Accurate Mass and Spectral Accuracy of Agilent LC/MS Single Quadrupole Systems with Cerno MassWorks

Technical notes | 2017 | Agilent TechnologiesInstrumentation
Software, LC/MS, LC/SQ
Industries
Manufacturer
Agilent Technologies

Summary

Importance of the Topic


Single quadrupole LC/MS systems are widely used but traditionally limited by moderate mass accuracy, restricting formula identification. Improving mass and spectral accuracy enhances their utility for unknown compound identification in routine labs.

Study Objectives and Overview


This application evaluates Agilent 6120B LC/MS combined with Cerno MassWorks software for enhanced mass accuracy and spectral accuracy, enabling confident elemental composition determination of unknowns using reserpine as a benchmark.

Methodology and Equipment


  • Calibration sample: Agilent standard tune solution (G2421-60001) covering m/z 117–2722.
  • Instrumentation: Agilent 6120B Single Quadrupole LC/MS in raw profile mode; scan range 100–1000 Da; zero ion threshold; Full Scan Data Storage.
  • Test compound: Reserpine (C33H40N2O9, MW 608.68) at 100 ppm; 1 µg injected, 10 replicates.
  • Calibration and analysis: MassWorks software for advanced calibration and lineshape modeling; formula search using CLIPS algorithm within 10 mDa tolerance.

Key Results and Discussion


  • Mass accuracy improved to ±5 mDa (0.3 mDa average error, 0.4 ppm), ~100× better than uncalibrated performance.
  • Spectral accuracy reached ~99.0%, enabling distinct formula ranking (>99.1% for correct match vs. next best <98.7%).
  • Repeatability over 10 injections showed standard deviation of 1.7 mDa and 0.15% spectral accuracy.
  • Calibration stability persists for days, reducing need for frequent recalibration.

Benefits and Practical Applications


This approach transforms standard single quadrupole LC/MS into a cost-effective tool for formula identification of pure compounds, bridging the performance gap with high-resolution instruments and streamlining workflows in QA/QC and research labs.

Future Trends and Opportunities


  • Integration of advanced calibration software to other low-resolution platforms.
  • Automation of spectral accuracy algorithms for real-time formula ID.
  • Expansion to complex mixtures and LC/MS screening applications.

Conclusion


Combining Agilent LC/MS single quadrupole with MassWorks yields high mass and spectral accuracy, enabling reliable formula identification with routine instrumentation, offering time and cost savings compared to specialized high-resolution systems.

Instrument Used


  • Agilent 6120B Single Quadrupole LC/MS
  • Agilent standard tune solution G2421-60001
  • Cerno MassWorks calibration and analysis software

References


  1. Wang Y., Gu M. The Concept of Spectral Accuracy for MS. Anal. Chem. 2010, 82, 7055–7062.
  2. Cerno Bioscience. Support notes for instrument parameter setup in MassWorks.
  3. Mullis J., Qiu F., Wang Y. The Robustness of Formula Determination on a Single Quadrupole GC/MS. ASMS 2008.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Rapid analysis of additives in food packaging
Rapid analysis of additives in food packaging
2018|Agilent Technologies|Applications
Application Note Rapid analysis of additives in food packaging Using an Agilent 6120 Single Quadrupole MS System with an atmospheric pressure solid analysis probe (ASAP) and accurate mass identification by MassWorks Author Hirokazu Sawada, PhD. Agilent Techhnologies, Inc. Abstract This…
Key words
massworks, massworksasap, asapelemental, elementalmass, massionization, ionizationrapid, rapiddirect, directatmospheric, atmosphericanalysis, analysiswrapper, wrapperaccurate, accurateformula, formulafood, foodquadrupole, quadrupolecalibration
Analysis of Additives in a Polymer
Analysis of Additives in a Polymer
2018|Agilent Technologies|Applications
Application Note Analysis of Additives in a Polymer Using the Agilent 6120B Single Quadrupole MS and MassWorks accurate mass elemental composition determination software Author Hirokazu Sawada, PhD. Agilent Technologies, Inc. Abstract The Agilent 6120B Single Quadrupole LC/MS System was used…
Key words
massworks, massworksmass, massformula, formulaelemental, elementalunknown, unknownaccurate, accuratecomposition, compositionquadrupole, quadrupolespectrum, spectrumsingle, singleadditives, additivessubstance, substancecalibration, calibrationpolymer, polymerprofile
Evaluation of Mass Accuracy Robustness for the Agilent InfinityLab Pro iQ Plus
Application Note Toxicology Evaluation of Mass Accuracy Robustness for the Agilent InfinityLab Pro iQ Plus Author Dallas Freitas Agilent Technologies, Inc. Abstract This application note demonstrates the robustness and mass accuracy of the Agilent InfinityLab Pro iQ Plus single quadrupole…
Key words
sim, simmassworks, massworksinjections, injectionsmean, meanpositive, positivecerno, cernobioscience, biosciencemass, massnumber, numbercarisoprodol, carisoprodolphencyclidine, phencyclidinetrazodone, trazodonebenzoylecgonine, benzoylecgoninehydrocodone, hydrocodoneplus
Accurate Mass Spectral Deconvolution of Multiply-charged Oligonucleotides using Unit Resolution Single Quadrupole LC/MS
Poster Reprint ASMS 2024 Poster number WP 480 Accurate Mass Spectral Deconvolution of Multiply-charged Oligonucleotides using Unit Resolution Single Quadrupole LC/MS Patrick M Batoon1; Lee Bertram2; Yongdong Wang3 1Agilent Technologies Inc., Santa Clara, CA; 2Agilent Technologies, Santa Clara, CA; 3Cerno…
Key words
cerno, cernomassworks, massworksmass, masstttttttttt, ttttttttttsammi, sammiaccurate, accuratespectral, spectralweight, weightmolecular, moleculartruecal, truecalassignment, assignmentspectrally, spectrallynominal, nominalexternally, externallycalculations
Other projects
GCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike