A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer
Applications | 2017 | Agilent TechnologiesInstrumentation
Accurate and reliable quantitation of immunosuppressant drugs in whole blood is critical for therapeutic drug monitoring in transplant patients and for quality control in clinical laboratories. High-throughput and robust LC-MS/MS methods reduce analysis time, increase sample throughput, and minimize instrument downtime, supporting large-scale clinical studies and routine patient management.
This study evaluated the long-term robustness of the Agilent 6470 Triple Quadrupole LC-MS/MS system for the quantitation of four immunosuppressants—cyclosporin A, everolimus, sirolimus, and tacrolimus—using a 2-minute high-throughput method. The performance was assessed over 14 days and more than 14,000 injections under stress conditions, alternating between calibration and test batches.
Whole blood samples were prepared by protein precipitation with a zinc sulfate–methanol reagent containing isotopically labeled internal standards. An automated online cleanup employed a trapping column for matrix removal, followed by elution onto an analytical column. Calibration curves were constructed using 11 concentration levels (0.1 ng/mL to 2,000 ng/mL) in triplicate, applying 1/x weighting.
The LC system comprised an Agilent 1260 Infinity Binary Pump, Thermostatted Column Compartment, and Autosampler with inline filter. A trapping column (ZORBAX Eclipse Plus C18) and an analytical column (Poroshell 120 EC-C18) were operated at 60 °C with a 2-minute gradient. The MS was an Agilent 6470 Triple Quadrupole with JetStream source, operating in positive ion mode with optimized drying gas, sheath gas, voltages, and MRM transitions.
Peak area variation remained below 10% RSD for cyclosporin A, everolimus, sirolimus, and tacrolimus through 13 batches (14,495 injections), exceeding the threshold only at batch 14. Calibration linearity was excellent with R2>0.997 before and >0.994 after robustness testing. Column performance showed no significant degradation over 8,500 injections, and internal standard correction maintained quantitation accuracy throughout.
Integration of advanced automation and artificial intelligence for predictive maintenance, expansion to multiplexed assays for broader panels of analytes, and further reduction of cycle times could enhance throughput and data quality. Coupling with digital laboratory information systems and cloud-based data analysis will support real-time decision making and remote monitoring.
The Agilent 6470 LC-MS/MS system demonstrated exceptional robustness and reproducibility for high-throughput quantitation of key immunosuppressants in whole blood. The method delivers reliable performance over thousands of injections, making it suitable for routine clinical and research environments.
Rapid Analysis of Cyclosporine A, Everolimus, Sirolimus, and Tacrolimus Drugs in Whole Blood Using an Agilent Triple Quadrupole LC/MS/MS System with Automated Online Sample Cleanup. Agilent Technologies Application Note, publication number 5991-3344EN.
LC/MS, LC/MS/MS, LC/QQQ
IndustriesClinical Research
ManufacturerAgilent Technologies
Summary
Importance of the topic
Accurate and reliable quantitation of immunosuppressant drugs in whole blood is critical for therapeutic drug monitoring in transplant patients and for quality control in clinical laboratories. High-throughput and robust LC-MS/MS methods reduce analysis time, increase sample throughput, and minimize instrument downtime, supporting large-scale clinical studies and routine patient management.
Objectives and overview of the study
This study evaluated the long-term robustness of the Agilent 6470 Triple Quadrupole LC-MS/MS system for the quantitation of four immunosuppressants—cyclosporin A, everolimus, sirolimus, and tacrolimus—using a 2-minute high-throughput method. The performance was assessed over 14 days and more than 14,000 injections under stress conditions, alternating between calibration and test batches.
Methodology
Whole blood samples were prepared by protein precipitation with a zinc sulfate–methanol reagent containing isotopically labeled internal standards. An automated online cleanup employed a trapping column for matrix removal, followed by elution onto an analytical column. Calibration curves were constructed using 11 concentration levels (0.1 ng/mL to 2,000 ng/mL) in triplicate, applying 1/x weighting.
Instrumentation
The LC system comprised an Agilent 1260 Infinity Binary Pump, Thermostatted Column Compartment, and Autosampler with inline filter. A trapping column (ZORBAX Eclipse Plus C18) and an analytical column (Poroshell 120 EC-C18) were operated at 60 °C with a 2-minute gradient. The MS was an Agilent 6470 Triple Quadrupole with JetStream source, operating in positive ion mode with optimized drying gas, sheath gas, voltages, and MRM transitions.
Key results and discussion
Peak area variation remained below 10% RSD for cyclosporin A, everolimus, sirolimus, and tacrolimus through 13 batches (14,495 injections), exceeding the threshold only at batch 14. Calibration linearity was excellent with R2>0.997 before and >0.994 after robustness testing. Column performance showed no significant degradation over 8,500 injections, and internal standard correction maintained quantitation accuracy throughout.
Benefits and practical applications of the method
- High-throughput 2-minute cycle time supports large sample loads.
- Automated online cleanup reduces matrix effects and instrument maintenance.
- Low peak area variability ensures reliable long-term quantitation.
- Applicable to clinical research, therapeutic drug monitoring, and QA/QC workflows.
Future trends and possibilities
Integration of advanced automation and artificial intelligence for predictive maintenance, expansion to multiplexed assays for broader panels of analytes, and further reduction of cycle times could enhance throughput and data quality. Coupling with digital laboratory information systems and cloud-based data analysis will support real-time decision making and remote monitoring.
Conclusion
The Agilent 6470 LC-MS/MS system demonstrated exceptional robustness and reproducibility for high-throughput quantitation of key immunosuppressants in whole blood. The method delivers reliable performance over thousands of injections, making it suitable for routine clinical and research environments.
Reference
Rapid Analysis of Cyclosporine A, Everolimus, Sirolimus, and Tacrolimus Drugs in Whole Blood Using an Agilent Triple Quadrupole LC/MS/MS System with Automated Online Sample Cleanup. Agilent Technologies Application Note, publication number 5991-3344EN.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Rapid Analysis of Cyclosporine A, Everolimus, Sirolimus, and Tacrolimus Drugs in Whole Blood Using an Agilent Triple Quadrupole LC/MS/MS System with Automated Online Sample Cleanup
2015|Agilent Technologies|Applications
Rapid Analysis of Cyclosporine A, Everolimus, Sirolimus, and Tacrolimus Drugs in Whole Blood Using an Agilent Triple Quadrupole LC/MS/MS System with Automated Online Sample Cleanup Application Note Clinical Research Authors Abstract Linda Côté, Siji Joseph, A highly sensitive and specific…
Key words
relative, relativetacrolimus, tacrolimusresponses, responsescyclosporin, cyclosporineverolimus, everolimussirolimus, sirolimusvariation, variationwere, weretac, tacconcentration, concentrationwhole, wholeblood, bloodanalytes, analytesthreshold, thresholdcsa
Simultaneous Analysis of Tacrolimus, Sirolimus, Everolimus, and Cyclosporin A in Whole Blood Using the Agilent RapidFire High-Throughput Mass Spectrometry System
2014|Agilent Technologies|Applications
Simultaneous Analysis of Tacrolimus, Sirolimus, Everolimus, and Cyclosporin A in Whole Blood Using the Agilent RapidFire High-Throughput Mass Spectrometry System Application Note Clinical Research Authors Introduction Kari E. Schlicht and Vaughn P. Miller Agilent Technologies, Inc. Wakefield, MA USA In…
Key words
auc, auccyclosporin, cyclosporinrapidfire, rapidfireeverolimus, everolimussirolimus, sirolimustacrolimus, tacrolimusqualifier, qualifierquantifier, quantifierprecision, precisionaccuracy, accuracywhole, wholequant, quantmid, midwere, wereblood
Simultaneous Analysis of Tacrolimus, Sirolimus, Everolimus, and Cyclosporine A in Whole Blood Using an Agilent Ultivo Triple Quadrupole LC/MS
2020|Agilent Technologies|Applications
Application Note Clinical Research Simultaneous Analysis of Tacrolimus, Sirolimus, Everolimus, and Cyclosporine A in Whole Blood Using an Agilent Ultivo Triple Quadrupole LC/MS Authors Gökhan Günay and Gökçe Göksu Gürsu SEM Laboratuar Cihazları Paz. San. Ve Tic. A.Ş., R&D Center,…
Key words
cyclosporine, cyclosporinejasem, jasemeverolimus, everolimustacrolimus, tacrolimussirolimus, sirolimusimmunosuppressants, immunosuppressantsrelative, relativepositive, positivewhole, wholeblood, blooddiverter, diverterultivo, ultivoresponses, responsesmin, mintime
Evaluation of Benchtop Quadrupole Orbitrap Ultra-High-Resolution Mass Spectrometer in Rapid Quantitative Analysis of Immunosuppressant Drugs in Blood Samples
2016|Thermo Fisher Scientific|Posters
Methods Calibrators Calibrators containing cyclosporin A, everolimus, sirolimus, and tacrolimus in whole blood were purchased from Chromsystems Instruments & Chemicals GmbH (TABLE 1). Evaluation of Benchtop Quadrupole Orbitrap Ultra-High-Resolution Mass Spectrometer in Rapid Quantitative Analysis of Immunosuppressant Evaluation of Benchtop…
Key words
cyclosporin, cyclosporintacrolimus, tacrolimuseverolimus, everolimussirolimus, sirolimusquantifier, quantifierimmunosuppressant, immunosuppressantrelative, relativeanalyte, analyteabundance, abundanceblood, bloodascomycin, ascomycinquadrupole, quadrupolepositive, positiveclinical, clinicalorbitrap