Chiral separation of Cyanofenphos

Applications |  | KNAUERInstrumentation
Consumables, LC columns, HPLC
Industries
Manufacturer
KNAUER

Summary

Significance of the topic


Chiral separation of cyanofenphos is essential for evaluating the enantiomeric purity of this pesticide, as different enantiomers can exhibit distinct biological activities and environmental behaviors. Reliable enantioselective analysis supports regulatory compliance and helps assess safety and efficacy in agricultural and environmental monitoring.

Objectives and study overview


This study aimed to establish an efficient and reproducible chiral HPLC method for separating the enantiomers of O-4-cyanophenyl-O-ethylphenylphosphonothioate (cyanofenphos) using a cellulose-based stationary phase. Key goals included optimizing chromatographic conditions to achieve baseline resolution and assessing method robustness.

Methodology


The developed protocol employed a Eurocel 03 column (250 × 4.6 mm, 5 µm) with a matching precolumn. Chromatographic parameters were:
  • Mobile phase: Hexane/2-propanol (80:20 v/v) with 0.1% trifluoroacetic acid
  • Isocratic elution at a flow rate of 1.0 mL/min
  • Column temperature maintained at 20 °C
  • Injection volume: 10 µL
  • UV detection at 234 nm

Main results and discussion


The method achieved clear enantiomeric separation with retention factors k′1 = 5.50 and k′2 = 8.55, yielding a selectivity factor (α) of 1.56. Chromatograms demonstrated sharp, symmetrical peaks and reproducible retention times. The use of cellulose-based Eurocel 03 provided strong chiral interactions, ensuring consistent enantioselectivity across runs.

Benefits and practical applications


The optimized HPLC method offers a robust tool for quality control, allowing accurate determination of enantiomeric excess in cyanofenphos formulations. Laboratories can implement this protocol for routine pesticide monitoring, environmental studies, and regulatory testing.

Future trends and potential applications


Advancements could include coupling the method with mass spectrometry for enhanced sensitivity, exploring greener mobile phase modifiers, and applying the approach to other chiral organophosphorus pesticides. Miniaturization and high-throughput adaptations may further streamline enantiomeric analyses.

Conclusion


The presented chiral HPLC procedure delivers reliable and reproducible separation of cyanofenphos enantiomers, supporting comprehensive enantiomeric purity assessments in analytical and regulatory contexts.

Used instrumentation


Eurocel 03 chiral column (250 × 4.6 mm, 5 µm) with precolumn, HPLC system, UV detector set at 234 nm.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Chiral separation of Praziquantel
Chiral separation of Praziquantel Method HPLC VCR0043J Chiral HPLC Column: Eurocel 03, 250 x 4.6 mm ID with precolumn Order No. 25VM400ECJ Phase: Eurocel 03 Conditions: Eluent: Gradient: Flow rate: Temperature: Volume: Detection: UV at 210 nm Substances: Praziquantel: (2S)-2-(cyclohexylcarbonyl)-2,3,4,6,7,11b-hexahydro-1H-pyrazino…
Key words
chiral, chiralpraziquantel, praziquantelprecolumn, precolumnhplc, hplcisocratic, isocraticmethanol, methanolseparation, separationmethod
Parallel screening for chiral separation of methyl phenyl sulfoxide
Application Note ► Parallel screening for chiral separation of methyl phenyl sulfoxide Category Matrix Method Keywords Analytes ID Chiral analysis HPLC parallel chiral screening, cellulose based chiral stationary phases, Eurocel 01, Eurocel 02 Methyl phenyl sulfoxide VCR1, January 2008 Summary…
Key words
chiral, chiralcolumn, columnsulfoxide, sulfoxidehex, hexeluent, eluentphenyl, phenylmethyl, methylrequest, requestscreening, screeningpolysaccharide, polysaccharidechromgate, chromgateupon, uponcellulose, celluloseparallel, paralleletoh
Chiral separation of Diclofop – methyl
Chiral separation of Diclofop – methyl Method HPLC VCR0023J Chiral HPLC Column: Eurocel 01, 250 x 4.6 mm ID with precolumn Order No. Phase: Eurocel 01, 5 µm Conditions: Eluent: Gradient: Flow rate: Temperature: Volume: Detection: UV at 210 nm…
Key words
chiral, chiraldiclofop, diclofopprecolumn, precolumnhplc, hplcisocratic, isocraticmethyl, methylseparation, separationmethod
Chiral separation of Dihydrofuranone
Chiral separation of Dihydrofuranone Method HPLC VCR0024J Chiral HPLC Column: Eurocel 01, 250 x 4.6 mm ID with precolumn Phase: Eurocel 01, 5 µm Conditions: Eluent: Gradient: Flow rate: Temperature: Volume: Detection: UV at 210 nm Substances: Dihydrofuranone, 4-(s-tolyl)dihydrofuran-2(3M)-one Keywords:…
Key words
chiral, chiraldihydrofuranone, dihydrofuranoneprecolumn, precolumnhplc, hplcisocratic, isocraticseparation, separationmethod
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike