Spot-on UV/Vis accuracy with Lunatic & Stunner
Applications | 2020 | Unchained LabsInstrumentation
Verifying UV/Vis spectrometer performance is essential for reliable quantification of proteins and nucleic acids in biologics research and quality control. Rapid, accurate instrument qualification saves time and resources by ensuring measurements meet regulatory and scientific standards before large‐scale studies begin.
This work assesses the accuracy, precision, and linearity of Unchained Labs’ Lunatic and Stunner microvolume UV/Vis instruments. Performance was evaluated using: certified reference materials from NIST (RM 8671 monoclonal antibody, SRM 2082 tryptophan and uracil standards), the Fundamentals L-tryptophan kit, and gravimetrically prepared protein dilution series. Goals included full‐range absorbance verification at 260 and 280 nm and demonstration of inter- and intra-instrument reproducibility.
The main instruments were:
NIST Evaluation
• Five Lunatic units; octuplicate measurements of 10 mg/mL NISTmAb at 280 nm.
• SRM 2082 tryptophan (1.4 mM) and uracil (1.0 mM) standards at 260/280 nm to calculate pathlength vs. nominal 10 mm.
Fundamentals Kit Assessment
• L-tryptophan standards (1–8 mg/mL) in octuplicate using Lunatic & Stunner v7.0 software.
• Linear regression of measured vs. target concentrations to evaluate accuracy, precision, linearity.
Protein Dilution Series
• Human serum albumin (0.323–365.5 mg/mL) and polyclonal IgG (0.1–200 mg/mL) prepared gravimetrically.
• Measurements on five Lunatic instruments with quick-load and standard protocols.
NIST Standards
• NISTmAb concentrations were within ±1.25% bias of the 10 mg/mL reference.
• SRM 2082 pathlengths deviated by less than 2% from nominal.
Fundamentals Kit
• Linear regression slope ≈0.9945, R² = 0.9999 across 1–8 mg/mL tryptophan.
• Mean recoveries between 99.1% and 100.6%, RSDs below 1%.
Protein Series
• Inter-instrument %CVs for HSA samples below 1.5% across 230–310 nm.
• IgG in PBS and acetate buffers showed linear concentration response from 0.1 to 200 mg/mL with R² ≈ 0.9999.
• Fast, low-volume (<2 µL) qualification simplifies routine instrument checks.
• High throughput (up to 96 samples) supports screening workflows.
• Certified standards and automated software reporting meet QA/QC and regulatory requirements.
• Expanded use of certified microvolume kits for additional wavelengths and analytes.
• Integration with Laboratory Information Management Systems for audit-ready records.
• Adoption in regulated biologics development and manufacturing environments.
Lunatic and Stunner UV/Vis spectrometers deliver robust accuracy, precision, and linearity across a wide concentration range. Validation with NIST reference materials and the Fundamentals kit confirms performance within tight bias and variability limits, streamlining instrument qualification in research and quality control labs.
Particle characterization, Particle size analysis, UV–VIS spectrophotometry
IndustriesProteomics
ManufacturerUnchained Labs
Summary
Importance of the Topic
Verifying UV/Vis spectrometer performance is essential for reliable quantification of proteins and nucleic acids in biologics research and quality control. Rapid, accurate instrument qualification saves time and resources by ensuring measurements meet regulatory and scientific standards before large‐scale studies begin.
Study Objectives and Overview
This work assesses the accuracy, precision, and linearity of Unchained Labs’ Lunatic and Stunner microvolume UV/Vis instruments. Performance was evaluated using: certified reference materials from NIST (RM 8671 monoclonal antibody, SRM 2082 tryptophan and uracil standards), the Fundamentals L-tryptophan kit, and gravimetrically prepared protein dilution series. Goals included full‐range absorbance verification at 260 and 280 nm and demonstration of inter- and intra-instrument reproducibility.
Instrumentation Used
The main instruments were:
- Lunatic microvolume UV/Vis spectrometer: fixed-pathlength microcuvettes, 2 µL sample, up to 96 samples per run.
- Stunner instrument: combines UV/Vis spectral analysis with dynamic light scattering in micro-volume plates.
- Fundamentals Check kit: five certified aqueous L-tryptophan concentrations for absorbance 230–310 nm qualification.
- NIST reference materials: RM 8671 human IgG1K for concentration accuracy; SRM 2082 tryptophan and uracil for pathlength validation.
Methodology
NIST Evaluation
• Five Lunatic units; octuplicate measurements of 10 mg/mL NISTmAb at 280 nm.
• SRM 2082 tryptophan (1.4 mM) and uracil (1.0 mM) standards at 260/280 nm to calculate pathlength vs. nominal 10 mm.
Fundamentals Kit Assessment
• L-tryptophan standards (1–8 mg/mL) in octuplicate using Lunatic & Stunner v7.0 software.
• Linear regression of measured vs. target concentrations to evaluate accuracy, precision, linearity.
Protein Dilution Series
• Human serum albumin (0.323–365.5 mg/mL) and polyclonal IgG (0.1–200 mg/mL) prepared gravimetrically.
• Measurements on five Lunatic instruments with quick-load and standard protocols.
Main Results and Discussion
NIST Standards
• NISTmAb concentrations were within ±1.25% bias of the 10 mg/mL reference.
• SRM 2082 pathlengths deviated by less than 2% from nominal.
Fundamentals Kit
• Linear regression slope ≈0.9945, R² = 0.9999 across 1–8 mg/mL tryptophan.
• Mean recoveries between 99.1% and 100.6%, RSDs below 1%.
Protein Series
• Inter-instrument %CVs for HSA samples below 1.5% across 230–310 nm.
• IgG in PBS and acetate buffers showed linear concentration response from 0.1 to 200 mg/mL with R² ≈ 0.9999.
Benefits and Practical Applications
• Fast, low-volume (<2 µL) qualification simplifies routine instrument checks.
• High throughput (up to 96 samples) supports screening workflows.
• Certified standards and automated software reporting meet QA/QC and regulatory requirements.
Future Trends and Applications
• Expanded use of certified microvolume kits for additional wavelengths and analytes.
• Integration with Laboratory Information Management Systems for audit-ready records.
• Adoption in regulated biologics development and manufacturing environments.
Conclusion
Lunatic and Stunner UV/Vis spectrometers deliver robust accuracy, precision, and linearity across a wide concentration range. Validation with NIST reference materials and the Fundamentals kit confirms performance within tight bias and variability limits, streamlining instrument qualification in research and quality control labs.
References
- National Institute of Standards and Technology. RM 8671 NISTmAb Humanized IgG1K Monoclonal Antibody (2020).
- National Institute of Standards and Technology. SRM 2082 Pathlength Absorbance Standards for Microliter Volume Spectrophotometers (2020).
- United States Pharmacopeia. Chapter <857> Ultraviolet-Visible Spectroscopy, USP 42.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Lunatic
2022|Unchained Labs|Brochures and specifications
Get your quant on Lunatic makes batch quantification of protein, DNA and RNA a no-brainer. All you need is 2 µL and 10 minutes to measure up to 96 samples. Run them straightup, even at high concentrations, without ever diluting.…
Key words
lunatic, lunaticigg, iggtryptophan, tryptophanplate, platewavelength, wavelengthprotect, protectvis, vislight, lightyour, youruse, usedna, dnavial, vialunchained, unchainedgets, getsonly
UV/Vis Pharmacopeia Performance Verification on Lunatic & Stunner
2020|Unchained Labs|Technical notes
TECHNICAL NOTE UV/Vis Pharmacopeia Performance Verification on Lunatic & Stunner Introduction A Compliance with international standards for analytical testing is a must-have for drug manufacturing and quality assurance/quality control (QA/QC). The United States Pharmacopeia (USP) chapter <857> and European Pharmacopeia…
Key words
lunatic, lunaticstunner, stunnervis, vispass, passverification, verificationwavelength, wavelengthabsorbance, absorbancepharmacopeia, pharmacopeiastray, straylight, lightacceptance, acceptanceperformance, performancespectrometers, spectrometerskit, kitabsorbances
A look under the hood of Lunatic
2020|Unchained Labs|Technical notes
Technical Note A look under the hood of Lunatic Introduction A Lunatic is the concentration liberator for next-gen protein and nucleic acid quantification. From 2 uL of sample, Lunatic delivers results with a full-spectrum spectrophotometer, a wide absorbance range from…
Key words
lunatic, lunaticabsorbance, absorbancehood, hoodvis, vislook, looklight, lightflashlamp, flashlampapps, appsmicrocuvette, microcuvettenucleic, nucleicprotein, proteinxenon, xenonextinction, extinctionturbidity, turbidityunder
DIY in Lunatic and Stunner with the Homebrew toolkit
2020|Unchained Labs|Applications
Application Note DIY in Lunatic and Stunner with the Homebrew toolkit Introduction A Quantifying biologics often requires many kinds of measurements and a lot of tedious downstream data processing. That can mean time-consuming, hands-on work with different instruments and software…
Key words
homebrew, homebrewstunner, stunnerlunatic, lunatictoolkit, toolkitdiy, diyyou, youviscosity, viscosityoutputs, outputsabsorbance, absorbancecorrection, correctionvis, visviscosities, viscositiescoefficient, coefficientdiffusion, diffusionyour