High-Resolution Charge Variant Analysis for Top-Selling Monoclonal Antibody Therapeutics Using a Linear pH Gradient Separation Platform
Applications | 2016 | Thermo Fisher ScientificInstrumentation
Monoclonal antibody (mAb) therapeutics represent a major segment in biopharma. Characterizing charge variants is critical for ensuring safety and efficacy of both innovator products and biosimilars. Charge heterogeneity arises from post-translational modifications such as deamidation, sialylation, and C-terminal lysine clipping, which can affect biological activity. A robust, high-resolution charge variant analysis platform allows streamlined quality control and comparability studies.
This application note demonstrates development of a versatile charge variant separation method using a linear pH gradient on a cation-exchange column (MAbPac SCX-10). Four top-selling mAbs—rituximab, trastuzumab, adalimumab, and bevacizumab—were evaluated under both full and half pH gradient conditions and compared against traditional salt gradient approaches. Objectives include assessing resolution, method optimization ease, and general applicability across different mAbs.
A linear pH gradient from pH 5.6 to 10.2 was generated by mixing CX-1 pH Gradient Buffers A and B on a Thermo Scientific Dionex UltiMate 3000 BioRS LC system. Samples were injected onto a MAbPac SCX-10 column (4×250 mm, 10 μm) at 30 °C with 1 mL/min flow rate; detection was by UV at 280 nm. Both full gradient (0–100 % B) and half gradient (0–50 % B) profiles were applied, each run for 40 min.
The linear pH gradient method provided clear separation of acidic, major, and basic variants (peaks 1–3) for all four mAbs. Comparison against salt gradient separations showed significantly greater resolution (larger retention time differences between variants) for rituximab, trastuzumab, and bevacizumab. Adalimumab displayed similar performance under both approaches. Optimization was simplified by adjusting gradient slope in a linear system, enabling higher resolution runs with reduced pH range.
As biosimilar development accelerates, platform charge variant analysis tools will be in greater demand. Emerging trends include coupling pH gradients with mass spectrometry for direct variant identification and automation of gradient optimization. Expanding buffer sets to cover broader pI ranges will further streamline method transfer among diverse mAbs.
The linear pH gradient cation-exchange method on a MAbPac SCX-10 column offers a robust, high-resolution platform for mAb charge variant analysis. It surpasses traditional salt gradient approaches in separation power and method simplicity, making it ideal for biosimilar development and routine quality control.
HPLC
IndustriesPharma & Biopharma
ManufacturerThermo Fisher Scientific
Summary
Significance of the topic
Monoclonal antibody (mAb) therapeutics represent a major segment in biopharma. Characterizing charge variants is critical for ensuring safety and efficacy of both innovator products and biosimilars. Charge heterogeneity arises from post-translational modifications such as deamidation, sialylation, and C-terminal lysine clipping, which can affect biological activity. A robust, high-resolution charge variant analysis platform allows streamlined quality control and comparability studies.
Study Objectives and Overview
This application note demonstrates development of a versatile charge variant separation method using a linear pH gradient on a cation-exchange column (MAbPac SCX-10). Four top-selling mAbs—rituximab, trastuzumab, adalimumab, and bevacizumab—were evaluated under both full and half pH gradient conditions and compared against traditional salt gradient approaches. Objectives include assessing resolution, method optimization ease, and general applicability across different mAbs.
Methodology and Instrumentation
A linear pH gradient from pH 5.6 to 10.2 was generated by mixing CX-1 pH Gradient Buffers A and B on a Thermo Scientific Dionex UltiMate 3000 BioRS LC system. Samples were injected onto a MAbPac SCX-10 column (4×250 mm, 10 μm) at 30 °C with 1 mL/min flow rate; detection was by UV at 280 nm. Both full gradient (0–100 % B) and half gradient (0–50 % B) profiles were applied, each run for 40 min.
Instrumentation Used
- Thermo Scientific Dionex UltiMate 3000 BioRS system (HPG-3400RS pump, WPS-3000TBRS autosampler, TCC-3000RS column compartment, VWD-3400RS UV detector)
- MAbPac SCX-10 column, 4×250 mm, 10 μm
- CX-1 pH Gradient Buffer Kit (Buffers A and B)
- Chromeleon 6.8 chromatography data system
Main Results and Discussion
The linear pH gradient method provided clear separation of acidic, major, and basic variants (peaks 1–3) for all four mAbs. Comparison against salt gradient separations showed significantly greater resolution (larger retention time differences between variants) for rituximab, trastuzumab, and bevacizumab. Adalimumab displayed similar performance under both approaches. Optimization was simplified by adjusting gradient slope in a linear system, enabling higher resolution runs with reduced pH range.
Benefits and Practical Applications
- A single platform method accommodates multiple mAbs, reducing development time.
- Linear pH gradient yields higher resolution than salt gradients without extensive buffer screening.
- Improved detection of minor charge variants supports biosimilar comparability and quality control.
Future Trends and Potential Applications
As biosimilar development accelerates, platform charge variant analysis tools will be in greater demand. Emerging trends include coupling pH gradients with mass spectrometry for direct variant identification and automation of gradient optimization. Expanding buffer sets to cover broader pI ranges will further streamline method transfer among diverse mAbs.
Conclusion
The linear pH gradient cation-exchange method on a MAbPac SCX-10 column offers a robust, high-resolution platform for mAb charge variant analysis. It surpasses traditional salt gradient approaches in separation power and method simplicity, making it ideal for biosimilar development and routine quality control.
References
- Top 50 Pharmaceutical Products by Global Sales, PMLive, accessed 2016.
- Lin S, Baek J, Decrop W, Rao S, Agroskin Y, Pohl C. Development of a Cation-Exchange pH Gradient Separation Platform. 39th International Symposium on HPLC, 2013.
- Vlasak J, Ionescu R. Heterogeneity of Monoclonal Antibodies Revealed by Charge-Sensitive Methods. Curr Pharm Biotechnol. 2008;9(6):468–481.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Evaluation and application of salt- and pH-based ion-exchange chromatography gradients for analysis of therapeutic monoclonal antibodies
2018|Thermo Fisher Scientific|Applications
APPLICATION NOTE 21778 Evaluation and application of salt- and pH-based ion-exchange chromatography gradients for analysis of therapeutic monoclonal antibodies Authors Amy Farrell, Craig Jakes, and Jonathan Bones Characterization and Comparability Laboratory, NIBRT – The National Institute for Bioprocessing Research and…
Key words
absorbance, absorbancecharge, chargevariant, variantbuffers, buffersmau, maumab, mabgradient, gradientvariants, variantsmabs, mabstherapeutic, therapeuticiex, iexsalt, saltmin, mintime, timephosphate
Thermo Bio LC Columns - See your protein therapeutics in high resolution
2020|Thermo Fisher Scientific|Brochures and specifications
See your protein therapeutics in high resolution Bio LC Columns Contents 2 Introduction to monoclonal antibody (mAb) characterization 3 Affinity titer determination 4 Aggregate separation 5 Charge variant analysis 6 Oxidation variant analysis 11 Antibody drug conjugates (ADCs) analysis 12…
Key words
mabpac, mabpacmab, mabcolumn, columncolumns, columnshplc, hplcformat, formatsize, sizescfc, scfcmau, maumin, minparticle, particleantibody, antibodycharge, chargemmae, mmaevariant
Simple charge variant profile comparison of an innovator monoclonal antibody and a biosimilar candidate
2018|Thermo Fisher Scientific|Applications
APPLICATION NOTE 21777 Innovator Biosimilar Simple charge variant profile comparison of an innovator monoclonal antibody and a biosimilar candidate Authors Silvia Millán, Anne Trappe, Amy Farrell, and Jonathan Bones Characterization and Comparability Laboratory, NIBRT – The National Institute for Bioprocessing…
Key words
cetuximab, cetuximabbiosimilar, biosimilarcharge, chargecandidate, candidatevariant, variantvariants, variantsmau, mauabsorbance, absorbanceinnovator, innovatortime, timemonoclonal, monoclonalmin, minprotein, proteingradient, gradienttruncation
BioLC columns and accessories
2024|Thermo Fisher Scientific|Brochures and specifications
BioLC columns Connected chromatography solutions BioLC columns and accessories Introduction Your complete tool kit Thermo Fisher Scientific has innovative Thermo Scientific™ BioLC columns for each step of your therapeutic protein characterization, no matter how challenging your separation. Here is just…
Key words
biolc, biolcaccesories, accesoriesmobile, mobilephase, phasereading, readingcolumn, columngradient, gradientsize, sizecolumns, columnslinks, linksvariants, variantsparticle, particlemabpac, mabpaclength, lengthformat