Agilent InfinityLab Pro iQ Series Site Preparation Checklist
Manuals | 2025 | Agilent TechnologiesInstrumentation
Proper site preparation is critical to ensure that sophisticated analytical instruments perform reliably and deliver accurate results over their service life. By addressing space, environmental, utility, and safety requirements before delivery, laboratories can minimize installation delays, reduce troubleshooting, and maximize return on investment.
This document serves as a comprehensive pre-installation checklist for the Agilent InfinityLab Pro iQ Series. It guides customers through all responsibilities needed to prepare laboratory space, utilities, and supplies. Ultimately, it aims to streamline the handover from delivery to operational readiness of LC-MS systems.
The checklist covers the following key system components and related requirements:
Key site preparation parameters include:
Adhering to this checklist minimizes installation setbacks, ensures compliance with safety and environmental regulations, and equips laboratory personnel with the information to manage utilities and consumables effectively. Early involvement of users during install promotes knowledge transfer of operation, maintenance, and safety procedures.
Future advancements may include digital site-readiness assessments, automated notifications for utility thresholds, remote monitoring of environmental parameters, and integration of green laboratory practices. Standardized electronic checklists could further accelerate installations and support predictive maintenance.
Thorough site preparation is a foundational step in achieving reliable LC-MS performance. This structured checklist empowers laboratories to meet all prerequisites for the Agilent InfinityLab Pro iQ Series, reducing downtime and safeguarding instrument integrity from day one.
Agilent Technologies, Inc. Agilent CrossLab Start Up Services: Agilent InfinityLab Pro iQ Series Site Preparation Checklist, Document Number D0129562, Revision A.01, Issued 5 June 2025.
LC/MS, LC/SQ
IndustriesOther
ManufacturerAgilent Technologies
Summary
Importance of the topic
Proper site preparation is critical to ensure that sophisticated analytical instruments perform reliably and deliver accurate results over their service life. By addressing space, environmental, utility, and safety requirements before delivery, laboratories can minimize installation delays, reduce troubleshooting, and maximize return on investment.
Objectives and overview
This document serves as a comprehensive pre-installation checklist for the Agilent InfinityLab Pro iQ Series. It guides customers through all responsibilities needed to prepare laboratory space, utilities, and supplies. Ultimately, it aims to streamline the handover from delivery to operational readiness of LC-MS systems.
Methodology and instrumentation
The checklist covers the following key system components and related requirements:
- Primary instrument: Agilent InfinityLab Pro iQ Series (models G6160B Pro iQ and G6170A Pro iQ Plus).
- Vacuum components: MS45 rough pump or Edwards nXR40i dry pump.
- Ion sources: ESI, APCI, MMI, or AJS for mass spectrometry.
- High-pressure liquid chromatography (HPLC) modules when stacked above the MS detector.
- Supporting utilities: nitrogen gas supply, exhaust venting, electrical power, and environmental controls.
- Operating supplies: LC/MS-grade solvents and reagents, waste management containers, and basic tool kits.
Main results and discussion
Key site preparation parameters include:
- Space and weight: Adequate bench dimensions with at least 30 cm clearance on each side and support for up to 68 kg of stacked modules. Avoid overhanging shelves and vibrations.
- Environmental conditions: Maintain 15 – 35 °C (±3 °C) with non-condensing humidity (< 85 % RH), altitude limits (≤ 3 300 m or 3 700 m depending on pump), and vibration-free surfaces.
- Exhaust venting: Two independent negative-pressure lines for rough pump and spray chamber, 1/2″ ID tubing, flow rates up to 33.9 ft³/h depending on source. Separate venting is mandatory to preserve warranty.
- Equipment stacking: Sequence HPLC detector, column holder, pump, and solvent tray on top of the MS. Stack height ≤ 74 cm and total weight ≤ 68 kg to ensure stability.
- Power requirements: Single-phase 100 – 240 VAC @ 50/60 Hz or 208-Wye configuration with isolated, noise-free grounding. Outlets rated 10 – 16 A and line voltage tolerance ±10/–5 %. UPS must not create floating ground.
- Nitrogen supply: LN₂ or generator providing ≥ 95 % purity, hydrocarbon-free, 80 – 100 psi pressure, flow rates up to 30 L/min for AJS-equipped systems.
- Operating supplies and tools: LC/MS grade solvents (water, methanol, acetonitrile, formic acid), waste containers with secondary containment and charcoal filters, and Agilent toolkit accessories.
Benefits and practical application of the method
Adhering to this checklist minimizes installation setbacks, ensures compliance with safety and environmental regulations, and equips laboratory personnel with the information to manage utilities and consumables effectively. Early involvement of users during install promotes knowledge transfer of operation, maintenance, and safety procedures.
Future trends and potential applications
Future advancements may include digital site-readiness assessments, automated notifications for utility thresholds, remote monitoring of environmental parameters, and integration of green laboratory practices. Standardized electronic checklists could further accelerate installations and support predictive maintenance.
Conclusion
Thorough site preparation is a foundational step in achieving reliable LC-MS performance. This structured checklist empowers laboratories to meet all prerequisites for the Agilent InfinityLab Pro iQ Series, reducing downtime and safeguarding instrument integrity from day one.
Reference
Agilent Technologies, Inc. Agilent CrossLab Start Up Services: Agilent InfinityLab Pro iQ Series Site Preparation Checklist, Document Number D0129562, Revision A.01, Issued 5 June 2025.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Agilent Cary Sipper Accessory Site Preparation Checklist
2023|Agilent Technologies|Manuals
Agilent CrossLab Start Up Services Agilent Cary Sipper Accessory Site Preparation Checklist Thank you for purchasing an instrument from Agilent Technologies. CrossLab Start Up is focused on helping customers shorten the time it takes to start realizing the full value…
Key words
sipper, sippercary, caryaccessory, accessorysite, sitechecklist, checklistagilent, agilentpower, powerdescription, descriptionpreparation, preparationspecial, specialoperating, operatingservice, serviceweight, weightrequirements, requirementscustomer
G9853A Agilent Cary Sipper Site Preparation Checklist
2020|Agilent Technologies|Manuals
G9853A Agilent Cary Sipper Site Preparation Checklist Thank you for purchasing an Agilent accessory. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation is the…
Key words
sipper, sippercary, carychecklist, checklistsite, siteaccessory, accessorypower, powerdescription, descriptionoperating, operatingagilent, agilentspecial, specialpreparation, preparationcustomer, customerconsumption, consumptionweight, weightnotes
Agilent Cary 3500 UV-Vis - System Site Preparation Checklist
2023|Agilent Technologies|Manuals
Agilent CrossLab Start Up Services Agilent Cary 3500 UV-Vis System Site Preparation Checklist Thank you for purchasing an instrument from Agilent Technologies. CrossLab Start Up is focused on helping customers shorten the time it takes to start realizing the full…
Key words
description, descriptionpower, powercord, cordspecial, specialoperating, operatingrequirements, requirementsinstrument, instrumentweight, weightvis, visservice, servicecustomer, customersite, sitelbs, lbspeltier, peltierpart
Agilent Cary 3500 UV-Vis Systems Site Preparation Checklist
2020|Agilent Technologies|Manuals
Agilent Cary 3500 UV-Vis Systems Site Preparation Checklist Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation is…
Key words
checklist, checklistsite, sitepower, powerdescription, descriptioncord, cordspecial, specialoperating, operatingpreparation, preparationvis, visinstrument, instrumentcustomer, customernotes, notesweight, weightlbs, lbspeltier