Quantification of 25-hydroxyvitamin D2 and D3 in human plasma by liquid chromatography coupled to high-resolution Orbitrap mass spectrometry for clinical research
Applications | 2019 | Thermo Fisher ScientificInstrumentation
Vitamin D status is critical for bone health, calcium and phosphate homeostasis, and modulation of immune and cardiovascular functions. Precise measurement of its main circulating forms, 25-hydroxyvitamin D2 and D3, is essential in clinical research to diagnose deficiencies, monitor supplementation, and investigate associations with various diseases. High selectivity and sensitivity are required to accurately quantify low serum concentrations and to minimize interference from matrix components.
The primary goal was to develop and validate a robust LC–HRMS method for the quantification of 25-OH-vitamin D2 and D3 in human plasma. This included implementing an offline sample preparation workflow using a ClinMass® kit, achieving rapid chromatographic separation with online SPE, and comparing full-scan and parallel reaction monitoring (PRM) acquisition modes on an Orbitrap platform.
The sample preparation involved:
Chromatographic setup on a Vanquish Duo UHPLC system:
Mass spectrometry on a Q Exactive Plus quadrupole-Orbitrap:
Linearity and sensitivity:
Accuracy and precision:
Comparison of acquisition modes:
The developed method provides:
It is readily adoptable in research laboratories investigating vitamin D status, disease associations, or supplementation efficacy.
Emerging opportunities include:
A precise, fast, and reproducible LC–HRMS method for 25-OH-vitamin D2 and D3 quantification was established using offline sample prep, online SPE, and Orbitrap detection. Both FullMS and PRM modes met stringent criteria for sensitivity, linearity, accuracy, and precision, supporting its use in clinical research settings.
De Nardi C., Kern K., Peters S. Quantification of 25-hydroxyvitamin D2 and D3 in human plasma by liquid chromatography coupled to high-resolution Orbitrap mass spectrometry for clinical research. Technical Note No. 73234, Thermo Fisher Scientific, 2019.
LC/HRMS, LC/MS, LC/MS/MS, LC/Orbitrap
IndustriesClinical Research
ManufacturerThermo Fisher Scientific, RECIPE
Summary
Significance of the Topic
Vitamin D status is critical for bone health, calcium and phosphate homeostasis, and modulation of immune and cardiovascular functions. Precise measurement of its main circulating forms, 25-hydroxyvitamin D2 and D3, is essential in clinical research to diagnose deficiencies, monitor supplementation, and investigate associations with various diseases. High selectivity and sensitivity are required to accurately quantify low serum concentrations and to minimize interference from matrix components.
Objectives and Study Overview
The primary goal was to develop and validate a robust LC–HRMS method for the quantification of 25-OH-vitamin D2 and D3 in human plasma. This included implementing an offline sample preparation workflow using a ClinMass® kit, achieving rapid chromatographic separation with online SPE, and comparing full-scan and parallel reaction monitoring (PRM) acquisition modes on an Orbitrap platform.
Methodology and Instrumentation
The sample preparation involved:
- Addition of d6-25-hydroxyvitamin D3 internal standard to 50 µL plasma
- Protein precipitation with organic solvent, vortex mixing, cooling, centrifugation
- Transfer of supernatant for LC–MS analysis
Chromatographic setup on a Vanquish Duo UHPLC system:
- Online SPE cartridge for sample clean-up
- Analytical column separation with a 3.0-min total runtime
- Binary gradient to enrich and elute analytes
Mass spectrometry on a Q Exactive Plus quadrupole-Orbitrap:
- Atmospheric pressure chemical ionization (APCI) in positive mode
- FullMS and PRM acquisition to compare sensitivity and specificity
- TraceFinder software for data processing
Main Results and Discussion
Linearity and sensitivity:
- Calibration ranges: 9.8–81.0 ng/mL for D2, 9.0–78.9 ng/mL for D3
- Linear response with 1/x weighting across full range and down to LLOQ
- Lower limits of quantification (LLOQ): 4.92 ng/mL (D2) and 1.81 ng/mL (D3) in FullMS; PRM yielded 4.52 ng/mL for D3
Accuracy and precision:
- Bias within ±15% (±20% at LLOQ)
- Intra-assay CVs below 11.3%; inter-assay CVs below 9.3%
- Carryover below 0.4%
Comparison of acquisition modes:
- FullMS offered slightly lower LLOQ for D3
- PRM enhanced selectivity with comparable precision
Benefits and Practical Applications
The developed method provides:
- High selectivity and minimal matrix interference
- Rapid turnaround compatible with clinical research throughput
- Reliable quantification across the physiologically relevant range
It is readily adoptable in research laboratories investigating vitamin D status, disease associations, or supplementation efficacy.
Future Trends and Potential Applications
Emerging opportunities include:
- Automation of sample preparation to further increase throughput
- Expansion to multiplexed panels of vitamin D metabolites
- Integration with AI-driven data analysis for pattern recognition in large cohorts
- Application in pharmacokinetic and epidemiological studies
Conclusion
A precise, fast, and reproducible LC–HRMS method for 25-OH-vitamin D2 and D3 quantification was established using offline sample prep, online SPE, and Orbitrap detection. Both FullMS and PRM modes met stringent criteria for sensitivity, linearity, accuracy, and precision, supporting its use in clinical research settings.
Reference
De Nardi C., Kern K., Peters S. Quantification of 25-hydroxyvitamin D2 and D3 in human plasma by liquid chromatography coupled to high-resolution Orbitrap mass spectrometry for clinical research. Technical Note No. 73234, Thermo Fisher Scientific, 2019.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Quantification of 25-hydroxyvitamin D2 and D3 in human plasma by LC-HRAM-MS for clinical research
2021|Thermo Fisher Scientific|Applications
TECHNICAL NOTE 66025 Quantification of 25-hydroxyvitamin D2 and D3 in human plasma by LC-HRAM-MS for clinical research Author: Mariana Barcenas, Thermo Fisher Scientific, Les Ulis, France Keywords: Orbitrap Exploris 120 MS, Vanquish Flex UHPLC, TraceFinder, 25-hydroxyvitamin D, high-resolution mass spectrometry,…
Key words
average, averagecalculated, calculatedconcentration, concentrationlevel, levelbias, biasrecipe, recipecalibrators, calibratorscalibrator, calibratoranalyte, analyteassay, assaylloq, lloqscan, scancontrol, controlnominal, nominalsource
Quantification of ethyl glucuronide and ethyl sulfate in human urine by LC-HRAM(MS) for clinical research
2021|Thermo Fisher Scientific|Applications
TECHNICAL NOTE 66004 Quantification of ethyl glucuronide and ethyl sulfate in human urine by LC-HRAM(MS) for clinical research Authors: Claudio De Nardi1, Katharina Kern2, Sebastian Berger2, Thermo Fisher Scientific GmbH, Dreieich, Germany 1 RECIPE Chemicals + Instruments GmbH, Munich, Germany…
Key words
fullms, fullmsethyl, ethylglucuronide, glucuronidelevel, levelprm, prmcalculated, calculatedaverage, averageconcentration, concentrationetg, etgbias, biassulfate, sulfateets, etsurine, urineiii, iiimode
Quantification of 25-hydroxyvitamin D2 and D3 in human plasma by liquid chromatography-tandem mass spectrometry for clinical research
2019|Thermo Fisher Scientific|Applications
TECHNICAL NOTE 65471 Quantification of 25-hydroxyvitamin D2 and D3 in human plasma by liquid chromatography-tandem mass spectrometry for clinical research Authors Application benefits Claudio De Nardi, Thermo Fisher Scientific GmbH, Dreieich, Germany • Increased accuracy of method by implementation of…
Key words
calculated, calculatedaverage, averageassay, assaypercentage, percentageevaluated, evaluatedmethod, methodterms, termsbias, biasinter, internominal, nominalatmospheric, atmosphericconcentration, concentrationequilibration, equilibrationmin, mincalibrators
Confident quantitation of 25-hydroxyvitamin D2 and D3 in human plasma for clinical research by liquid chromatography-tandem mass spectrometry with simple, robust sample preparation
2019|Thermo Fisher Scientific|Applications
TECHNICAL NOTE 65559 Confident quantitation of 25-hydroxyvitamin D2 and D3 in human plasma for clinical research by liquid chromatography-tandem mass spectrometry with simple, robust sample preparation Authors Application benefits Claudio De Nardi, Thermo Fisher Scientific GmbH, Dreieich, Germany • Simple…
Key words
calculated, calculatedcalibrators, calibratorsaverage, averagepercentage, percentagemass, massbias, biasprecipitation, precipitationnominal, nominalevaluated, evaluatedquantis, quantisatmospheric, atmosphericconcentration, concentrationcontrols, controlscontrol, controlterms