Best Practices for Using an Agilent LC System - Technical Note
Technical notes | 2022 | Agilent TechnologiesInstrumentation
This technical note outlines essential maintenance and operational guidelines for an Agilent LC system to ensure reliable performance, consistent analytical results, and extended instrument life.
The note aims to consolidate best practices across all stages of liquid chromatography workflow on Agilent systems, including solvent handling, routine tasks, start-up and shutdown procedures, degasser and pump operation, sampler management, inline filter usage, column care, detector maintenance, and biocompatible system protocols.
Best practices are derived from systematic evaluation of daily and periodic maintenance tasks, cleaning procedures, operational settings, and preventative measures. Key instrumentation covered includes:
The guidelines highlight:
Implementing these practices leads to:
Advances in automation and software diagnostics will enable predictive maintenance and remote monitoring. Integration of AI-based advisors and cloud analytics promises further optimization of chromatographic workflows and instrument uptime.
Adopting these comprehensive best practices ensures robust, efficient, and reproducible performance of Agilent LC systems across diverse analytical applications.
HPLC
IndustriesManufacturerAgilent Technologies
Summary
Significance of the topic
This technical note outlines essential maintenance and operational guidelines for an Agilent LC system to ensure reliable performance, consistent analytical results, and extended instrument life.
Objectives and study overview
The note aims to consolidate best practices across all stages of liquid chromatography workflow on Agilent systems, including solvent handling, routine tasks, start-up and shutdown procedures, degasser and pump operation, sampler management, inline filter usage, column care, detector maintenance, and biocompatible system protocols.
Methodology and instrumentation
Best practices are derived from systematic evaluation of daily and periodic maintenance tasks, cleaning procedures, operational settings, and preventative measures. Key instrumentation covered includes:
- Agilent LC pumps (1260, 1290 Infinity, 1290 Infinity II series)
- Degassers G1322A, G7122A, and G4225A
- Multisamplers (including Multiwash option)
- PDA/DAD and RI detectors
- Inline filter assemblies and biocompatible flow path components
Main results and discussion
The guidelines highlight:
- Strict solvent management: use HPLC-grade reagents, daily exchanges, proper labeling, brown bottles for aqueous phases
- Sample preparation: match diluent to mobile phase, remove particulates via 0.2 μm filtration or centrifugation
- Routine cleaning: sequential flushing with water, NaOH, HCl to prevent biofilm and salt deposits
- Seal wash protocols: mandatory use of 10 % isopropanol (or pure for normal phase), periodic operation, fresh bottles
- Multiwash use for carryover control in reversed-phase mode; daily water flush for salt applications
- Inline filter benefits: 0.3 μm pore size, minimal dead volume, high-pressure compatibility to protect columns
Benefits and practical applications
Implementing these practices leads to:
- Enhanced method reproducibility and baseline stability
- Reduced downtime and maintenance costs
- Extended lifetime of columns and consumables
- Improved sample integrity and data quality in QA/QC and research settings
Future trends and possibilities
Advances in automation and software diagnostics will enable predictive maintenance and remote monitoring. Integration of AI-based advisors and cloud analytics promises further optimization of chromatographic workflows and instrument uptime.
Conclusion
Adopting these comprehensive best practices ensures robust, efficient, and reproducible performance of Agilent LC systems across diverse analytical applications.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Best Practices for Using an Agilent LC System - Technical Note
2016|Agilent Technologies|Technical notes
Best Practices for Using an Agilent LC System Technical Note Best Practices - Technical Note This technical note describes best practices for using an Agilent LC system. Contents Housekeeping 2 How to deal with solvents 2 How to prepare samples…
Key words
seal, sealwash, washrecommendations, recommendationsmandatory, mandatorysolvent, solventweekly, weeklyusage, usagepumps, pumpsinstalled, installeddegassers, degassersshut, shutwhen, whenpump, pumpsamplers, samplerspower
Reduce PFAS Background with the Agilent PFC-Free HPLC Conversion Kit
2020|Agilent Technologies|Technical notes
Technical Overview Reduce PFAS Background with the Agilent PFC-Free HPLC Conversion Kit An ideal solution for trace level PFAS analysis with LC/MS/MS Authors Matthias Kamuf, Marcus Wälz, and Andreas Borowiak Agilent Technologies, Inc. Introduction Per- and polyfluoroalkyl substances (PFAS) are…
Key words
pfc, pfcdelay, delaypfas, pfasacid, acidkit, kitbackground, backgroundfree, freetubing, tubingmultiwash, multiwashsetup, setupcolumn, columnfep, fepperistaltic, peristalticwash, washconversion
Your Essential Resource for INFINITYLAB LC SUPPLIES
2020|Agilent Technologies|Brochures and specifications
Your Essential Resource for INFINITYLAB LC SUPPLIES ACROSS THE LAB, AROUND THE WORLD, WITH YOU EVERY STEP OF THE WAY. Real stories from the lab. WHO WE ARE. WHAT WE DO. Whatever your lab needs, Agilent CrossLab is ready to…
Key words
valve, valvesst, sstsupplies, suppliespeek, peekdescription, descriptionfitting, fittingpreswaged, preswagedinfinitylab, infinitylabcapillaries, capillariesbio, biokit, kitkits, kitspump, pumpstainless, stainlessseal
Care and Feeding of your HPLC System: Preventative Maintenance Tips
2024|Agilent Technologies|Presentations
Care and Feeding of your HPLC System: Preventative Maintenance Tips Mark Powell Applications Engineer Columns and Supplies Technical Support Instrument Supplies • Critical points of instrument maintenance • Supplies to have on hand Not keeping up with regular preventive maintenance…
Key words
feeding, feedingcare, carehplc, hplcyour, yourpump, pumpdad, dadvwd, vwdsystem, systemfilter, filtercompartment, compartmentfitting, fittingvalve, valvebottle, bottleneedle, needlequick