Peptide Mapping - Hardware and Column Optimization

Presentations | 2014 | Agilent TechnologiesInstrumentation
Consumables, HPLC, LC columns
Industries
Manufacturer
Agilent Technologies

Summary

Significance of the Topic


Peptide mapping by reversed-phase liquid chromatography is a cornerstone technique for characterizing protein therapeutics, biopharmaceuticals and complex biological samples. High-resolution separation and accurate gradient delivery are essential for reproducible identification of post-translational modifications, sequence variants and impurities. Optimizing hardware and minimizing extra-column dispersion are critical to achieving sharp peaks, consistent retention times and reliable quantitation in both research and quality-control laboratories.

Objectives and Study Overview


This work evaluates the performance of a modern quaternary UHPLC pump, mixing accessories and low-dispersion flow cells in peptide mapping applications. Key aims include reducing dwell and delay volumes, improving composition and flow accuracy, and demonstrating reproducible peptide separations with online modifier dilution. A companion software tool for automated chromatogram comparison is also introduced.

Methodology and Instrumentation


System dispersion and delay volume were quantified using narrow-bore columns and short, shallow gradients. Dwell volume profiles were measured across different pump configurations. Peak dispersion in tubing and flow cells was assessed by comparing standard and optimized plumbing. Performance metrics included retention-time precision, peak width at half height and separation efficiency. Reproducible peptide maps were generated for BSA digest under isocratic and gradient conditions with pre-mixed solvents versus the BlendAssist online dilution tool.

Main Results and Discussion


Use of the quaternary UHPLC pump with a low-volume mixer reduced delay volume to <350 µL, delivering composition precision below 0.15 % RSD and flow accuracy <0.07 % RSD. Extra-column volume dropped from 9.7 to 3.9 µL, sharpening isocratic alkylphenone peaks by over 50 %. The optofluidic Max-Light cartridge cell provided high sensitivity with a small internal volume, minimizing dispersion. BlendAssist enabled accurate online dilution of trifluoroacetic acid modifiers, yielding peptide maps with retention-time variation under 0.1 % across replicates. The OpenLAB CDS MatchCompare add-on automated alignment and area‐percent comparison of peptide chromatograms against references.

Benefits and Practical Applications

  • Enhanced peak resolution and reproducibility in peptide mapping workflows
  • Reduced analysis times by preserving shallow gradient fidelity
  • Cost-effective UHPLC performance using existing HPLC columns and methods
  • Simplified solvent preparation through online modifier dilution
  • Objective, software-driven chromatogram comparison for QC release

Future Trends and Opportunities


Continued miniaturization of flow paths and integration of ultralow-dispersion components will further improve sensitivity for low-abundance peptides. AI-driven method development and chromatogram matching may automate identification of sequence variants and PTMs. Emerging microfluidic and optofluidic detection schemes promise to lower sample requirements and enable real-time process monitoring in biomanufacturing.

Conclusion


Optimizing pump configuration, tubing dimensions and flow-cell design markedly enhances peptide mapping performance. The combination of low-delay-volume quaternary mixing, advanced dispersion control and software-assisted comparison delivers highly reproducible separations critical for biopharma R&D and QA/QC. These improvements accelerate method transfer, reduce operational costs and support robust characterization of complex protein therapeutics.

References

  • Dolan W. LCGC 2006;24(5):458–466.
  • Rehder D et al. J Chromatogr A. 2006;1102:164–175.
  • Bank R A et al. Peptide Mapping with Lys-C, Free Univ Amsterdam. 1988;238(1):105–108.
  • Wypych J et al. J Biol Chem. 2008;283:16194–16205.
  • Wang W et al. Anal Sci. Human Genome Sci Inc. deamidation study.
  • Palaniswamy M S. Agilent application note. Diphenyl column disulfide analysis.
  • Yoo C et al. Anal R&D Amgen Inc. Top-down PEGylation site mapping.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Eliminate  your Application and Chromatography Challenges - Optimizing  your Application/Method Transfer  considerations
Eliminate your Application and Chromatography Challenges LC Application Scientist Session Information Contributed by USA HPLC Applications Scientist Team Speaker: 1 June 22, 2020 Lorena Lopez LC Application Scientist Lexington, MA DE.7419791667 1260 Infinity II Prime LC 800 bar 1260 Infinity…
Key words
volume, volumense, nseiset, isetdelay, delaydispersion, dispersioncolumn, columnretention, retentiondamper, damperemulation, emulationminus, minusmixer, mixergradient, gradientmethod, methodtime, timeparticle
Agilent 1260 Infinity Binary LC Optimization Guide
Agilent 1260 Infinity Binary LC Optimization Guide Authors: Angelika Gratzfeld-Huesgen, Michael Frank, Christian Gotenfels Table of Contents Introduction 5 1. Theory of using smaller particles in liquid chromatography 6 1.1. Benefits of columns packed with sub-2-micron particles 12 Frictional heating…
Key words
application, applicationcolumn, columnresolution, resolutionrrht, rrhtrapid, rapiddelay, delaysheet, sheetdata, datavolume, volumeagilent, agilentcolumns, columnsdatasheet, datasheetzorbax, zorbaxfast, fastthreaded
Your Essential Resource for INFINITYLAB LC SUPPLIES
Your Essential Resource for INFINITYLAB LC SUPPLIES
2020|Agilent Technologies|Brochures and specifications
Your Essential Resource for INFINITYLAB LC SUPPLIES ACROSS THE LAB, AROUND THE WORLD, WITH YOU EVERY STEP OF THE WAY. Real stories from the lab. WHO WE ARE. WHAT WE DO. Whatever your lab needs, Agilent CrossLab is ready to…
Key words
valve, valvesst, sstsupplies, suppliespeek, peekdescription, descriptionfitting, fittingpreswaged, preswagedinfinitylab, infinitylabcapillaries, capillariesbio, biokit, kitkits, kitspump, pumpstainless, stainlesscolumn
Overcoming Barriers with UHPLC to Achieve Maximum Performance
Overcoming Barriers with UHPLC to Achieve Maximum Performance Anne Mack, William Long, Jason Link, Maureen Joseph Agilent Technologies, Inc., 2850 Centerville Road, Wilmington, DE 19808 Abstract Experimental UHPLC systems and columns improve laboratory productivity by enabling shorter analysis times and…
Key words
extra, extravolume, volumemau, maudelay, delaymin, mincolumn, columncollection, collectionuhplc, uhplcisocratic, isocraticrates, ratesimproperly, improperlygradient, gradientcan, cancapillary, capillarytcc
Other projects
GCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike