Development and Validation of a Sensitive UHPLC-MS/MS Method for Teduglutide Analysis in Human Plasma on LCMS-8060RX

Posters | 2025 | Shimadzu | ASMSInstrumentation
LC/MS, LC/MS/MS, LC/QQQ
Industries
Clinical Research
Manufacturer
Shimadzu

Summary

Significance of the Topic


The accurate and sensitive quantification of teduglutide, a synthetic GLP-2 analog used in short bowel syndrome therapy, is critical for bioequivalence assessments and therapeutic monitoring. Reliable plasma analysis ensures proper dose optimization, supports generic formulation development, and underpins regulatory compliance in pharmacokinetic and clinical studies.

Study Objectives and Overview


This study aimed to develop and partially validate a robust ultra-high performance liquid chromatography tandem mass spectrometry (UHPLC-MS/MS) method for precise measurement of teduglutide in human plasma. Key goals included achieving a wide dynamic range, high recovery, minimal matrix effects, and full compliance with ICH M10 guidelines.

Methodology and Instrumentation


Sample Preparation:
  • Human K₂EDTA plasma was spiked with calibration standards (0.5–100 ng/mL) and QC samples (0.5–50 ng/mL).
  • Protein precipitation with methanol, centrifugation, followed by SPE on mixed-mode anion exchange cartridges.
  • Wash steps with ammonia/methanol:water and acidified acetonitrile, elution with formic acid in acetonitrile:water.

Chromatography and Mass Spectrometry:
  • Shimadzu Nexera X2 UHPLC with Shim-pack Scepter C18 column (1.9 µm, 100 × 3.0 mm).
  • Mobile phase A: 0.1% formic acid in water; B: 0.1% formic acid in acetonitrile; gradient elution at 0.5 mL/min and 50 °C.
  • Shimadzu LCMS-8060RX ESI-MS/MS in MRM mode, optimized gas flows and temperatures.

Main Results and Discussion


Selectivity and Linearity:
The method demonstrated minimal endogenous interference (<5.7%) across six plasma lots. An eight-point calibration curve (0.5–100 ng/mL) with 1/x² weighting achieved excellent linearity (r² = 0.9930).

Precision and Accuracy:
Intra- and inter-day precision (CV) remained below 10.7%, while accuracy ranged from 100.2% to 105.4% across LLOQ, LQC, MQC, and HQC levels.

Recovery and Matrix Effect:
Global recovery averaged 97.83% (RSD 2.43%). Matrix factor values at LQC and HQC were 0.96 and 0.98, indicating negligible ion suppression or enhancement.

Carry-Over:
No detectable carry-over was observed following the highest calibration standard injection.

Benefits and Practical Applications


The validated UHPLC-MS/MS method offers rapid throughput, high sensitivity (LLOQ 0.5 ng/mL), and reliable quantification, making it suitable for pharmacokinetic profiling, bioequivalence trials, and therapeutic drug monitoring of teduglutide. The streamlined sample preparation and robust chromatographic performance support high sample workloads in research and quality control laboratories.

Future Trends and Potential Applications


Advancements in microflow UHPLC and high-resolution MS may further lower detection limits and reduce solvent consumption. Integration with automated sample preparation platforms and AI-driven data analysis could enhance throughput and data quality. Expanding the method to quantify related peptide analogs will support broader clinical applications and novel GLP-2 analog development.

Conclusion


A sensitive and reliable UHPLC-MS/MS assay for teduglutide in human plasma was successfully developed and partially validated according to regulatory guidelines. The method demonstrates excellent linearity, precision, accuracy, and recovery with negligible matrix effects and carry-over, positioning it as a valuable tool for bioanalytical studies.

Reference


  • ICH M10 Bioanalytical Method Validation and Study Sample Analysis. Step 4, May 2022.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Sub-picogram level Bio-analytical method for quantification of Desmopressin in human plasma using LCMS-8060NX
WP 610 Sub-picogram level Bio-analytical method for quantification of Desmopressin in human plasma using LCMS-8060NX Chaitanya Krishna Atmakuri1, Avinash B Gaikwad1, Yogesh G Arote1 1 ADC-Shimadzu Analytical (India) Pvt. Ltd., Nerul, Navi Mumbai-400706, Maharashtra, India. 1. Introduction Desmopressin (dDAVP), a…
Key words
desmopressin, desmopressinext, extaqsample, aqsamplepesample, pesamplelqc, lqchqc, hqcextracted, extractedlloq, lloqmatrix, matrixplasma, plasmacarryover, carryoverrsd, rsdsample, sampleaccuracy, accuracyday
Highly Sensitive and Selective Method for Estimation of Formoterol at Sub-pg/mL in Human Plasma Using Shimadzu LCMS-8060NX
MP 803 Highly Sensitive and Selective Method for Estimation of Formoterol at Sub-pg/mL in Human Plasma Using Shimadzu LCMS-8060NX Avinash B Gaikwad1, Chaitanya Krishna Atmakuri1, Yogesh G Arote1 1 ADC-Shimadzu Analytical (India) Pvt. Ltd., Nerul, Navi Mumbai-400706, Maharashtra, India. 1.…
Key words
formoterol, formoterolext, extratio, ratioextracted, extractedsample, samplematrix, matrixmean, meanlqc, lqchqc, hqcarea, arearesponse, responselloq, lloqfactor, factorprelabelled, prelabelledcarryover
Highly Sensitive and Selective Method for Estimation of Formoterol at Sub-pg/mL in Human Plasma Using Shimadzu LCMS-8060NX
Liquid Chromatograph Mass Spectrometer LCMS-8060NX Application News Highly Sensitive and Selective Method for Estimation of Formoterol at Sub-pg/mL in Human Plasma Using Shimadzu LCMS-8060NX Avinash Gaikwad1, Chaitanya krishna Atmakuri1, Yogesh Arote1, Jitendra Kelkar2, Pratap Rasam2 1 ADC - Shimadzu Analytical…
Key words
formoterol, formoterolhqc, hqcprecision, precisionmatrix, matrixrsd, rsdaccuracy, accuracynominal, nominalsample, sampleextracted, extractedext, extlloq, lloqlqc, lqcconc, concfactor, factormethod
Rapid, Sensitive and Direct Quantitation of Tiotropium at sub-pg/mL in Plasma using Shimadzu LCMS-8060NX
TP 794 Rapid, Sensitive and Direct Quantitation of Tiotropium at sub-pg/mL in Plasma using Shimadzu LCMS-8060NX Yogesh G Arote1 , Avinash B Gaikwad1, Chaitanya Krishna A1 1 ADC-Shimadzu Analytical (India) Pvt. Ltd., Nerul, Navi Mumbai-400706, Maharashtra, India.. 1. Introduction 4.…
Key words
tiotropium, tiotropiumext, extratio, ratioextracted, extractedlloq, lloqsample, samplemean, meanrsd, rsdlqc, lqchqc, hqcmatrix, matrixresponse, responsefactor, factorarea, areacholinergic
Other projects
GCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike