Impact of USP <1058> - Regulatory Spotlight on Analytical Instrument Qualification (AIQ)
Technical notes | 2017 | Agilent TechnologiesInstrumentation
The USP general chapter <1058> on Analytical Instrument Qualification (AIQ) defines risk-based strategies and performance requirements to ensure that analytical instruments deliver reliable, accurate data. The 2017 revision reinforces laboratory audit readiness, strengthens data integrity demands, and aligns qualification practices with global good manufacturing practice guidelines.
This application note examines the key changes introduced in the 2017 update of USP <1058>, evaluates their impact on laboratory qualification processes, and illustrates how risk assessment and modern software tools can support full compliance.
While USP <1058> does not list specific instrument examples, typical analytical platforms affected include:
To facilitate AIQ execution, Agilent’s Automated Compliance Engine (ACE) and Network Distributed ACE (NDA) software provide configurable qualification protocols aligned with USP <1058> and Annex 15 requirements.
The updated chapter mandates a structured qualification lifecycle:
Standard operating procedures (SOPs) must be revised to incorporate these elements and to establish traceable, controlled workflows. Manual calculations and unvalidated spreadsheets represent audit risks and should be replaced with validated software solutions.
The analysis highlights several critical implications of the 2017 revision:
The increased audit visibility of AIQ deficiencies has led to more FDA 483 observations for missing PQ tests, uncalibrated tools, and incomplete qualification reports.
Implementing the revised USP <1058> framework delivers multiple benefits:
Looking ahead, laboratories are likely to adopt fully integrated qualification solutions that leverage cloud-based compliance engines, real-time monitoring, and advanced risk-management dashboards. Greater harmonization across global pharmacopoeias and tighter integration with laboratory information management systems (LIMS) will further streamline qualification lifecycles.
The 2017 update to USP <1058> elevates the importance of risk-based qualification, URS-driven testing, and electronic compliance tools. Proactive adoption of structured SOPs and validated software platforms ensures audit readiness, strengthens data integrity, and supports consistent instrument performance.
GC, GC/MSD, HPLC, LC/MS, SFC
IndustriesPharma & Biopharma
ManufacturerAgilent Technologies
Summary
Importance of the Topic
The USP general chapter <1058> on Analytical Instrument Qualification (AIQ) defines risk-based strategies and performance requirements to ensure that analytical instruments deliver reliable, accurate data. The 2017 revision reinforces laboratory audit readiness, strengthens data integrity demands, and aligns qualification practices with global good manufacturing practice guidelines.
Objectives and Study Overview
This application note examines the key changes introduced in the 2017 update of USP <1058>, evaluates their impact on laboratory qualification processes, and illustrates how risk assessment and modern software tools can support full compliance.
Used Instrumentation
While USP <1058> does not list specific instrument examples, typical analytical platforms affected include:
- High-performance liquid chromatography (HPLC) and HPLC-MS
- Gas chromatography (GC) and GC-MS
- Supercritical fluid chromatography (SFC)
- Capillary electrophoresis (CE)
- Inductively coupled plasma techniques (ICP-OES, ICP-MS)
- Dissolution testers; infrared and ultraviolet spectroscopy (FT-IR, UV)
- Atomic absorption (AA) and laboratory software systems
To facilitate AIQ execution, Agilent’s Automated Compliance Engine (ACE) and Network Distributed ACE (NDA) software provide configurable qualification protocols aligned with USP <1058> and Annex 15 requirements.
Methodology and Instrument Qualification Approach
The updated chapter mandates a structured qualification lifecycle:
- Risk Assessment: Classify instruments into Group A, B or C based on criticality.
- User Requirements Specification (URS) and Design Qualification (DQ): Define intended use and performance targets.
- Operational Qualification (OQ): Validate instrument operation across the URS/DQ range.
- Performance Qualification (PQ): Demonstrate fitness for purpose under routine analytical conditions.
Standard operating procedures (SOPs) must be revised to incorporate these elements and to establish traceable, controlled workflows. Manual calculations and unvalidated spreadsheets represent audit risks and should be replaced with validated software solutions.
Main Results and Discussion
The analysis highlights several critical implications of the 2017 revision:
- Explicit requirement for URS and risk-based classification increases upfront planning.
- Clarification that both OQ and PQ are mandatory reduces interpretative gaps.
- Enhanced software guidance promotes alignment with GAMP and simplifies documentation.
- Stronger regulatory focus on data integrity drives harmonization of qualification across multiple instrument types.
The increased audit visibility of AIQ deficiencies has led to more FDA 483 observations for missing PQ tests, uncalibrated tools, and incomplete qualification reports.
Contributions and Practical Applications
Implementing the revised USP <1058> framework delivers multiple benefits:
- Reduced audit findings by aligning qualification protocols with regulatory expectations.
- Improved data integrity through validated workflows and electronic record-keeping.
- Scalable qualification across diverse analytical platforms using configurable software.
- Enhanced consistency in instrument performance and reporting.
Future Trends and Possibilities for Use
Looking ahead, laboratories are likely to adopt fully integrated qualification solutions that leverage cloud-based compliance engines, real-time monitoring, and advanced risk-management dashboards. Greater harmonization across global pharmacopoeias and tighter integration with laboratory information management systems (LIMS) will further streamline qualification lifecycles.
Conclusion
The 2017 update to USP <1058> elevates the importance of risk-based qualification, URS-driven testing, and electronic compliance tools. Proactive adoption of structured SOPs and validated software platforms ensures audit readiness, strengthens data integrity, and supports consistent instrument performance.
References
- USP <1058>, USP 40-NF 35, first supplement, 2017.
- Agilent Automated Compliance Engine (ACE) EQP Software Overview.
- Agilent Network Distributed ACE (NDA) Compliance Flyer, 2017.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Regulatory 101. What is AIQ? Complete DQ/IQ
2015|Agilent Technologies|Presentations
Regulatory 101 What is AIQ ? Complete DQ/IQ Paul Smith Global Strategic Compliance Program Manager Agilent Technologies, Inc. July 9, 2015 Confidentiality Label July 10, 2015 1 10 Steps to a “New Instrument”…. User Requirement Specification URS 1. • •…
Key words
confidentiality, confidentialitylabel, labelqualification, qualificationinstrument, instrumenturs, urswhat, whatapproval, approvalgamp, gampcompliance, complianceaiq, aiqqualify, qualifygpg, gpgrisk, riskcompanies, companiescycle
Agilent Equipment Qualification Solutions
2020|Agilent Technologies|Brochures and specifications
Agilent Equipment Qualification Solutions Qualify the Instruments in Your Laboratory with a Harmonized, and Data Integrity Compliant Solution Instrument Qualification – A Regulatory Requirement For your analytical results to be valid, your analytical instruments must be suitable for their intended…
Key words
qualification, qualificationprotocols, protocolstranscribed, transcribedagilent, agilentelectronic, electroniccalculations, calculationspaper, paperintegrity, integrityservices, servicesexcel, excelaudit, auditdata, dataharmonized, harmonizedreports, reportsace
Analytical Instrument Qualification with Data Integrity in Mind
2022|Waters|Technical notes
[ ARTICLE ] Analytical Instrument Qualification with Data Integrity in Mind Heather Longden, Senior Marketing Manager, Pharmaceutical Regulatory Compliance Waters Corporation, Milford, MA, USA It is critical to understand the data integrity implications regarding analytical instrument qualification. OVERVIEW Analytical Instrument…
Key words
qualification, qualificationaiq, aiqintegrity, integritydata, datainstrument, instrumentmind, mindarticle, articlerecords, recordssystem, systemanalytical, analyticalusp, uspdocumentation, documentationrequirements, requirementsguidance, guidanceintended
Analytical Instrument Qualification with Data Integrity in Mind
2022|Waters|Technical notes
[ ARTICLE ] Analytical Instrument Qualification with Data Integrity in Mind Heather Longden, Senior Marketing Manager, Pharmaceutical Regulatory Compliance Waters Corporation, Milford, MA, USA It is critical to understand the data integrity implications regarding analytical instrument qualification. OVERVIEW Analytical Instrument…
Key words