Analysis of Chloride and Sulfate Ions in Fuel Ethanol - No. 446
Applications | 2023 | ShimadzuInstrumentation
Fuel ethanol may contain inorganic anions such as chloride and sulfate that can impact equipment corrosion, catalyst poisoning and fuel quality compliance. Precise quantification of these ions ensures adherence to ASTM D7319 standards and supports quality control in biofuel production.
This study demonstrates the determination of chloride (Cl−) and sulfate (SO42−) in fuel ethanol using suppressed ion chromatography. The goal is to provide a rapid, reliable and sensitive method according to ASTM D7319-07, offering an alternative to conventional titration or spectrophotometric techniques.
A suppressed conductivity detector monitors ionic elution after separation on an anion exchange column. The mobile phase consists of 0.6 mmol/L sodium carbonate and 12 mmol/L sodium hydrogen carbonate at a flow rate of 1.0 mL/min. The column temperature is maintained at 30 °C and sample injections of 5 µL are performed using standard vials.
The method achieved clear baseline separation of chloride and sulfate ions, with consistent peak shapes and reproducible retention. Sensitivity allowed detection at low ppm levels, meeting performance criteria for fuel ethanol analysis. The suppression technique minimized background conductivity, enhancing signal-to-noise ratio.
Advances may include coupling with mass spectrometry for broader ionic profiling, miniaturization for on-site testing, and automated sample preparation to increase throughput in industrial laboratories.
The suppressed ion chromatography method using Shim-pack columns offers a robust approach for chloride and sulfate determination in fuel ethanol. Its simplicity, speed and compliance with industry standards make it a valuable tool for biofuel quality control.
Ion chromatography, Consumables, LC columns
IndustriesEnergy & Chemicals
ManufacturerShimadzu
Summary
Importance of the Topic
Fuel ethanol may contain inorganic anions such as chloride and sulfate that can impact equipment corrosion, catalyst poisoning and fuel quality compliance. Precise quantification of these ions ensures adherence to ASTM D7319 standards and supports quality control in biofuel production.
Objectives and Study Overview
This study demonstrates the determination of chloride (Cl−) and sulfate (SO42−) in fuel ethanol using suppressed ion chromatography. The goal is to provide a rapid, reliable and sensitive method according to ASTM D7319-07, offering an alternative to conventional titration or spectrophotometric techniques.
Methodology
A suppressed conductivity detector monitors ionic elution after separation on an anion exchange column. The mobile phase consists of 0.6 mmol/L sodium carbonate and 12 mmol/L sodium hydrogen carbonate at a flow rate of 1.0 mL/min. The column temperature is maintained at 30 °C and sample injections of 5 µL are performed using standard vials.
Instrumentation Used
- Chromatography system: HIC-ESP suppressor ion chromatograph
- Analytical column: Shim-pack IC-SA2, 250 mm × 4.0 mm, 9 µm
- Guard column: Shim-pack IC-SA2 (G), 10 mm × 4.6 mm, 9 µm
- Detection: Conductivity
- Sample vial: Shimadzu 4 mL polypropylene
Main Results and Discussion
The method achieved clear baseline separation of chloride and sulfate ions, with consistent peak shapes and reproducible retention. Sensitivity allowed detection at low ppm levels, meeting performance criteria for fuel ethanol analysis. The suppression technique minimized background conductivity, enhancing signal-to-noise ratio.
Benefits and Practical Applications
- High specificity for target anions
- Compliance with ASTM D7319
- Rapid analysis suitable for routine QA/QC
- Reduced reagent consumption and waste
Future Trends and Potential Applications
Advances may include coupling with mass spectrometry for broader ionic profiling, miniaturization for on-site testing, and automated sample preparation to increase throughput in industrial laboratories.
Conclusion
The suppressed ion chromatography method using Shim-pack columns offers a robust approach for chloride and sulfate determination in fuel ethanol. Its simplicity, speed and compliance with industry standards make it a valuable tool for biofuel quality control.
References
- Shimadzu Corporation. Application News an_01-00265, First Edition August 2023, ERAS-1000-0446.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Analysis of Chloride ion and sulfate ion in chondroitin sulfate
2023|Shimadzu|Applications
ERAS-1000-0441 Ion Chromatography (IC) Shim-pack Series Shim-packTM IC-SA2, Analysis of Chloride ion and sulfate ion in chondroitin sulfate 441 Keywords: counter ion analysis, Japanese Pharmacopeia, suppressor ion chromatogram, anion analysis 1. Cl2. SO42- Column : Shim-pack IC-SA2, (250 mm L…
Key words
sulfate, sulfatechondroitin, chondroitinion, ionshim, shimpack, packguard, guardchloride, chloridecolumn, columnvial, vialseries, seriesmobile, mobilerate, ratephase, phasechromatography, chromatographydetection
Analysis of Anions by Ion Chromatography
2021|Shimadzu|Applications
ERAS-1000-0208 Ion Chromatography (IC) Shim-pack Series Shim-packTM IC-SA2 Analysis of Anions by Ion Chromatography 208 Keywords: fluoride, chloride, bromide, nitrate, nitrite, phosphate, sulfate µS/cm 5.0 3 1. F2. Cl3. NO24. Br5. NO36. PO437. SO42- 1 4.0 3.0 4 5 6…
Key words
anions, anionsshim, shimpack, packion, ionguard, guardchromatography, chromatographyvial, vialcolumn, columnseries, seriesmobile, mobilevolume, volumerate, ratephase, phaseinjection, injectiondetection
Analysis of Dissolved Hexavalent Chromium
2023|Shimadzu|Applications
ERAS-1000-0451 Ion Chromatography (IC) Shim-pack Series Shim-packTM IC-SA2 Analysis of Dissolved Hexavalent Chromium 451 Keywords: EPA 218.6, Post-column derivatization, Industrial wastewater analysis 1. Cr(VI) Column : Guard Column : Mobile Phase : Mobile phase Flow rate Post Column Reagents :…
Key words
post, postcolumn, columnreagents, reagentshexavalent, hexavalentcoil, coilchromium, chromiummobile, mobileshim, shimrate, ratephase, phasepack, packdissolved, dissolvedguard, guardflow, flowvial
Analysis of Iodate Ion and Iodide Ion
2022|Shimadzu|Applications
ERAS-1000-0368 Ion Chromatography (IC) Shim-pack Series Shim-packTM IC-SA2 Analysis of Iodate Ion and Iodide Ion 368 Keywords: Halogen mAU 22.5 20.0 210 nm 225 nm 2 17.5 1. Iodate ion (IO3-) (20 mg/L) 2. Iodide ion (I-) (14.5 mg/L) 15.0…
Key words
prominencetm, prominencetmiodate, iodateion, ionhic, hiciodide, iodideshim, shimpack, packguard, guardmau, maucolumn, columnseries, seriesmobile, mobilerate, ratephase, phasechromatography