Quantitation of 33 benzodiazepines by LCMS- 8030 from human serum using RECIPE ClinMass LC MS/MS Complete Kit MS6000

Applications | 2013 | ShimadzuInstrumentation
Consumables, LC/MS, LC/MS/MS, LC/QQQ
Industries
Clinical Research
Manufacturer
Shimadzu, RECIPE

Summary

Significance of the Topic


The accurate quantitation of benzodiazepines in human serum is critical for therapeutic drug monitoring, forensic investigations and clinical toxicology. Benzodiazepines are widely prescribed central nervous system (CNS) depressants with a narrow therapeutic index and potential for dependence. Rapid, sensitive and reliable assays enable clinicians and laboratories to optimize dosing regimens, detect abuse and ensure patient safety.

Goals and Overview of the Study


This work describes the evaluation of a pre-optimized commercial kit (RECIPE ClinMass® LC MS/MS Complete Kit MS6000) for the simultaneous quantitation of 33 benzodiazepines and metabolites in human serum. The primary goals were to assess method sensitivity, linearity and reproducibility when applied on a Shimadzu LCMS-8030 triple quadrupole mass spectrometer coupled to a Nexera UHPLC system with on-line solid phase extraction (SPE).

Methodology and Instrumentation Used


Sample preparation is automated via an on-line SPE column integrated with the UHPLC using a six-port switching valve. Analytical separation is achieved within 10 minutes using a gradient elution provided by the RECIPE kit mobile phases. Positive APCI mode and multiple reaction monitoring (MRM) are employed on the LCMS-8030. Key instrumental parameters include:
  • UHPLC System: Shimadzu Nexera
  • LCMS: Shimadzu LCMS-8030; APCI positive, interface 425 °C, DL 200 °C, heat block 300 °C
  • Flow Rates: Analytical pumps A & B 0.7 mL/min; loading pump C varied for SPE at 5 mL/min
  • Injection Volume: 20 µL; Column temperature: 30 °C
  • MRM Transitions: Precursor ions [M+H]+ and two product ions per analyte, dwell time 20 ms

Main Results and Discussion


All 33 compounds were baseline separated within 10 minutes following on-line SPE extraction. Calibration curves demonstrated excellent linearity over clinically relevant concentration ranges (R2 ≥ 0.991; typical R2 > 0.998). Diazepam calibration showed R2 = 0.9985 with a lower limit of quantitation around 0.1 mg/L. Accuracy and precision were evaluated using low and high concentration controls in duplicate; inter-day and intra-day relative standard deviations were consistently below 5%.

The method covers parent drugs (e.g., diazepam, lorazepam, alprazolam) and key metabolites (e.g., nordiazepam, 7-aminoclonazepam). On-line SPE reduced manual sample handling and improved throughput while maintaining sensitivity and selectivity.

Benefits and Practical Applications


The pre-configured kit accelerates method development by supplying calibrated standards, deuterated internal standards and optimized mobile phases. Laboratories gain:
  • Rapid implementation without extensive local optimization
  • High throughput with a 10-minute cycle per sample
  • Robust quantitation across a broad panel of benzodiazepines
  • Low carry-over and reduced risk of sample contamination

Future Trends and Opportunities


Emerging trends include miniaturized SPE cartridges to further decrease solvent usage and integration with high-resolution mass spectrometry for non-targeted screening of novel psychoactive substances. Advances in automated sample preparation and data processing algorithms will enable even greater throughput and real-time decision support in clinical settings. Expanding kits to include other CNS agents could streamline multi-drug monitoring platforms.

Conclusion


The combination of the RECIPE ClinMass® LC MS/MS kit with Shimadzu LCMS-8030 and Nexera UHPLC offers a fast, reliable and reproducible method for comprehensive benzodiazepine quantitation in human serum. The workflow meets clinical and forensic requirements for sensitivity, linearity and precision while minimizing development time.

Reference


Shimadzu Corporation; Recipe Chemicals + Instruments GmbH. Quantitation of 33 benzodiazepines by LCMS-8030 from human serum using RECIPE ClinMass® LC MS/MS Complete Kit MS6000, 2013.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Offline and online sample extraction for the quantification of benzodiazepines in human plasma or serum for clinical research
TECHNICAL NOTE 64913 Valium (diazepam) structure Offline and online sample extraction for the quantification of benzodiazepines in human plasma or serum for clinical research Authors Claudio De Nardi,1 Katharina Kern,2 Steffen Peters2 Application benefits Thermo Fisher Scientific GmbH, Dreieich, Germany…
Key words
midazolam, midazolamtriazolam, triazolamalprazolam, alprazolamdesmethylflunitrazepam, desmethylflunitrazepambias, biastrazodon, trazodonbromazepam, bromazepamnorclobazam, norclobazamdiazepam, diazepamtetrazepam, tetrazepamzolpidem, zolpidemnordiazepam, nordiazepamdemoxepam, demoxepamflunitrazepam, flunitrazepamaverage
Quantification of 35 benzodiazepines in human plasma by LC-HRAM-MS for clinical research
TECHNICAL NOTE 000306 Quantification of 35 benzodiazepines in human plasma by LC-HRAM-MS for clinical research Authors: Gaëtan Renoulin1, Claudio De Nardi2, Mariana Barcenas1 1 Thermo Fisher Scientific, Les Ulis, France 2 Thermo Fisher Scientific, Reinach, Switzerland Keywords: Orbitrap Exploris 120…
Key words
level, levelintensity, intensityalpha, alphacalculated, calculatedaverage, averagemin, minzaleplone, zaleplonenorclobazam, norclobazamtetrazepam, tetrazepamassay, assaydesmethylflunitrazepam, desmethylflunitrazepamdemoxepam, demoxepamhydroxymidazolam, hydroxymidazolammedazepam, medazepamratio
Quantification of 34 benzodiazepines in human plasma by LC-HRAM-MS for clinical research
TECHNICAL NOTE 66051 Quantification of 34 benzodiazepines in human plasma by LC-HRAM-MS for clinical research Authors: Gaëtan Renoulin1, Mariana Barcenas1, Claudio De Nardi2 Thermo Fisher Scientific, Les Ulis, France 1 Thermo Fisher Scientific, Reinach, Switzerland 2 Keywords: Orbitrap Exploris 120…
Key words
level, levelalpha, alphazaleplone, zaleplonenorclobazam, norclobazamtetrazepam, tetrazepamdesmethylflunitrazepam, desmethylflunitrazepamdemoxepam, demoxepammedazepam, medazepamhydroxymidazolam, hydroxymidazolamaverage, averagelormetazepam, lormetazepamprazepam, prazepamhydroxyalprazolam, hydroxyalprazolamhydroxytriazolam, hydroxytriazolamclobazam
Quantitation of Benzodiazepines in Serum by UHPLC-Triple Quadrupole Mass Spectrometry
Quantitation of Benzodiazepines in Serum by UHPLC-Triple Quadrupole Mass Spectrometry This study demonstrates a simple, rapid, and reliable method for the simultaneous detection and quantitation of 35 benzodiazepines in serum using the Bruker Elute UHPLC coupled to the EVOQ LC-TQ…
Key words
benzodiazepines, benzodiazepinesevoq, evoqclinmass, clinmasstdm, tdmbruker, brukerelite, eliteclonazepam, clonazepamserum, serumnorclobazam, norclobazamuhplc, uhplcnordazepam, nordazepamtetrazepam, tetrazepamdesmethylflunitrazepam, desmethylflunitrazepamdemoxepam, demoxepammedazepam
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike