Quick and Routine Research Quantification of Melatonin in Plasma with the Agilent Ultivo LC/TQ
Applications | 2019 | Agilent TechnologiesInstrumentation
Melatonin is a key hormone for regulation of circadian rhythm and sleep. Low level quantification in human plasma supports research in pharmacokinetics and endogenous production. Sensitive methods are required to measure sub ppt concentrations reliably in complex biological matrices.
The study demonstrates a rapid and sensitive method for quantification of melatonin in plasma using an Agilent 1260 Infinity II Prime liquid chromatograph coupled with Ultivo triple quadrupole mass spectrometer with electrospray ionization. The target limit of quantification is 10 pg per mL using a simple protein precipitation extraction.
The procedure uses protein precipitation of plasma samples with acetonitrile followed by centrifugation and vacuum drying. Extracts are reconstituted in methanol and subjected to LC MS analysis. A stacked injection approach with a 3 microL injection volume is applied. Calibration standards of deuterated melatonin are prepared in plasma extract over a range from 10 pg per mL to 100 ng per mL.
The optimized method achieves a limit of detection of 5 pg per mL in extract corresponding to 10 pg per mL in plasma and a lower limit of quantification of 10 pg per mL. The calibration curve is linear over four orders of magnitude with correlation coefficient of 0.998. Method precision is better than 10 percent relative standard deviation and average accuracy ranges from 95 to 105 percent. Extraction recovery of deuterated melatonin is 95 percent with one standard deviation of 6 percent. Measurement of endogenous melatonin in pooled plasma yields an average concentration of 40.8 pg per mL with 7.3 percent RSD.
The Agilent Ultivo LC TQ with ESI source provides an economical fit for purpose solution for trace level quantification of melatonin in plasma. The method offers high sensitivity precision and reproducibility with minimal sample handling and is well suited for routine research applications.
None
LC/MS, LC/MS/MS, LC/QQQ
IndustriesClinical Research
ManufacturerAgilent Technologies
Summary
Importance of the topic
Melatonin is a key hormone for regulation of circadian rhythm and sleep. Low level quantification in human plasma supports research in pharmacokinetics and endogenous production. Sensitive methods are required to measure sub ppt concentrations reliably in complex biological matrices.
Goals and overview
The study demonstrates a rapid and sensitive method for quantification of melatonin in plasma using an Agilent 1260 Infinity II Prime liquid chromatograph coupled with Ultivo triple quadrupole mass spectrometer with electrospray ionization. The target limit of quantification is 10 pg per mL using a simple protein precipitation extraction.
Methodology
The procedure uses protein precipitation of plasma samples with acetonitrile followed by centrifugation and vacuum drying. Extracts are reconstituted in methanol and subjected to LC MS analysis. A stacked injection approach with a 3 microL injection volume is applied. Calibration standards of deuterated melatonin are prepared in plasma extract over a range from 10 pg per mL to 100 ng per mL.
Used instrumentation
- Agilent 1260 Infinity II Prime flexible pump
- Agilent 1260 Infinity II multisampler with cooler
- Agilent 1260 Infinity II multicolumn thermostat
- Agilent Ultivo triple quadrupole LC MS system
- Agilent electrospray ionization source
Main results and discussion
The optimized method achieves a limit of detection of 5 pg per mL in extract corresponding to 10 pg per mL in plasma and a lower limit of quantification of 10 pg per mL. The calibration curve is linear over four orders of magnitude with correlation coefficient of 0.998. Method precision is better than 10 percent relative standard deviation and average accuracy ranges from 95 to 105 percent. Extraction recovery of deuterated melatonin is 95 percent with one standard deviation of 6 percent. Measurement of endogenous melatonin in pooled plasma yields an average concentration of 40.8 pg per mL with 7.3 percent RSD.
Benefits and practical applications
- Fast sample preparation by simple protein precipitation
- High sensitivity suitable for low ppt level detection
- Robust linear response across a wide concentration range
- Fit for routine research use in clinical and pharmacokinetic studies
Future trends and possibilities
- Integration with high throughput automation
- Expansion to multiplex hormone panels
- Further miniaturization and increased sensitivity
- Applications in chronobiology and personalized dosing strategies
Conclusion
The Agilent Ultivo LC TQ with ESI source provides an economical fit for purpose solution for trace level quantification of melatonin in plasma. The method offers high sensitivity precision and reproducibility with minimal sample handling and is well suited for routine research applications.
Reference
None
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Catch Them Sleeping: Quick and Routine Quantification of Melatonin in Plasma with Ultivo LC/TQ
2019|Agilent Technologies|Posters
Poster Reprint ASMS 2019 MP013 Catch Them Sleeping: Quick and Routine Quantification of Melatonin in Plasma with Ultivo LC/TQ Mark J. Sartain and Aaron Boice Agilent Technologies, Inc., Santa Clara, CA, USA Introduction Melatonin (MEL) is an endogenous hormone and…
Key words
mel, melmelatonin, melatoninultivo, ultivomrm, mrmendogenous, endogenousesi, esiquantification, quantificationradioimmunoassays, radioimmunoassaysplasma, plasmafragmentor, fragmentorcircadian, circadiananalytical, analyticalsource, sourcemasshunter, masshunterneurotransmitter
Agilent LC Triple Quadrupole Clinical Research Compendium
2021|Agilent Technologies|Guides
Agilent LC Triple Quadrupole Clinical Research Compendium Application Compendium Table of contents Contents Introduction 4 Routine analysis 5 Hormones 5 Progesterone Metabolism in Serum 5 Analytical Determination of Testosterone in Human Serum Using an Agilent Ultivo Triple Quadrupole LC/MS 7…
Key words
atturos, atturoscounts, countsplasma, plasmaagilent, agilentstreamselect, streamselectserum, serummel, meltriple, triplequadrupole, quadrupolecaptiva, captivathc, thcusing, usingtestosterone, testosteronemethod, methodresponses
Rapid and Sensitive Quantitation of Metanephrines and 3-Methoxytyramine in Human Plasma Using an Agilent Ultivo Triple Quadrupole LC/MS
2022|Agilent Technologies|Applications
Application Note Clinical Research Rapid and Sensitive Quantitation of Metanephrines and 3-Methoxytyramine in Human Plasma Using an Agilent Ultivo Triple Quadrupole LC/MS Abstract Authors Suet Ying Lee , Amaury Cazenave‑Gassiot1,2, Leroy S. Pakkiri3,4, Chester L. Drum3,4,5, Markus R. Wenk1,2, and…
Key words
counts, countschromsystems, chromsystemsplasma, plasmarecovery, recoveryspe, spethree, threemqc, mqclqc, lqchqc, hqcyong, yongnmn, nmnloo, looacquisition, acquisitioncalculated, calculatedworkflow
Analysis of Multiclass Steroids in Serum Using Agilent Ultivo Triple Quadrupole LC/MS
2022|Agilent Technologies|Applications
Application Note Clinical Research Analysis of Multiclass Steroids in Serum Using Agilent Ultivo Triple Quadrupole LC/MS Automation-ready SLE sample preparation using Agilent Chem Elut S, 96-well plates Abstract Authors Cherie Chin , Xue Ping Goh , Suet Ying Lee1,2, Amaury…
Key words
counts, countsdheas, dheassle, sleacquisition, acquisitiondht, dhtdhea, dheaaldosterone, aldosteronecorticosterone, corticosteroneandrostenedione, androstenedionecortisol, cortisolsteroids, steroidscortisone, cortisonemin, mintarget, targetprogesterone