Peak MS Table (All Models) - User Manual
Manuals | 2017 | Peak ScientificInstrumentation
The seamless integration of gas generation and instrument housing transforms mass spectrometry workflows by ensuring reliable, high-purity gas supply while reducing laboratory footprint, noise and maintenance burdens.
This user manual aims to guide end users through the installation, safe operation, maintenance and technical specifications of the MS Table series. It covers generator models with and without the integrated Infinity 1031 nitrogen generator, detailing the key performance parameters and compliance requirements.
The MS Table is tailored to accommodate Sciex triple quadrupole, QTRAP and IVD mass spectrometers. Premium versions integrate the Infinity 1031 compressor-less nitrogen generator, delivering curtain, source and exhaust gases. An optional wall-mounted Infinity 1031 kit is available for basic tables.
Installation requires a stable, level indoor environment with ambient temperatures between 5 °C and 30 °C, relative humidity below 70%. The 1N and 2N models connect to an oil-free compressed air source (ISO 8753-1:2010 Class 1.4.1). Electrical supply is 110-240 VAC, 50/60 Hz. Safety symbols, conformity declarations and environmental directives ensure legal and safe operation.
Pressure and flow outputs are factory-set:
By consolidating gas generation and instrument enclosure, the MS Table delivers sound-proofed, integrated vacuum pump housing, minimizing lab footprint and setup time. The system's modular design simplifies relocation and reduces dependencies on cylinder logistics, enhancing uptime and safety.
Advances in sensor integration, remote monitoring and predictive maintenance will further optimize generator performance. Future generations may adopt smart diagnostics, cloud connectivity and greener adsorbent materials, driving sustainability and automation in analytical laboratories.
The MS Table series offers a comprehensive solution for mass spectrometry gas management, combining robust performance, safety compliance and user-friendly maintenance. Its design philosophy underscores reliability, environmental responsibility and adaptability to evolving analytical workflows.
Laboratory gases and gas generators
IndustriesManufacturerPeak Scientific
Summary
Significance of the Topic
The seamless integration of gas generation and instrument housing transforms mass spectrometry workflows by ensuring reliable, high-purity gas supply while reducing laboratory footprint, noise and maintenance burdens.
Objectives and Overview
This user manual aims to guide end users through the installation, safe operation, maintenance and technical specifications of the MS Table series. It covers generator models with and without the integrated Infinity 1031 nitrogen generator, detailing the key performance parameters and compliance requirements.
Used Instrumentation
The MS Table is tailored to accommodate Sciex triple quadrupole, QTRAP and IVD mass spectrometers. Premium versions integrate the Infinity 1031 compressor-less nitrogen generator, delivering curtain, source and exhaust gases. An optional wall-mounted Infinity 1031 kit is available for basic tables.
Methodology and Instrumentation
Installation requires a stable, level indoor environment with ambient temperatures between 5 °C and 30 °C, relative humidity below 70%. The 1N and 2N models connect to an oil-free compressed air source (ISO 8753-1:2010 Class 1.4.1). Electrical supply is 110-240 VAC, 50/60 Hz. Safety symbols, conformity declarations and environmental directives ensure legal and safe operation.
Main Specifications and Discussion
Pressure and flow outputs are factory-set:
- Curtain gas: up to 4.5 bar, 19 L/min;
- Source gas: up to 7.6 bar, 26 L/min;
- Exhaust gas: up to 4.1 bar, 25 L/min.
Benefits and Practical Applications
By consolidating gas generation and instrument enclosure, the MS Table delivers sound-proofed, integrated vacuum pump housing, minimizing lab footprint and setup time. The system's modular design simplifies relocation and reduces dependencies on cylinder logistics, enhancing uptime and safety.
Future Trends and Applications
Advances in sensor integration, remote monitoring and predictive maintenance will further optimize generator performance. Future generations may adopt smart diagnostics, cloud connectivity and greener adsorbent materials, driving sustainability and automation in analytical laboratories.
Conclusion
The MS Table series offers a comprehensive solution for mass spectrometry gas management, combining robust performance, safety compliance and user-friendly maintenance. Its design philosophy underscores reliability, environmental responsibility and adaptability to evolving analytical workflows.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Genius 3030 & 3031 User Manual
2019|Peak Scientific|Manuals
Genius 3030 & 3031 User Manual Genius3030-3031 User Manual Rev 3 RSID 128 PN UM-0005 06/03/19 Contents Change History How to use this Manual Introduction Warranties and Liabilities Safety Notices Declaration of Conformity Environmental Declaration Technical Specification Unpacking Installation Generator…
Key words
generator, generatorservice, serviceindication, indicationoutlets, outletsmaximum, maximumpeak, peakduty, dutycaution, cautionwarning, warningprotected, protectedcompressor, compressorgas, gasyour, youryellow, yellowdeclaration
Genius 1050, 1051/61, 1052 & 1053 - User Manual
2019|Peak Scientific|Manuals
Genius 1050, 1051/61, 1052 & 1053 User Manual Genius1050-1051-1052-1053-1061 User Manual Rev 5 PN UM-0002 RSID 1484 26/09/19 Contents User Manual Change History How to use this Manual Introduction Warranties and Liabilities Safety Notices Symbols Safety Notice to Users Declaration…
Key words
generator, generatormaximum, maximumservice, serviceindication, indicationduty, dutystorage, storageoutlet, outletlbs, lbsoutlets, outletscaution, cautionpeak, peakpressure, pressuretemperature, temperatureambient, ambientled
Peak Genius 1024 - User Manual
2020|Peak Scientific|Manuals
Genius 1024 User Manual Genius1024 User Manual Rev 4 PN UM-0018 RSID 1917 09/03/20 Contents Change History How to use this Manual Introduction Warranties and Liabilities Safety Notices Symbols Safety Notice to Users Declaration of Conformity Environmental Declaration Technical Specification…
Key words
generator, generatorservice, servicecaution, cautionpeak, peakindication, indicationyour, yourwarning, warningprotected, protectedmaximum, maximumpressure, pressuredeclaration, declarationscientific, scientificair, airreset, resetshould
Genius 1022, 1023 & NM32LA (All Models) - User Manual
2019|Peak Scientific|Manuals
Genius 1022, 1023 & NM32LA (All Models) User Manual Copyright © Peak Scientific Instruments Ltd 2019 — Genius1022-1023-NM32LA User Manual Rev 11 PN UM-0001 RSID 1483 EN —21/10/2019 Contents User Manual 1 Change History 3 How to use this Manual…
Key words
generator, generatorservice, servicemaximum, maximumindication, indicationstorage, storageduty, dutyoutlets, outletslbs, lbspressure, pressurecaution, cautionpeak, peaktemperature, temperatureambient, ambientwarning, warningoutlet