Sensitive and Accurate Quantitation of the Antibody-Drug Conjugate Ado-Trastuzumab Emtansine in Rat Plasma
Applications | 2017 | SCIEXInstrumentation
Accurate quantitation of antibody–drug conjugates (ADCs) such as ado-trastuzumab emtansine is essential in pharmacokinetic and bioanalytical studies to support dose selection, safety assessment, and regulatory approval. The complex nature of plasma matrices and heterogeneity of lysine-linked ADCs necessitate methods with high sensitivity, broad dynamic range, and robust reproducibility.
This work developed and validated a trap-and-elute microflow LC-MS/MS method using the SCIEX BioBA sample enrichment kit combined with an M3 MicroLC-TE system and QTRAP 6500+ mass spectrometer. The goal was to improve lower limit of quantitation (LLOQ), signal-to-noise ratio, and linear dynamic range for ADC quantitation in rat plasma, and to benchmark performance against a conventional direct-injection LC-MS/MS workflow.
Sample Preparation
The trap-and-elute microflow approach yielded a 5× lower LLOQ (1 ng/mL vs. 5 ng/mL), 4× higher signal-to-noise ratios, and a 5-order dynamic range (r>0.995) for peptides IYPTNGYTR and FTISADTSK. Carryover remained below 0.01%. The microflow method reduced solvent consumption and protected the MS source, without compromising throughput.
This workflow enables sensitive quantitation of ADCs at low nanogram levels with minimal sample volume. The trap-and-elute format extends column life, lowers operational costs, and is well suited for high-throughput PK, toxicology, and biomarker studies in pharmaceutical research.
Advancements may include further miniaturization to nano-flow LC, full automation of immunocapture and digestion steps, integration with laboratory information management systems, and adaptation to emerging biotherapeutics such as bispecific antibodies and antibody fragments.
The integration of the SCIEX BioBA enrichment kit with trap-and-elute microflow LC-MS/MS on the M3 MicroLC and QTRAP 6500+ delivers a robust, high-sensitivity assay for ADC quantitation in complex matrices, meeting current bioanalytical demands.
1. Antibody Drug Conjugate Bioanalysis using the BioBA Solution, SCIEX Technical Note RUO-MKT-02-4864-A.
2. BioBA High Capacity Sample Enrichment Kit Protocol, RUO-IDV-05-3429-A.
3. Automating Biologics Signature Peptide Quantitation, SCIEX Technical Note RUO-MKT-02-4029-B.
4. Improving Sensitivity in Bioanalysis using Trap-and-Elute MicroLC-MS/MS, SCIEX Technical Note RUO-MKT-02-2849.
5. Improving Sensitivity for an Immunocapture LC-MS/MS Assay of Infliximab in Rat Plasma Using Trap-and-Elute MicroLC-MS/MS, SCIEX Technical Note RUO-MKT-02-3413.
Consumables, LC/MS, LC/MS/MS, LC/QTRAP
IndustriesClinical Research
ManufacturerSCIEX
Summary
Significance of the Topic
Accurate quantitation of antibody–drug conjugates (ADCs) such as ado-trastuzumab emtansine is essential in pharmacokinetic and bioanalytical studies to support dose selection, safety assessment, and regulatory approval. The complex nature of plasma matrices and heterogeneity of lysine-linked ADCs necessitate methods with high sensitivity, broad dynamic range, and robust reproducibility.
Objectives and Study Overview
This work developed and validated a trap-and-elute microflow LC-MS/MS method using the SCIEX BioBA sample enrichment kit combined with an M3 MicroLC-TE system and QTRAP 6500+ mass spectrometer. The goal was to improve lower limit of quantitation (LLOQ), signal-to-noise ratio, and linear dynamic range for ADC quantitation in rat plasma, and to benchmark performance against a conventional direct-injection LC-MS/MS workflow.
Methodology and Instrumentation
Sample Preparation
- Magnetic streptavidin bead immunocapture of total antibody via anti-human IgG.
- Optimized reduction, alkylation, and trypsin/Lys-C digestion steps to enhance peptide recovery.
- Spiking of ado-trastuzumab emtansine into Sprague-Dawley rat plasma over 0.5–100 000 ng/mL range with SILuMab internal standard.
- Microflow LC: SCIEX M3 MicroLC-TE with trap column (10 × 0.3 mm) and analytical column (50 × 0.3 mm) at 10 µL/min.
- Traditional LC: Shimadzu Prominence HPLC with 100 × 2.1 mm Kinetex C18 at 500 µL/min.
- Mass Spectrometer: SCIEX QTRAP 6500+ with IonDrive Turbo V source and optimized MRM transitions for signature peptides.
Main Results and Discussion
The trap-and-elute microflow approach yielded a 5× lower LLOQ (1 ng/mL vs. 5 ng/mL), 4× higher signal-to-noise ratios, and a 5-order dynamic range (r>0.995) for peptides IYPTNGYTR and FTISADTSK. Carryover remained below 0.01%. The microflow method reduced solvent consumption and protected the MS source, without compromising throughput.
Benefits and Practical Applications
This workflow enables sensitive quantitation of ADCs at low nanogram levels with minimal sample volume. The trap-and-elute format extends column life, lowers operational costs, and is well suited for high-throughput PK, toxicology, and biomarker studies in pharmaceutical research.
Future Trends and Applications
Advancements may include further miniaturization to nano-flow LC, full automation of immunocapture and digestion steps, integration with laboratory information management systems, and adaptation to emerging biotherapeutics such as bispecific antibodies and antibody fragments.
Conclusion
The integration of the SCIEX BioBA enrichment kit with trap-and-elute microflow LC-MS/MS on the M3 MicroLC and QTRAP 6500+ delivers a robust, high-sensitivity assay for ADC quantitation in complex matrices, meeting current bioanalytical demands.
References
1. Antibody Drug Conjugate Bioanalysis using the BioBA Solution, SCIEX Technical Note RUO-MKT-02-4864-A.
2. BioBA High Capacity Sample Enrichment Kit Protocol, RUO-IDV-05-3429-A.
3. Automating Biologics Signature Peptide Quantitation, SCIEX Technical Note RUO-MKT-02-4029-B.
4. Improving Sensitivity in Bioanalysis using Trap-and-Elute MicroLC-MS/MS, SCIEX Technical Note RUO-MKT-02-2849.
5. Improving Sensitivity for an Immunocapture LC-MS/MS Assay of Infliximab in Rat Plasma Using Trap-and-Elute MicroLC-MS/MS, SCIEX Technical Note RUO-MKT-02-3413.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Antibody Drug Conjugate Bioanalysis using the BioBA Solution
2016|SCIEX|Applications
Antibody Drug Conjugate Bioanalysis using the BioBA Solution A Robust and Accurate Quantification of Ado-Trastuzumab Emtansine Using the BioBA Solution and the SCIEX QTRAP® 6500+ System Michael Deng and Ian Moore SCIEX, Concord, Canada Key Challenges of Antibody-Drug Conjugate Bioanalysis…
Key words
bioba, biobaiyptngytr, iyptngytrsignature, signaturesurfactant, surfactantemtansine, emtansinedtlmisr, dtlmisrantibody, antibodypeptide, peptidestd, stdado, adoadcs, adcsheterogeneous, heterogeneousbioanalysis, bioanalysisadc, adctrastuzumab
Ultra-Sensitive Quantification of Trastuzumab Emtansine in Mouse Plasma using Trap-Elute MicroLC MS Method
2018|SCIEX|Applications
Ultra-Sensitive Quantification of Trastuzumab Emtansine in Mouse Plasma using Trap-Elute MicroLC MS Method Featuring SCIEX QTRAP® 6500+ System with OptiFlow™ Turbo V source and M5 MicroLC system Lei Xiong, Ji Jiang, and Remco van Soest SCIEX, Redwood City, California, USA…
Key words
microflow, microflowimmunoaffinity, immunoaffinitymicrolc, microlcflow, flowphase, phasemobile, mobilerate, rateemtansine, emtansinetrap, trapvalue, valueslurry, slurryparameter, parameterquantification, quantificationelute, elutebead
Improving Sensitivity for Trastuzumab Emtansine Quantification using Trap-Elute MicroLC-MS with Large Volume Sample Loading
2018|SCIEX|Applications
Improving Sensitivity for Trastuzumab Emtansine Quantification using Trap-Elute MicroLC-MS with Large Volume Sample Loading Using SCIEX 6500+ QTRAP® System with OptiFlow™ Turbo V source and M5 MicroLC System Lei Xiong and Remco van Soest SCIEX, Redwood City, California, USA LC-MS…
Key words
emtansine, emtansinetrastuzumab, trastuzumabmicrolc, microlcvolume, volumemouse, mousesample, samplequantification, quantificationimmunoaffinity, immunoaffinityloading, loadingtrap, trapflow, flowslurry, slurryphase, phaseparameter, parameterelute
Overview of LC-MS Quantitative Solutions for Biotherapeutic Analysis
2019|SCIEX|Others
Overview of LC-MS Quantitative Solutions for Biotherapeutic Analysis Featuring SCIEX Instrumentation and Technologies Lei Xiong, Elliott Jones SCIEX, Redwood City, California, USA Figure 1. Summary of SCIEX Quantitative Solutions and Instrumentations to Solve Analytical Challenges in Biotherapeutic Industry. Darkest green…
Key words
sciex, sciexmrmhr, mrmhrworkflow, workflowprotein, proteinmicrolc, microlcquantification, quantificationions, ionsmrm, mrmqtrap, qtrapswath, swathquantitation, quantitationhrams, hramsspectrometers, spectrometersmobility, mobilityselexion