Analysis of Formaldehyde Using HPLC and Post-Column Derivatization with Acetylacetone
Applications | 2020 | ShimadzuInstrumentation
Formaldehyde is widely used in wood preservatives and resin products but is a known irritant and suspected carcinogen. In cosmetic formulations, regulatory agencies restrict or prohibit formaldehyde due to health concerns. Accurate monitoring at trace levels is essential for product safety and regulatory compliance.
The study aims to develop and validate a sensitive HPLC method using post-column derivatization with acetylacetone to quantify formaldehyde in cosmetics. The approach is based on established health science protocols and adapts ultra high performance liquid chromatography to achieve trace-level detection.
The analysis employs a Nexera XR ultra high performance liquid chromatography system configured with:
Calibration was linear from 0.01 to 1.0 mg/L with a correlation coefficient exceeding 0.9999. Retention time of the derivative was approximately 2.8 minutes. Matrix samples from shampoo, conditioner, and skin lotion were extracted with water and filtered before analysis. Spiked recovery rates ranged from 101 to 109 percent, confirming the method reliability at trace levels.
The validated HPLC method with post-column derivatization using acetylacetone demonstrates high sensitivity, accuracy, and robustness for formaldehyde analysis in cosmetic products. It meets regulatory requirements and supports quality assurance in manufacturing workflows.
1. Ministry of Health and Welfare of Japan Notification No. 331 Standards for Cosmetics 2000
2. European Commission Regulation (EC) No. 1223/2009 Annex III 2009
3. The Pharmaceutical Society of Japan Methods of Analysis in Health Science 2015
HPLC
IndustriesEnergy & Chemicals
ManufacturerShimadzu
Summary
Importance of the Topic
Formaldehyde is widely used in wood preservatives and resin products but is a known irritant and suspected carcinogen. In cosmetic formulations, regulatory agencies restrict or prohibit formaldehyde due to health concerns. Accurate monitoring at trace levels is essential for product safety and regulatory compliance.
Objectives and Overview of the Study
The study aims to develop and validate a sensitive HPLC method using post-column derivatization with acetylacetone to quantify formaldehyde in cosmetics. The approach is based on established health science protocols and adapts ultra high performance liquid chromatography to achieve trace-level detection.
Methodology and Instrumentation
The analysis employs a Nexera XR ultra high performance liquid chromatography system configured with:
- Auto sampler SIL-40C XR
- Column oven CTO-40C and CTO-40S
- Degassing unit DGU-403
- Pump LC-40B XR
- PDA detector SPD-M40 set at 414 nm
Main Results and Discussion
Calibration was linear from 0.01 to 1.0 mg/L with a correlation coefficient exceeding 0.9999. Retention time of the derivative was approximately 2.8 minutes. Matrix samples from shampoo, conditioner, and skin lotion were extracted with water and filtered before analysis. Spiked recovery rates ranged from 101 to 109 percent, confirming the method reliability at trace levels.
Benefits and Practical Applications
- The method offers high selectivity by detecting a specific derivatized product at 414 nm.
- No organic solvents in the mobile phase avoid peak distortion and simplify sample handling.
- The integrated temperature control of the column oven doubles as a chemical reactor for post-column derivatization.
- Fast analysis cycle and minimal sample preparation promote routine quality control in cosmetic testing.
Future Trends and Opportunities
- Integration of automated sample preparation to enhance throughput.
- Miniaturized and portable HPLC systems for on-site testing.
- Use of advanced detectors to push detection limits even lower.
- Adaptation to other regulatory matrices such as environmental or food analysis.
Conclusion
The validated HPLC method with post-column derivatization using acetylacetone demonstrates high sensitivity, accuracy, and robustness for formaldehyde analysis in cosmetic products. It meets regulatory requirements and supports quality assurance in manufacturing workflows.
References
1. Ministry of Health and Welfare of Japan Notification No. 331 Standards for Cosmetics 2000
2. European Commission Regulation (EC) No. 1223/2009 Annex III 2009
3. The Pharmaceutical Society of Japan Methods of Analysis in Health Science 2015
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Analysis of Formaldehyde Using Post-Column Derivatization with Acetylacetone
2021|Shimadzu|Applications
Application News No. L574 High Performance Liquid Chromatography Analysis of Formaldehyde Using Post-Column Derivatization with Acetylacetone Formaldehyde, which is used as an ingredient of wood preservatives and resin products, has attracted considerable attention because it is a causative agent of…
Key words
formaldehyde, formaldehydeacetylacetone, acetylacetonenexera, nexeraderivatization, derivatizationspiking, spikingcoil, coilmau, maupost, postwater, waterwithout, withoutreaction, reactioncolumn, columnspiked, spikeddissolved, dissolvedregent
Ultra High Performance Liquid Chromatograph Shimadzu Nexera series - Technical Reports / Applications
2020|Shimadzu|Guides
C190-E256 Ultra High Performance Liquid Chromatograph Nexera series Technical Reports / Applications The Nexera Series: Reinvent Your Workflow Just as the internet and smartphones have changed our lifestyles, and robots with AI capabilities continue to make our lives more convenient,…
Key words
acid, acidmau, mauanalysis, analysismobile, mobilemin, mincolumn, columnphase, phaseshim, shimnexera, nexerainjection, injectionpack, packorganic, organicatp, atpacids, acidshigh
Analysis of Polyether Antibiotics in Animal Feeds Using Post-Column Derivatization
2022|Shimadzu|Applications
Nexera™ lite High Performance Liquid Chromatograph/CRB-40 Application News Analysis of Polyether Antibiotics in Animal Feeds Using Post-Column Derivatization E. Ando, H. Terada User Benefits Three polyether antibiotics can be stably quantified over a wide concentration range in accordance with…
Key words
feed, feedbroilers, broilerschickens, chickenscoil, coilpolyether, polyetherreaction, reactionpotency, potencyantibiotics, antibioticsaverage, averagemau, mauanimal, animalcattle, cattlepost, postfeeds, feedsderivatization
Application Handbook Liquid Chromatography
2016|Shimadzu|Guides
Application Handbook Liquid Chromatography Introduction HPLC, UHPLC as well as SFC systems are able to quantitatively analyze substances in mixtures containing multiple ingredients by separating and detecting target substances. They are used to separate, quantify, qualify or purify single components…
Key words
sfe, sfeflowrate, flowrateprominence, prominencenews, newsnexera, nexeraanalysis, analysiscolumn, columnmin, minextraction, extractionsfc, sfcwithout, withoutpeaks, peaksmau, maumobile, mobilephase