Automated and accurate component detection using reference mass spectra

Technical notes | 2018 | Thermo Fisher ScientificInstrumentation
GC/MSD, Software, LC/MS, IC-MS
Industries
Manufacturer
Thermo Fisher Scientific

Summary

Importance of the Topic


The identification of chemical components in complex matrices is critical across food safety, environmental monitoring, and clinical analysis.
GC-MS with full scan acquisition provides rich spectral data but can generate false positives in complex samples or when retention times shift.
Using reference mass spectra to augment targeted screening enhances confidence in compound detection.

Objectives and Scope of the Study


This study aims to demonstrate the capability of the Chromeleon Chromatography Data System to apply laboratory reference mass spectra during full scan screening.
The goals include reducing false positives, accommodating retention time variations, and enabling tentative identification when retention times are unknown.

Methodology and Instrumentation


The Chromeleon CDS processing method was configured with components linked to reference mass spectra from commercial libraries or prior injections.
Component match settings include spectrum only or spectrum and time matching with flexible retention time windows.

Instrumentation Used

  • Gas chromatograph coupled to mass spectrometer with electron ionization source
  • Thermo Scientific Chromeleon Chromatography Data System software
  • NIST Mass Spectral Library (NIST 17) and Search Program

Key Results and Discussion


An unspiked tea sample produced false positive pesticide identifications under standard retention time and ion ratio criteria.
By assigning reference spectra, Chromeleon CDS correctly rejected false positives and confirmed analytes based on spectral similarity.
In cases of retention time shifts such as 100 fg octafluoronaphthalene, the software identified the correct peak at 3.79 min despite a 4.00 min expected time due to spectrum-based matching.
Spectrum only matching allowed accurate detection even when retention times were unknown, facilitating preliminary screening prior to standard injections for RT confirmation.

Benefits and Practical Applications of the Method

  • Increased confidence and reduced false positives in complex matrices
  • Adaptability to retention time drift and unknown RTs without additional standards
  • Efficient targeted screening of large compound lists in food, environmental, or forensic analysis

Future Trends and Potential Applications


Expansion of reference spectrum strategies to LC-MS and IC-MS platforms.
Integration with machine learning algorithms for adaptive spectral matching and peak picking.
Development of cloud based and collaborative spectral libraries for real time identification.
Real time QA QC workflows using reference spectra during routine analysis.

Conclusion


The incorporation of reference mass spectra into Chromeleon CDS significantly enhances the reliability of component detection.
Eliminating false positives and allowing spectral only matching streamlines targeted screening workflows.
This approach delivers robust data quality and time savings in analytical laboratories.

References

  1. Hübschmann H J Handbook of GC MS Wiley VCH 2015
  2. Zweigenbaum J Mass Spectrometry in Food Safety Methods and Protocols Humana Press 2011
  3. Bruner F Gas Chromatographic Environmental Analysis Principles Techniques Instrumentation Wiley VCH 1993
  4. Gerhards P Bons U Sawazki J Szigan J Wertmann A GC MS in Clinical Chemistry Wiley VCH 1999
  5. NIST EPA NIH Mass Spectral Library NIST 17 Mass Spectral Search Program Version 2.3 NIST Mass Spectrometry Data Center 2017

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
How can the Orbitrap Exploris GC 240 mass spectrometer and Compound Discoverer software lead discovery in metabolomics studies?
SMART NOTE 000070 Smart Notes How can the Orbitrap Exploris GC 240 mass spectrometer and Compound Discoverer software lead discovery in metabolomics studies? For laboratories whose primary focus is metabolomic studies, the capability to comprehensively and accurately characterize and quantify…
Key words
metabolomics, metabolomicsdiscoverer, discovererpci, pciorbitrap, orbitrapcompound, compoundhram, hrammass, massmolecular, molecularsoftware, softwaredisplayed, displayedlibrary, libraryputative, putativestatistical, statisticaldeconvolution, deconvolutiondata
Screening Workflow Using Ion Mobility-Mass Spectrometry for The Analysis of Extractables and Leachables
Application Note Screening Workflow Using Ion Mobility-Mass Spectrometry for The Analysis of Extractables and Leachables Jane Cooper, Rachel Sanig, Gitte Barknowitz, Baiba Cabovska Waters Corporation Abstract Identification and characterization of extractable and leachable (E&L) components in various industries is growing…
Key words
unifi, unifitargeted, targetedscreening, screeningmobility, mobilityaids, aidsworkflow, workflowunknown, unknownscientific, scientificelucidation, elucidationextractable, extractableion, ionims, imsidentification, identificationspectrometry, spectrometrynon
Automated Targeted Screening of Leachables in Pharmaceutical QC Labs Using the Agilent LC/MSD XT and Agilent OpenLAB CDS
Automated Targeted Screening of Leachables in Pharmaceutical QC Labs Using the Agilent LC/MSD XT and Agilent OpenLAB CDS Application Note Pharmaceuticals QA/QC Authors Abstract Syed Salman Lateef, Vinayak AK, and This Application Note describes a routine analysis workflow for pharmaceutical…
Key words
leachables, leachablestargeted, targetedodp, odpscreening, screeningcds, cdsopenlab, openlabwere, weresim, simlibrary, librarycompounds, compoundsused, usedophthalmic, ophthalmicsuspect, suspectphthalic, phthalicanhydride
Accelerate your Quantitative LC/MS Workflows with a Fully Integrated UHPLC-QQQ Platform
Accelerate your Quantitative LC/MS Workflows with a Fully Integrated UHPLC-QQQ Platform Remco Swart Pittcon March 8-12, 2015 The world leader in serving science 1 Outline • Thermo Scientific™ Vanquish™ UHPLC System • Thermo Scientific™ TSQ Endura™ Triple Quadrupole Mass Spectrometer…
Key words
tsq, tsqvanquish, vanquishquantiva, quantivacomponent, componentendura, enduradata, datanew, newcds, cdsmore, morechromeleon, chromeleonprocessing, processingreporting, reportingthermo, thermoscientific, scientificreport
Other projects
GCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike