Highly Efficient LC-MS/MS Analysis of Organophosphorous Pesticides Utilizing ARC-18 Column Selectivity with Inert Column Technology
Posters | 2024 | Restek | RAFAInstrumentation
Organophosphorous pesticides are widely used in agriculture to boost crop yields but pose significant risks to human health and the environment. Sensitive and reliable analytical methods are essential to monitor trace levels of these compounds. However, many organophosphates exhibit nonspecific adsorption (NSA) or binding (NSB) to metal surfaces, resulting in poor peak shape and low sensitivity. The use of inert, coated columns such as Raptor ARC-18 can mitigate these issues and improve method performance.
The study aimed to evaluate the performance of an inert Raptor ARC-18 chromatographic column versus standard stainless steel hardware in LC-MS/MS analysis of a panel of 15 organophosphorous pesticides. Key goals included assessing sensitivity improvements, peak shape, and chromatographic resolution, especially for challenging isomers and isobars.
The method employed an Raptor (Inert) ARC-18 column (100 mm × 2.1 mm, 2.7 µm) maintained at 50 °C. A multiresidue pesticide standard at 1 ng/mL in water/0.1% formic acid was injected (5 µL) and analyzed using LC-MS/MS with ESI+ in MRM mode. Mobile phase A was water with 0.1% formic acid and 2 mM ammonium acetate; phase B was methanol with 0.1% formic acid and 2 mM ammonium acetate. The binary gradient ranged from 95% A to 100% B over 6 minutes, with a total run time of 9 minutes at a flow rate of 0.4 mL/min.
Comparison of the inert and stainless steel columns revealed that 14 out of 15 pesticides exhibited improved peak area and height with inert hardware, with area ratios ranging up to 2.07 and height ratios up to 2.00. On average, there was a 1.6-fold increase in peak area and 1.7-fold in peak height across six most problematic compounds (e.g., Methamidophos, Acephate, Omethoate). Carbaryl was the only analyte with slightly reduced sensitivity on the inert column (area ratio 0.91). Enhanced signal-to-noise and reduced peak tailing confirmed reduced NSA/NSB interactions. Chromatograms demonstrated clear separation of isomeric pairs such as Mevinphos and Dimethomorph.
Emerging trends include development of novel inert surface chemistries for broader analyte classes, integration with high-resolution MS for non-target screening, microflow LC to reduce solvent consumption, and advanced data processing for rapid multi-residue quantitation across diverse matrices.
The use of inert Raptor ARC-18 columns in LC-MS/MS analysis of organophosphorous pesticides significantly enhances sensitivity, peak quality, and resolution, while simplifying method development and maintenance. This technology offers robust and efficient solutions for routine pesticide monitoring in environmental and food safety laboratories.
LC/MS, LC/MS/MS, LC/QQQ, Consumables, LC columns
IndustriesFood & Agriculture, Environmental
ManufacturerRestek
Summary
Importance of the Topic
Organophosphorous pesticides are widely used in agriculture to boost crop yields but pose significant risks to human health and the environment. Sensitive and reliable analytical methods are essential to monitor trace levels of these compounds. However, many organophosphates exhibit nonspecific adsorption (NSA) or binding (NSB) to metal surfaces, resulting in poor peak shape and low sensitivity. The use of inert, coated columns such as Raptor ARC-18 can mitigate these issues and improve method performance.
Objectives and Study Overview
The study aimed to evaluate the performance of an inert Raptor ARC-18 chromatographic column versus standard stainless steel hardware in LC-MS/MS analysis of a panel of 15 organophosphorous pesticides. Key goals included assessing sensitivity improvements, peak shape, and chromatographic resolution, especially for challenging isomers and isobars.
Materials and Methodology
The method employed an Raptor (Inert) ARC-18 column (100 mm × 2.1 mm, 2.7 µm) maintained at 50 °C. A multiresidue pesticide standard at 1 ng/mL in water/0.1% formic acid was injected (5 µL) and analyzed using LC-MS/MS with ESI+ in MRM mode. Mobile phase A was water with 0.1% formic acid and 2 mM ammonium acetate; phase B was methanol with 0.1% formic acid and 2 mM ammonium acetate. The binary gradient ranged from 95% A to 100% B over 6 minutes, with a total run time of 9 minutes at a flow rate of 0.4 mL/min.
Results and Discussion
Comparison of the inert and stainless steel columns revealed that 14 out of 15 pesticides exhibited improved peak area and height with inert hardware, with area ratios ranging up to 2.07 and height ratios up to 2.00. On average, there was a 1.6-fold increase in peak area and 1.7-fold in peak height across six most problematic compounds (e.g., Methamidophos, Acephate, Omethoate). Carbaryl was the only analyte with slightly reduced sensitivity on the inert column (area ratio 0.91). Enhanced signal-to-noise and reduced peak tailing confirmed reduced NSA/NSB interactions. Chromatograms demonstrated clear separation of isomeric pairs such as Mevinphos and Dimethomorph.
Benefits and Practical Applications
- Improved sensitivity and lower detection limits for trace pesticide residues.
- Consistent peak shape and reduced analyte loss due to NSA/NSB.
- Elimination of time-consuming column passivation and conditioning steps.
- Enhanced throughput in routine monitoring workflows.
Future Trends and Opportunities
Emerging trends include development of novel inert surface chemistries for broader analyte classes, integration with high-resolution MS for non-target screening, microflow LC to reduce solvent consumption, and advanced data processing for rapid multi-residue quantitation across diverse matrices.
Conclusion
The use of inert Raptor ARC-18 columns in LC-MS/MS analysis of organophosphorous pesticides significantly enhances sensitivity, peak quality, and resolution, while simplifying method development and maintenance. This technology offers robust and efficient solutions for routine pesticide monitoring in environmental and food safety laboratories.
Used Instrumentation
- LC Column: Raptor (Inert) ARC-18, 100 mm × 2.1 mm, 2.7 µm, 50 °C
- Standard: Multiresidue pesticide standard #1 (1 ng/mL, water/0.1% formic acid)
- Injection Volume: 5 µL
- Detector: LC-MS/MS, ESI+, MRM mode
- Mobile Phase A: Water, 0.1% formic acid, 2 mM ammonium acetate
- Mobile Phase B: Methanol, 0.1% formic acid, 2 mM ammonium acetate
- Gradient: 95:5 A/B to 0:100 A/B over 6 min, total run 9 min, flow 0.4 mL/min
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Highly Efficient LC-MS/MS Analysis of Multiple Mycotoxins Utilizing Biphenyl Column Selectivity with Inert Column Technology
2024|Agilent Technologies|Posters
Highly Efficient LC-MS/MS Analysis of Multiple Mycotoxins Utilizing Biphenyl Column Selectivity with Inert Column Technology Jan Pschierer, Diego A. Lopez, Shun-Hsin Liang, Paul Connolly, Restek, Bellefonte, PA, USA Results and Discussion Introduction With an ever-growing list of mycotoxins to analyse,…
Key words
inert, inerthardware, hardwarestainless, stainlessbiphenyl, biphenylwitz, witzsteel, steelcolumn, columnmycotoxin, mycotoxinalternariol, alternariolnsa, nsansb, nsbpassivation, passivationimprovement, improvementpeak, peakinteractions
Determination of Pesticides in Tobacco Automated with FREESTYLE QuEChERS and LC-MS/MS
2019|Agilent Technologies|Applications
APPLICATION NOTE | AN0032 Determination of Pesticides in Tobacco Automated with FREESTYLE QuEChERS and LC-MS/MS Dr. Hans Rainer Wollseifen (MACHEREY-NAGEL) APPLICATION NOTE | www.LCTech-online.com As of: November 2019, Version: 1.1 APPLICATION NOTE | AN0032 Content Introduction ................................................................................................................................................. 3 Method Development…
Key words
freestyle, freestylepesticide, pesticidequechers, quecherstobacco, tobaccoaldicarb, aldicarbrecovery, recoveryterbutylazin, terbutylazinspinosyn, spinosynspinosad, spinosadmatrix, matrixmethyl, methylcyamemazine, cyamemazinechlorthalonil, chlorthalonilfludioxinil, fludioxinilweekend
Determination of Pesticides in Tea Automated with FREESTYLE QuEChERS and LC-MS/MS
2019|Agilent Technologies|Applications
APPLICATION NOTE | AN0031 Determination of Pesticides in Tea Automated with FREESTYLE QuEChERS and LC-MS/MS Dr. Hans Rainer Wollseifen (MACHEREY-NAGEL) APPLICATION NOTE | www.LCTech-online.com As of: November 2019, Version: 1.2 APPLICATION NOTE | AN0031 Content 1. Introduction ................................................................................................................................................. 3 2.…
Key words
freestyle, freestylepesticide, pesticidequechers, quechersaldicarb, aldicarbrecovery, recoveryterbutylazin, terbutylazinspinosyn, spinosynspinosad, spinosadmethyl, methyltea, teasummarise, summarisecyamemazine, cyamemazinechlorthalonil, chlorthalonilfludioxinil, fludioxinilmatrix
Simultaneous Determination of Alternaria Toxins, Ergot Alkaloid Epimers, and Other Major Mycotoxins in Various Food Matrices by LC-MS/MS 
2024|Restek|Posters
Simultaneous Determination of Alternaria Toxins, Ergot Alkaloid Epimers, and Other Major Mycotoxins in Various Food Matrices by LC-MS/MS Diego A. Lopez, Shun-Hsin Liang, and Justin Steimling; Restek Corporation, Bellefonte, PA, USA. Abstract & Introduction Sample and Matrix-Matched Standards Preparation Further…
Key words
alternariol, alternarioltenuazonic, tenuazonicnsb, nsbergometrinine, ergometrininemonomethylether, monomethyletherergotaminine, ergotaminineergometrine, ergometrinealtenuene, altenueneergocorninine, ergocorninineergosine, ergosineergocristinine, ergocristinineergosinine, ergosinineergocristine, ergocristineergocryptinine, ergocryptinineergocornine