Automated Optimization and Intelligent Reflex Screening on the Agilent 6495D LC/TQ

Applications | 2025 | Agilent TechnologiesInstrumentation
LC/MS, LC/MS/MS, LC/QQQ
Industries
Forensics
Manufacturer
Agilent Technologies

Summary

Automated Optimization and Intelligent Reflex Screening on a Triple Quadrupole LC/MS


Significance of the topic


Forensic toxicology laboratories face growing demands for rapid, reliable screening of biological samples. Traditional methods rely on manual tuning of multiple reaction monitoring (MRM) transitions and two-tiered workflows, which can be laborious and prone to analyst variability. The integration of automated optimization tools and intelligent reflex screening streamlines method development and sample processing, enhancing throughput and data consistency while reducing hands-on time.

Objectives and study overview


This application note demonstrates an end-to-end workflow on the Agilent 6495D LC/TQ instrument. The goals are to automate compound and source parameter optimization, implement a fast tier-1 screening method, and apply an intelligent reflex mechanism that triggers tier-2 confirmatory analysis automatically when thresholds are exceeded. A combined report merges results from both tiers without manual intervention.

Methodology and instrumentation


Instrumentation used:
  • Agilent 6495D triple quadrupole LC/MS with Jet Stream ESI source
  • Agilent 1290 Infinity II LC system
  • InfinityLab Poroshell 120 EC-C18 column (2.1×30 mm, 2.7 μm) for tier-1
  • InfinityLab Poroshell 120 Phenyl-Hexyl column (2.1×100 mm, 2.7 μm) for tier-2
Chromatographic conditions:
  • Tier-1 gradient: 4 min, 0–98% methanol at 0.5 mL/min
  • Tier-2 gradient: 6 min, 10–98% methanol at 0.5 mL/min
  • Column temperature: 55 °C; injection volume: 1 μL
Software tools:
  • MassHunter Optimizer for automated MRM transition and collision energy selection
  • MassHunter Source Optimizer for gas and sheath parameters
  • Intelligent Reflex workflow in MassHunter Acquisition for real-time decision making

Main results and discussion


Automated compound optimization scanned precursor and product ion pairs across collision energies, generating high-sensitivity transitions with minimal user input. Source optimization refined gas temperature, sheath gas flow, and voltages to achieve robust ionization for all analytes. The two-tier chromatography delivered orthogonal selectivity: a rapid four-minute screen and a more selective six-minute confirmation run. Intelligent Reflex monitored tier-1 signals and automatically reinjected presumptive positives under tier-2 conditions, eliminating manual decision steps. Combined reporting in MassHunter Quantitative Analysis aggregated screening and confirmation data, simplifying result review and reporting.

Benefits and practical applications


  • Reduces method development time by automating MRM and source parameter selection
  • Increases laboratory throughput through fast tier-1 screening
  • Ensures consistency and reproducibility across analysts
  • Eliminates manual triggering of confirmatory injections
  • Produces a single integrated report for streamlined data review

Future trends and potential applications


Advances in machine learning and adaptive control could further refine automated optimization, enabling real-time method adjustment based on sample complexity. Integration with laboratory information management systems (LIMS) may automate data transfer and reporting. Similar workflows may be extended to environmental monitoring, food safety, and clinical pharmacokinetics, where fast, reliable two-tier screening is essential.

Conclusion


The combination of MassHunter Optimizer, Source Optimizer, and Intelligent Reflex screening on the Agilent 6495D LC/TQ delivers a fully automated, two-tier workflow for forensic toxicology. This approach reduces hands-on time, improves consistency, and accelerates turnaround without sacrificing the confidence provided by orthogonal confirmation.

References


Agilent Technologies, Application Note 5994-8832EN, 2025.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Intelligent Reflex Fast Screening for Drugs in Urine
Intelligent Reflex Fast Screening for Drugs in Urine
2023|Agilent Technologies|Applications
Application Note Forensic Toxicology Intelligent Reflex Fast Screening for Drugs in Urine Data-dependent reinjection logic for screening and confirmation of presumptive positives using the intelligent reflex fast screening function in MassHunter acquisition software Agilent 6495D triple quadrupole LC/MS system Agilent…
Key words
reflex, reflexreinjection, reinjectionscreening, screeningintelligent, intelligentconfirmatory, confirmatorypresumptive, presumptiveacquisition, acquisitionmasshunter, masshunterfast, fastlogic, logicintelligence, intelligencemethod, methodexperiment, experimenttime, timetoxicology
Advances in Mass Spectrometry Instrument Intelligence
Advances in Mass Spectrometry Instrument Intelligence
2022|Agilent Technologies|Brochures and specifications
Advances in Mass Spectrometry Instrument Intelligence Overview Table of Contents There is a global trend towards more self‑aware, intelligent tools meant to make life easier. The field of liquid chromatography/mass spectrometry is no different with the move to automate difficult…
Key words
passed, passedresult, resultuptime, uptimemaximize, maximizereflex, reflexreinjection, reinjectiondevelop, developfaster, fasteroptimization, optimizationconfidence, confidenceensure, ensurechecktune, checktuneoverview, overviewintelligent, intelligention
Intelligent Reflex Targeted Reinjection Workflow for Pesticide Screening in Food Matrices
Application Note Food Safety Intelligent Reflex Targeted Reinjection Workflow for Pesticide Screening in Food Matrices Using the Agilent Revident LC/Q-TOF Authors Cate Simmermaker, Olivier Chevallier, Kai Chen, and Wei Wei Agilent Technologies, Inc. Abstract This application note details a methodology…
Key words
reflex, reflexreinjection, reinjectiontargeted, targetedscreening, screeningintelligent, intelligentscreener, screenerions, ionsacquisition, acquisitionuntargeted, untargetedworklist, worklistall, allwere, werequant, quantdata, datasuspects
Boost Efficiency and Productivity with Instrument Intelligence and Workflow Automation
Technical Overview Boost Efficiency and Productivity with Instrument Intelligence and Workflow Automation For LC/TQ, LC/TOF, and LC/Q-TOF workflows Introduction Ever-increasing pressure for laboratories to minimize operating costs while maximizing output demands innovative approaches to their daily work. Recent developments in…
Key words
reflex, reflexworkflows, workflowsintelligent, intelligentappend, appendemf, emfworkflow, workflowinstrument, instrumentswarm, swarmworklist, worklistlogic, logicmaintenance, maintenancereinjections, reinjectionsautotune, autotunefeedback, feedbackstatus
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike