Data Quality Enhancement using Software-controlled Acquisition Strategies in MALDI-TOF MS
Technical notes | 2025 | ShimadzuInstrumentation
MALDI-TOF mass spectrometry is a cornerstone technique in analytical chemistry, valued for its rapid analysis, simple sample preparation and high sensitivity. However, variability in co-crystallization of matrix and analyte leads to spot-to-spot differences in ionization efficiency, degrading spectral quality and reproducibility. Software controlled acquisition strategies aim to overcome these challenges by dynamically selecting only spectra that meet user-defined quality criteria, thereby enhancing data consistency and reliability in applications ranging from peptide characterization to microbial identification.
This report evaluates two software controlled acquisition strategies available in Shimadzu MALDI data acquisition software: find sweet spots and exhaust raster points. The goals are to demonstrate how these features can improve spectral quality, signal yield, resolution and sensitivity in diverse applications, including off-centered sample spots, topographically complex preparations, low-concentration analytes and microbial protein profiling.
All experiments were conducted on Shimadzu MALDI-8000 series and Axima MALDI-TOF instruments using the MALDI Solutions Data Acquisition, AuraSolution and QC Reporter software. Sample types included:
Data quality criteria were defined by mass range, minimum signal intensity, minimum resolution and maximum failed acquisitions. Strategies were applied to accumulate only spectra meeting these thresholds.
Off-centered Sample Spots
The software controlled data quality approach delivers:
Future developments may include the integration of machine learning for real-time optimization of quality criteria, extension of these strategies to MALDI imaging workflows, and application to other desorption/ionization techniques. Automated feedback mechanisms could further reduce user intervention and adapt to novel sample types.
Software controlled acquisition strategies in MALDI-TOF MS significantly improve data quality by selectively accumulating spectra that meet predefined criteria. Both find sweet spots and exhaust raster points workflows enhance signal yield, resolution and sensitivity across diverse applications, offering robust solutions for high-throughput and high-confidence mass spectrometry analysis.
Shimadzu MALDI-8000 series and Axima MALDI-TOF mass spectrometers with MALDI Solutions Data Acquisition, AuraSolution and QC Reporter software.
No external literature references provided.
MALDI, LC/TOF, LC/MS, Software
IndustriesOther
ManufacturerShimadzu
Summary
Significance of the Topic
MALDI-TOF mass spectrometry is a cornerstone technique in analytical chemistry, valued for its rapid analysis, simple sample preparation and high sensitivity. However, variability in co-crystallization of matrix and analyte leads to spot-to-spot differences in ionization efficiency, degrading spectral quality and reproducibility. Software controlled acquisition strategies aim to overcome these challenges by dynamically selecting only spectra that meet user-defined quality criteria, thereby enhancing data consistency and reliability in applications ranging from peptide characterization to microbial identification.
Objectives and Overview of the Study
This report evaluates two software controlled acquisition strategies available in Shimadzu MALDI data acquisition software: find sweet spots and exhaust raster points. The goals are to demonstrate how these features can improve spectral quality, signal yield, resolution and sensitivity in diverse applications, including off-centered sample spots, topographically complex preparations, low-concentration analytes and microbial protein profiling.
Methodology and Instrumentation
All experiments were conducted on Shimadzu MALDI-8000 series and Axima MALDI-TOF instruments using the MALDI Solutions Data Acquisition, AuraSolution and QC Reporter software. Sample types included:
- Peptides: Glu1-Fibrinopeptide B and ACTH 7-38 fragment in CHCA matrix
- Oligonucleotides: 12-mer prepared with 3-Hydroxypicolinic acid matrix
- Protein digest: BSA tryptic digest at low femtomole concentrations
- Microbial smears: Escherichia coli in CHCA matrix
Data quality criteria were defined by mass range, minimum signal intensity, minimum resolution and maximum failed acquisitions. Strategies were applied to accumulate only spectra meeting these thresholds.
Main Results and Discussion
Off-centered Sample Spots
- The find sweet spots strategy reduced acquisition time by selecting promising raster positions in two passes, halving analysis time compared to exhaustive scanning.
- Signal yield increased by over twofold, with comparable mass resolution.
- For the 12-mer oligonucleotide, find sweet spots doubled resolution by focusing on flatter central areas and rejecting spectra from raised edges.
- ACTH 7-38 fragment achieved full isotopic resolution on a compact linear TOF when using resolution thresholds above 6000.
- Low-concentration BSA digest spectra acquired with data quality met Mascot peptide mass fingerprinting identification criteria, whereas conventional acquisition failed.
- The exhaust raster points strategy improved detection of high-mass bacterial proteins by ablating surface matrix and salts through repeated acquisitions until quality criteria were met.
Benefits and Practical Applications of the Method
The software controlled data quality approach delivers:
- Enhanced reproducibility and spectral consistency across heterogeneous sample spots
- Increased throughput by minimizing uninformative laser shots
- Improved mass resolution and sensitivity for challenging analytes
- Reliable microbial profiling for clinical and environmental laboratories
Future Trends and Potential Applications
Future developments may include the integration of machine learning for real-time optimization of quality criteria, extension of these strategies to MALDI imaging workflows, and application to other desorption/ionization techniques. Automated feedback mechanisms could further reduce user intervention and adapt to novel sample types.
Conclusion
Software controlled acquisition strategies in MALDI-TOF MS significantly improve data quality by selectively accumulating spectra that meet predefined criteria. Both find sweet spots and exhaust raster points workflows enhance signal yield, resolution and sensitivity across diverse applications, offering robust solutions for high-throughput and high-confidence mass spectrometry analysis.
Used Instrumentation
Shimadzu MALDI-8000 series and Axima MALDI-TOF mass spectrometers with MALDI Solutions Data Acquisition, AuraSolution and QC Reporter software.
References
No external literature references provided.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Use of Single Spot Imaging in MALDI Method Development
2025|Shimadzu|Technical notes
012-MO542-EN Technical Technical Report Report Use of Single Spot Imaging in MALDI Method Development Caroline J. Jones Shimadzu, Manchester, UK Abstract: In the process of analytical method development using MALDI, the choice of matrix, ratios of solvents and data acquisition…
Key words
annular, annularspot, spotraster, rasteracquisition, acquisitionmaldi, maldispatial, spatialcircular, circularimaging, imagingdistribution, distributionstrategy, strategyshots, shotsstrategies, strategiesanalyte, analytehomogenous, homogenousdata
FlexiFocus ™ Slides: Chemically Modified Sample-Focussing MALDI Targets to Enhance Productivity and Throughput
2023|Shimadzu|Technical notes
C146-E479 FlexiFocus™ Slides: Chemically Modified Sample-Focussing MALDI Targets to Enhance Productivity and Throughput Technical Report Simona Salivo1 A b stra c t: The FlexiFocus™ slides are chemically modified sample-focussing MALDI targets that have been developed to facilitate high-throughput, automated analysis…
Key words
flexifocus, flexifocusintensity, intensitychca, chcafocussing, focussingtarget, targetcentre, centrebsa, bsaslide, slidespotting, spottingdigest, digestslides, slidesacgtacgtacgtacgtacgt, acgtacgtacgtacgtacgtmaldi, maldidowex, dowexchemically
Modified target surface to enhance performance during MALDI analysis
2021|Shimadzu|Applications
MALDI Mass Spectrometry Application News Modified target surface to enhance performance during MALDI analysis Andreas Baumeister1, Maximilian Benz2, Wolfgang Sipos2 1 Shimadzu Europa GmbH, 2 Aquarray GmbH User Benefits u Higher number of samples on the target increases the throughput.…
Key words
aquarray, aquarrayomniphilic, omniphilicomniphobic, omniphobicmaldi, maldipatterned, patternedspots, spotsmicroarray, microarrayshimadzu, shimadzudispenser, dispensersurface, surfacedroplet, dropletmodified, modifiedmarrying, marryingmaldimini, maldiminiglufib
Compensation of Sample Topography Variability in MALDI-TOF MS using Lock Mass Correction
2025|Shimadzu|Technical notes
MO537A Compensation of Sample Topography Variability in MALDI-TOF MS using Lock Mass Correction Technical Technical Report Report Simona Salivo1 Abstract: In the typical MALDI process, the sample is mixed with a UV-energy absorbent MALDI matrix to aid ionisation. Depending on…
Key words
lock, lockmass, massmaldi, malditopography, topographycorrection, correctionacgtacgtacg, acgtacgtacgtacgtacgt, tacgtacgterror, errorionview, ionviewatctttggtgtt, atctttggtgttspatial, spatialdistribution, distributionresolution, resolutionimaging, imagingvariability