rates without protocol optimization
  • Measuring matrix effects across common TKIs
  • Demonstrating method simplicity and selectivity

Methodology and Instrumentation


Mixed-mode cation-exchange SPE was performed using Oasis MCX cartridges. The generic protocol involved:
  1. Cartridge conditioning with methanol and aqueous buffer
  2. Loading acidified plasma samples to favor cation exchange
  3. Washing to remove neutral and anionic interferences
  4. Elution of TKIs under basic conditions

Instrumentation


  • Waters Oasis MCX solid-phase extraction cartridges
  • Standard laboratory centrifuge and vacuum manifold setup

Key Results and Discussion


The SPE protocol yielded average recoveries exceeding 72.6% for all tested TKIs and matrix effects ranging from -3.8% to -11.0%, indicating minimal ion suppression. These results highlight the robustness and selectivity of Oasis MCX SPE for routine bioanalytical workflows, eliminating the need for extensive method development.

Benefits and Practical Applications


  • Universal extraction protocol for multiple drugs
  • High throughput sample preparation with reduced development time
  • Consistent recoveries and low matrix interference suitable for LC-MS/MS analysis
  • Applicability to clinical pharmacokinetic studies and therapeutic drug monitoring

Future Trends and Opportunities


  • Integration with automated SPE platforms for increased throughput
  • Combination with online SPE-LC-MS/MS for direct coupling and reduced hands-on time
  • Expansion to other drug classes and metabolites using mixed-mode SPE
  • Miniaturized SPE formats for low-volume samples and high-sensitivity assays

Conclusion


The Oasis MCX SPE method offers a simple, selective, and high-recovery approach for extracting tyrosine kinase inhibitors from plasma. Its minimal matrix effects and generic protocol make it an attractive solution for bioanalytical laboratories seeking efficient and reliable sample preparation.

References


  • Williams RE, Leatherwood HM. Top 200 Small Molecule Drugs by Retail Sales in 2023. (accessed 21 June 2024).

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Rapid Mixed-Mode SPE Method Development of Tyrosine Kinase Inhibitor Oncology Pharmaceuticals Oasis™ Method Development Plate
Application Note Rapid Mixed-Mode SPE Method Development of Tyrosine Kinase Inhibitor Oncology Pharmaceuticals Oasis™ Method Development Plate Mary Trudeau, Margot Lee Waters Corporation Abstract The following work demonstrates the fast mixed-mode SPE screening approach for bioanalytical sample preparation, achieving high…
Key words
oncology, oncologykinase, kinaseoasis, oasisinhibitor, inhibitordevelopment, developmenttyrosine, tyrosinepharmaceuticals, pharmaceuticalsspe, spemixed, mixedplate, platemethod, methodrapid, rapidmode, modebioanalytical, bioanalyticaltki
Fast and Simple Bioanalytical Plasma Sample Extraction Using Oasis® PRiME HLB SPE for High Analyte Recovery
Application Note Fast and Simple Bioanalytical Plasma Sample Extraction Using Oasis® PRiME HLB SPE for High Analyte Recovery Mary Trudeau, Margot Lee Waters Corporation This is an Application Brief and does not contain a detailed Experimental section. Abstract This application…
Key words
hlb, hlboasis, oasisprime, primeprivacy, privacyspe, speselling, sellingplasma, plasmarecovery, recoverypolicy, policyanalyte, analytebioanalytical, bioanalyticalnotices, noticescareers, careerscookie, cookiecookies
Oasis™ PRiME HLB Solid Phase Extraction for High Bioanalytical Plasma Analyte Recovery and Low Matrix Effects of Top-Selling Pharmaceuticals
Application Note Oasis™ PRiME HLB Solid Phase Extraction for High Bioanalytical Plasma Analyte Recovery and Low Matrix Effects of Top-Selling Pharmaceuticals Mary Trudeau, Margot Lee Waters Corporation Abstract The following work demonstrates a simple, generic, and broadly applicable sample preparation…
Key words
selling, sellingbioanalytical, bioanalyticalhlb, hlboasis, oasisprime, primepharmaceuticals, pharmaceuticalsplasma, plasmaextraction, extractiontop, topeffects, effectsmatrix, matrixsolid, solidanalyte, analyterecovery, recoveryphase
Rapid Mixed-Mode SPE Method Development Using the Oasis® μElution Sorbent Selection Plate
Application Note Rapid Mixed-Mode SPE Method Development Using the Oasis® µElution Sorbent Selection Plate Mary Trudeau, Margot Lee Waters Corporation This is an Application Brief and does not contain a detailed Experimental section. Abstract This application demonstrates the fast and…
Key words
µelution, µelutionoasis, oasismixed, mixedspe, spesorbent, sorbentplate, plateselection, selectiondevelopment, developmentrapid, rapidmode, modemethod, methodprivacy, privacyusing, usingpolicy, policynotices
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike