New To HPLC - Avoiding Beginner Pitfalls
Presentations | 2013 | Agilent TechnologiesInstrumentation
High-performance liquid chromatography (HPLC) is a cornerstone technique in analytical chemistry for separating, identifying and quantifying compounds in complex mixtures. New users often face challenges related to instrument setup, column selection, method conditions and sample handling that can compromise data quality, column life and overall productivity. Recognizing and avoiding these beginner pitfalls ensures reliable results, efficient use of resources and prolonged instrument and column performance.
This application note by Rita Steed provides a structured review of common errors encountered by novice HPLC operators. It covers four major areas: instrument hardware and connections, column characteristics and maintenance, method parameters (mobile phase, pH, temperature) and sample preparation and injection. Each section illustrates problems through chromatographic examples and proposes practical solutions for routine laboratory operation.
Critical instrument factors include:
The methodology emphasizes:
Key observations include:
By systematically addressing these pitfalls, laboratories can achieve:
Emerging developments will further streamline HPLC practice:
A thorough understanding of instrument setup, column selection, method parameters and sample handling is critical for effective HPLC operation. By recognizing and mitigating the described beginner pitfalls, analysts can ensure high data quality, method robustness and optimized resource use across research, QA/QC and industrial applications.
The original presentation and performance reports by Agilent Technologies (2013) were used to illustrate common issues and solutions in HPLC practice.
HPLC
IndustriesManufacturerAgilent Technologies
Summary
Importance of the Topic
High-performance liquid chromatography (HPLC) is a cornerstone technique in analytical chemistry for separating, identifying and quantifying compounds in complex mixtures. New users often face challenges related to instrument setup, column selection, method conditions and sample handling that can compromise data quality, column life and overall productivity. Recognizing and avoiding these beginner pitfalls ensures reliable results, efficient use of resources and prolonged instrument and column performance.
Objectives and Overview
This application note by Rita Steed provides a structured review of common errors encountered by novice HPLC operators. It covers four major areas: instrument hardware and connections, column characteristics and maintenance, method parameters (mobile phase, pH, temperature) and sample preparation and injection. Each section illustrates problems through chromatographic examples and proposes practical solutions for routine laboratory operation.
Instrumentation
Critical instrument factors include:
- Fittings and Tubing: Improperly swaged or mismatched ferrules and oversized tubing generate dead volume, peak tailing and loss of efficiency. Use minimal internal-diameter tubing (e.g. 0.12 mm ID for low-volume UHPLC) with pre-swaged ferrules.
- Detector Selection and Settings: Choose between UV/DAD, MS, RI, ELSD, FLD or ECD based on sample properties. Optimize flow cell volume and data rate to achieve ~20–30 data points per peak for accurate quantitation.
- Instrument Dwell Volume: Determine system dwell volume by gradient tracer experiments and adjust method delay times to ensure reproducible gradient formation.
Methodology and Instrumentation
The methodology emphasizes:
- Detector Flow Cells: Match cell volume to peak widths—micro, semi-micro and standard flow cells offer trade-offs between sensitivity and resolution.
- Column Performance Reports: Validate new columns on the actual instrument to establish baseline efficiency, selectivity and tailing factors under operating conditions.
- Extra-Column Volume: Quantify losses in efficiency with increasing tubing volume. Reducing extra-column volume by half can improve plate counts by up to 60 % for short UHPLC columns.
Main Results and Discussion
Key observations include:
- Poor Connections: Improperly swaged tubing introduces voids causing universal peak tailing. Replacing tubing restores symmetrical peaks.
- Particle Size Effects: Selectivity and efficiency vary with particle size (5 µm, 3.5 µm, 1.8 µm) and with mobile-phase proportioning versus premixing.
- Mobile Phase Preparation: Inconsistent mixing (v/v versus w/w) causes retention shifts and peak distortion. Consistent, documented buffer preparation is essential.
- pH and Temperature: Ionizable compounds show compound-specific retention changes with pH. Elevated temperature may improve peak shape but can degrade temperature-sensitive analytes or stationary phase.
- Sample and Injection Solvent: Unfiltered or strong-solvent injections cause split or broad peaks and column clogging. Use appropriate syringe filters and match injection solvent to initial mobile-phase strength.
Benefits and Practical Applications
By systematically addressing these pitfalls, laboratories can achieve:
- Improved peak shape, resolution and reproducibility.
- Extended column lifetime through proper pH, temperature control and column flushing protocols.
- Reduced downtime and maintenance costs by preventing blockages and degraded fittings.
- Robust method transfer between different HPLC/UHPLC systems.
Future Trends and Potential Applications
Emerging developments will further streamline HPLC practice:
- Advanced stationary phases with broader pH stability and specialized chemistries for challenging analytes.
- Micro- and nano-flow UHPLC systems requiring ultra-low dead volumes and precise fluidics.
- Automated method development software incorporating AI to predict optimal mobile phase composition, column selection and gradient profiles.
- Integration of on-line sample cleanup (SPE, microfluidics) to minimize sample-related column damage.
Conclusion
A thorough understanding of instrument setup, column selection, method parameters and sample handling is critical for effective HPLC operation. By recognizing and mitigating the described beginner pitfalls, analysts can ensure high data quality, method robustness and optimized resource use across research, QA/QC and industrial applications.
References
The original presentation and performance reports by Agilent Technologies (2013) were used to illustrate common issues and solutions in HPLC practice.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
There’s a lot to Learn: Tips for Avoiding Pitfalls
2015|Agilent Technologies|Presentations
There’s a lot to Learn: Tips for Avoiding Pitfalls Mark Powell LC Columns Application Engineer January 19, 2015 Overview Instrument • Connections • Performance Column • Characteristics • Lifetime Method • Mobile phase • Temperature Sample • Cleanup • Injection…
Key words
beginner, beginnerpitfalls, pitfallsavoiding, avoidingrestricted, restrictedagilent, agilentconfidentiality, confidentialitymobile, mobilecolumn, columnlabel, labelphase, phasemin, minfittings, fittingspitfall, pitfallfitting, fittingconnections
Avoiding Chromatography Heartbreak: Common LC Pitfalls
2017|Agilent Technologies|Presentations
Avoiding Chromatography Heartbreak: Common LC Pitfalls Rita Steed LC Columns Application Engineer February 28, 2017 Avoiding Overview Instrument • Connections • Performance Column • Characteristics • Lifetime Method • Mobile phase • Temperature Sample • Cleanup • Injection Avoiding Beginner…
Key words
pitfalls, pitfallsbeginner, beginneravoiding, avoidingmobile, mobilephase, phasemau, maucolumn, columnmin, minyour, yourfittings, fittingsavoid, avoidyou, youghost, ghostpitfall, pitfallbuffer
The Chromatography Detective: Troubleshooting Tips & Tools for LC & LCMS
2018|Agilent Technologies|Presentations
The Chromatography Detective: Troubleshooting Tips & Tools for LC & LCMS Rita Steed LC Columns Application Engineer January 31, 2018 What do you do when..... ➢ Your chromatography changes • Peak shape • Retention ➢ You can’t reproduce a method…
Key words
beginner, beginnerpitfalls, pitfallsavoiding, avoidingcolumn, columndwell, dwellvolume, volumephase, phasemobile, mobilepeak, peakecv, ecvinstrument, instrumenttailing, tailingshape, shaperetention, retentionconnections
Treat Yourself to Good Chromatography - Tips and tricks
2019|Agilent Technologies|Presentations
Treat Yourself to Good Chromatography Tips and tricks Rita Steed LC Application Engineer October 17, 2019 Column Related Column • Column Choices • New Column – Save Performance Report – Test mix (PN 01080-68704); 65/35 ACN/H2O or MeOH • Installation…
Key words
yourself, yourselftreat, treatrestricted, restrictedgood, goodagilent, agilentchromatography, chromatographycolumn, columnmin, minmobile, mobilephase, phasevolume, volumemau, mauyour, yourbuffers, buffersbuffer