Novel Methods for Detection and Quantitation of Impurities Using a New High Sensitivity Photodiode Array Detector
Posters | 2013 | ShimadzuInstrumentation
Monitoring trace impurities in chemical and pharmaceutical products is critical to ensure quality, efficacy and regulatory compliance. Ultrahigh performance liquid chromatography (UHPLC) requires highly sensitive detectors to identify and quantify low-level contaminants in complex matrices. The development of advanced detection and data-processing techniques addresses limitations of conventional photodiode array (PDA) systems and streamlines impurity analysis workflows.
This work introduces a novel high-sensitivity PDA detector (SPD-M30A) and two intelligent data processing methods (i-PDeA and i-DReC). The objectives are:
The analytical platform comprised a UHPLC system coupled with the SPD-M30A PDA detector featuring a low-dispersion capillary SR-Cell for sub-5×10−6 AU noise. Key elements:
High-sensitivity analyses of ethyl 4-hydroxybenzoate at 0.005% impurity level showed an S/N of ~9.5. i-PDeA successfully resolved co-eluted valerophenone (VP) and difluorobenzophenone (DFBP) peaks, achieving linear calibration (R2 >0.9999) over 1–200 mg/mL with ≤1.2% error. In mixed samples (DFBP/VP ratios 100:1 to 100:200), VP quantitation errors remained under 1%. i-DReC extended dynamic range in pharmaceutical samples (0.01–1 g/L) with maintained linearity beyond 0.5 g/L; reproducibility of main peak area was 0.06% RSD and impurity area ratios (0.005–2.97%) showed <1% RSD.
The combined hardware and software solutions provide:
Next steps may include:
The novel SPD-M30A PDA detector and the accompanying i-PDeA and i-DReC algorithms significantly improve the detection, deconvolution and quantitation of trace impurities. These innovations enhance data quality, streamline workflows and support stringent analytical requirements in pharmaceutical and industrial laboratories.
1) D. Nakayama et al., Shimadzu Application Note PO-CON1357E, First Edition June 2013
HPLC
IndustriesPharma & Biopharma
ManufacturerShimadzu
Summary
Importance of Topic
Monitoring trace impurities in chemical and pharmaceutical products is critical to ensure quality, efficacy and regulatory compliance. Ultrahigh performance liquid chromatography (UHPLC) requires highly sensitive detectors to identify and quantify low-level contaminants in complex matrices. The development of advanced detection and data-processing techniques addresses limitations of conventional photodiode array (PDA) systems and streamlines impurity analysis workflows.
Study Objectives and Overview
This work introduces a novel high-sensitivity PDA detector (SPD-M30A) and two intelligent data processing methods (i-PDeA and i-DReC). The objectives are:
- To evaluate the detector’s enhanced sensitivity and dynamic range.
- To apply i-PDeA for deconvolution of co-eluted peaks and purity assessment.
- To implement i-DReC for dynamic range extension in over-saturated signals.
- To demonstrate quantitative performance in standard mixtures and pharmaceutical samples.
Methodology and Used Instrumentation
The analytical platform comprised a UHPLC system coupled with the SPD-M30A PDA detector featuring a low-dispersion capillary SR-Cell for sub-5×10−6 AU noise. Key elements:
- UHPLC columns for rapid separation and high resolution.
- SPD-M30A detector enabling simultaneous major component and impurity analysis at concentration ratios down to 0.005%.
- i-PDeA (Intelligent Peak Deconvolution Analysis) to mathematically separate overlapping peaks using derivative spectra at optimized wavelengths.
- i-DReC (Intelligent Dynamic Range Calculation) to correct saturated signals by shifting to unsaturated wavelengths and applying sensitivity correction factors (k = Ia/Ib).
Main Results and Discussion
High-sensitivity analyses of ethyl 4-hydroxybenzoate at 0.005% impurity level showed an S/N of ~9.5. i-PDeA successfully resolved co-eluted valerophenone (VP) and difluorobenzophenone (DFBP) peaks, achieving linear calibration (R2 >0.9999) over 1–200 mg/mL with ≤1.2% error. In mixed samples (DFBP/VP ratios 100:1 to 100:200), VP quantitation errors remained under 1%. i-DReC extended dynamic range in pharmaceutical samples (0.01–1 g/L) with maintained linearity beyond 0.5 g/L; reproducibility of main peak area was 0.06% RSD and impurity area ratios (0.005–2.97%) showed <1% RSD.
Benefits and Practical Applications
The combined hardware and software solutions provide:
- Enhanced sensitivity for trace impurity detection.
- Automated peak purity evaluation and deconvolution without additional runs.
- Dynamic range extension that eliminates re-analysis or dilution steps.
- High throughput and robust quantitation in routine QA/QC and research environments.
Future Trends and Potential Applications
Next steps may include:
- Integration of machine learning algorithms for automated method selection and real-time anomaly detection.
- Expansion to multi-dimensional chromatographic techniques (LC×LC) coupled with advanced PDA arrays.
- Miniaturization for field deployable UHPLC-PDA systems in environmental and food safety monitoring.
- Broader application to biomolecule impurity profiling in biopharmaceutical development.
Conclusion
The novel SPD-M30A PDA detector and the accompanying i-PDeA and i-DReC algorithms significantly improve the detection, deconvolution and quantitation of trace impurities. These innovations enhance data quality, streamline workflows and support stringent analytical requirements in pharmaceutical and industrial laboratories.
Reference
1) D. Nakayama et al., Shimadzu Application Note PO-CON1357E, First Edition June 2013
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
New Data Processing Method for Photodiode Array Detector
2013|Shimadzu|Technical notes
C190-E167 Technical Report New Data Processing Method for Photodiode Array Detector Principle and Summary of i-DReC (Intelligent Dynamic Range Extension Calculator) Toshinobu Yanagisawa1 A b s tra c t: A new data processing method for a photo diode array (PDA)…
Key words
drec, dreccorrection, correctionwavelength, wavelengthdre, dreintensity, intensityspectrum, spectrumcorrected, correctedmau, maupeak, peakrange, rangemain, mainextension, extensionerror, errorarea, areadynamic
Pharmaceuticals - Technologies for Enhanced LC and SFC Workflow
2018|Shimadzu|Brochures and specifications
C10G-E069 Pharmaceuticals Technologies for Enhanced LC and SFC Workflow Introduction With the rapid growth of global population, the increased incidence and prevalence of diseases, and the continuous aging society in major economies, the global demand for pharmaceuticals can only increase.…
Key words
min, minreport, reportscouting, scoutingnexera, nexerapeak, peakmobile, mobiledata, datapdea, pdeadrec, drecsystem, systemspectrum, spectrumchiralpak, chiralpakmethod, methodtime, timemau
i-DReC (Intelligent Dynamic Range Extension Calculator), a New Photodiode Array Detector Function; Impurity Analysis Application
2013|Shimadzu|Technical notes
LAAN-A-LC-E229 Application News L451 No. High Performance Liquid Chromatography i-DReC (Intelligent Dynamic Range Extension Calculator), a New Photodiode Array Detector Function; Impurity Analysis Application Using the Intelligent Dynamic Range Extension Calculator (i-DReC), a new data analysis function for the photodiode…
Key words
drec, drecwavelength, wavelengthimpurity, impuritypeak, peakcorrection, correctionarea, areaintensity, intensityheight, heightcalculated, calculatedmau, mausensitivity, sensitivitychosen, chosencoefficient, coefficientspectrum, spectrumdynamic
Shimadzu Nexera series - Ultra High Performance Liquid Chromatograph
2020|Shimadzu|Brochures and specifications
C196-E096D Ultra High Performance Liquid Chromatograph Nexera series Mobile Phase Monitoring Auto-Diagnostics FlowPilot Maximizing Throughput Unattended Operation Integrated Lab Compact Design EXPERIENCE NEW BENCHMARKS A New Benchmark of Intelligence A New Benchmark of Efficiency A New Benchmark of Design Shimadzu…
Key words
pending, pendingpatent, patentbenchmark, benchmarkflowpilot, flowpilotnexera, nexerastatus, statususage, usagedrec, drecmau, maunew, newanalysis, analysispdea, pdeaintelligence, intelligencestandby, standbyimpurity