Rapid Analysis of Immuno-suppressants Using Triple Quadrupole LC-MS/MS

Applications | 2014 | ShimadzuInstrumentation
LC/MS, LC/MS/MS, LC/QQQ
Industries
Clinical Research
Manufacturer
Shimadzu

Summary

Importance of the Topic


Therapeutic drug monitoring of immunosuppressants is critical in transplant medicine to ensure patient safety and graft survival. Fast and sensitive analytical methods enable prompt dose adjustments, reduce toxicity risks, and support high-throughput clinical workflows.

Objectives and Study Overview


The study aimed to develop and validate an ultra-fast LC-MS/MS method for simultaneous quantification of four key immunosuppressants (tacrolimus, everolimus, rapamycin, cyclosporin A) alongside two internal standards (ascomycin, cyclosporin D) within a single 1.8-minute run.

Methodology and Instrumentation


Chromatography was performed on a Shimadzu Nexera X2 UHPLC system using a YMC-Triart C18 column (30 mm × 2 mm I.D., 1.9 µm). A binary gradient of 1 mmol/L ammonium acetate in water (mobile phase A) and methanol (mobile phase B) ramped from 60 % to 95 % B in under a minute. Electrospray ionization in negative mode on the Shimadzu LCMS-8050 triple quadrupole provided targeted multiple reaction monitoring (MRM). Key MS parameters included a drying gas flow of 5 L/min, nebulizer gas at 3 L/min, interface temperature at 400 ℃, and collision-induced dissociation to product ions ranging from m/z 321.20 to 1102.60.

Main Results and Discussion


The method achieved limits of quantitation down to 1 ng/mL for all analytes with only 1.5 µL injection volumes. Calibration was linear across therapeutic ranges (tacrolimus 0.5–500 ng/mL; everolimus 0.2–200 ng/mL; rapamycin 0.5–500 ng/mL; cyclosporin A 0.2–1000 ng/mL). Intra-run precision (%RSD) at the lower limit of quantitation ranged from 7.55 % to 19.2 %, and accuracy remained within 85–115 %. Chromatographic separation maintained peak resolution despite the sub-2-minute cycle, demonstrating robustness for routine analysis.

Benefits and Practical Applications


  • High throughput: sub-2-minute analysis enables processing of dozens of samples per hour.
  • Clinical relevance: meets sensitivity and precision requirements for therapeutic drug monitoring.
  • Resource efficiency: low solvent consumption and minimal sample volume.

Future Trends and Potential Applications


Further integration with automated sample preparation could streamline workflows in clinical laboratories. Expanding this approach to additional narrow-therapeutic-index drugs and incorporating high-resolution MS would enhance multiplexing capabilities and confirmatory power.

Conclusion


The developed triple quadrupole LC-MS/MS method delivers rapid, sensitive, and reliable quantification of critical immunosuppressants, supporting high-throughput clinical monitoring while maintaining regulatory standards for precision and accuracy.

References


  • Shimadzu Corporation. Rapid Analysis of Immuno-suppressants Using Triple Quadrupole LC-MS/MS. First Edition, June 2014.
  • Department of Pharmacy, Kyoto University Hospital. Acknowledgements.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Accelerated and robust monitoring for immunosuppressants using triple quadrupole mass spectrometry
PO-CON1449E Accelerated and robust monitoring for immunosuppressants using triple quadrupole mass spectrometry ASMS 2014 WP628 Natsuyo Asano1, Tairo Ogura1, Kiyomi Arakawa1 1 Shimadzu Corporation. 1, Nishinokyo Kuwabara-cho, Nakagyo-ku, Kyoto 604–8511, Japan Accelerated and robust monitoring for immunosuppressants using triple quadrupole…
Key words
immunosuppressants, immunosuppressantscyclosporin, cyclosporinascomycin, ascomycinneg, negrapamycin, rapamycinaccelerated, acceleratedeverolimus, everolimustacrolimus, tacrolimustriple, triplerobust, robustmonitoring, monitoringquadrupole, quadrupolespectrometry, spectrometrymass, massdeprotonated
Simultaneous Analysis of Tacrolimus, Sirolimus, Everolimus, and Cyclosporin A in Whole Blood Using the Agilent RapidFire High-Throughput Mass Spectrometry System
Simultaneous Analysis of Tacrolimus, Sirolimus, Everolimus, and Cyclosporin A in Whole Blood Using the Agilent RapidFire High-Throughput Mass Spectrometry System Application Note Clinical Research Authors Introduction Kari E. Schlicht and Vaughn P. Miller Agilent Technologies, Inc. Wakefield, MA USA In…
Key words
auc, auccyclosporin, cyclosporinrapidfire, rapidfireeverolimus, everolimussirolimus, sirolimustacrolimus, tacrolimusquantifier, quantifierqualifier, qualifierprecision, precisionaccuracy, accuracywhole, wholequant, quantmid, midwere, wereblood
Evaluation of Benchtop Quadrupole Orbitrap Ultra-High-Resolution Mass Spectrometer in Rapid Quantitative Analysis of Immunosuppressant Drugs in Blood Samples
Methods Calibrators Calibrators containing cyclosporin A, everolimus, sirolimus, and tacrolimus in whole blood were purchased from Chromsystems Instruments & Chemicals GmbH (TABLE 1). Evaluation of Benchtop Quadrupole Orbitrap Ultra-High-Resolution Mass Spectrometer in Rapid Quantitative Analysis of Immunosuppressant Evaluation of Benchtop…
Key words
cyclosporin, cyclosporintacrolimus, tacrolimuseverolimus, everolimussirolimus, sirolimusquantifier, quantifierrelative, relativeimmunosuppressant, immunosuppressantanalyte, analyteabundance, abundanceblood, bloodascomycin, ascomycinquadrupole, quadrupolepositive, positiveorbitrap, orbitrapclinical
Simultaneous Analysis of Tacrolimus, Sirolimus, Everolimus, and Cyclosporine A in Whole Blood Using an Agilent Ultivo Triple Quadrupole LC/MS
Application Note Clinical Research Simultaneous Analysis of Tacrolimus, Sirolimus, Everolimus, and Cyclosporine A in Whole Blood Using an Agilent Ultivo Triple Quadrupole LC/MS Authors Gökhan Günay and Gökçe Göksu Gürsu SEM Laboratuar Cihazları Paz. San. Ve Tic. A.Ş., R&D Center,…
Key words
cyclosporine, cyclosporinejasem, jasemeverolimus, everolimustacrolimus, tacrolimussirolimus, sirolimusimmunosuppressants, immunosuppressantsrelative, relativepositive, positivewhole, wholeblood, blooddiverter, diverterultivo, ultivoresponses, responsesmin, mintime
Other projects
GCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike