Comparison of the 800 bar and 600 bar Agilent 1260 Infinity II Vialsampler

Technical notes | 2017 | Agilent TechnologiesInstrumentation
HPLC
Industries
Manufacturer
Agilent Technologies

Summary

Significance of the Topic


Liquid chromatography systems with high pressure tolerance and reliable method transfer are essential in regulated environments to ensure consistent analytical performance and minimize the burden of revalidation.

Objectives and Study Overview


This overview compares the Agilent 1260 Infinity II Vialsampler models with 600 bar and 800 bar pressure limits regarding reproducibility, resolution, linearity, sensitivity, and carryover using four corticosteroids as analytes. The goal is to demonstrate backward compatibility and facilitate seamless method transfer.

Methodology and Instrumentation


Instrumentation used
  • Agilent 1260 Infinity II Flexible Pump (G7104C)
  • Agilent 1260 Infinity II Vialsampler 600 bar and 800 bar (G7129A/C) with integrated column compartment and sample cooler
  • Agilent 1260 Infinity II Diode Array Detector WR (G7115A)
Chromatographic conditions
  • Column: InfinityLab Poroshell 120 EC-C18, 3.0×100 mm, 2.7 μm
  • Mobile phase: water and acetonitrile gradient from 30 to 95% B over 7 min, 0.8 mL/min, 30 °C
  • Injection: 5 μL, needle wash in 50% acetonitrile
  • Detection wavelength: 254/10 nm (ref 360/100 nm)
Carryover assessment used an Agilent Pursuit XRs 3 C18 column with TFA modifiers and chlorhexidine spiking. Data were processed with Agilent OpenLAB CDS 2.2.

Main Results and Discussion


Both vialsamplers produced equivalent performance: retention time RSD ≤0.03%, area RSD ≤0.15%, resolution values >3 for all critical pairs, linearity coefficients (R2) >0.99996, and matched LOD/LOQ values. Carryover testing showed no detectable residue after blank injections.

Benefits and Practical Applications


The 800 bar vialsampler offers direct method compatibility with existing 600 bar protocols, reducing revalidation effort. High reproducibility and sensitivity support quantitative analysis of corticosteroids in quality control and research settings.

Future Trends and Opportunities


Advancements in high-pressure LC will enable faster separations and improved resolution for complex samples. Integration with novel detectors, miniaturization, and greener solvents will enhance throughput and sustainability. AI-driven data analysis may further streamline method development and compliance.

Conclusion


The 800 bar Agilent 1260 Infinity II Vialsampler demonstrates full backward compatibility with the 600 bar model, delivering equivalent chromatographic performance and enabling straightforward method transfer without additional revalidation.

References


1. U.S. FDA, Guidance for Industry: Analytical Procedures and Methods Validation for Drugs and Biologics, July 2015.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Performance Characteristics of the Agilent 1260 Infinity II LC System – an Entry System for Routine Analysis
Technical Overview Performance Characteristics of the Agilent 1260 Infinity II LC System – an Entry System for Routine Analysis Authors Elena Bühn, Beate Stahl, and Edgar Naegele Agilent Technologies, Inc. Introduction The Agilent 1260 Infinity II LC System is the…
Key words
rsd, rsdphthalate, phthalatecaffeine, caffeinearea, areamau, maugradient, gradientinjection, injectionparameter, parameterparaben, parabenvalue, valuesolvent, solventsample, samplepropiophenone, propiophenoneheptanophenone, heptanophenoneoctanophenone
Quantitative Screening of Multiresidue Veterinary Drugs in Seafood Using the Agilent 6470 Triple Quadrupole LC/MS
Application Note Food Testing and Agriculture Quantitative Screening of Multiresidue Veterinary Drugs in Seafood Using the Agilent 6470 Triple Quadrupole LC/MS Authors Siji Joseph, Aimei Zou, Limian Zhao, Patrick Batoon, and Chee Sian Gan Agilent Technologies, Inc. Abstract This application…
Key words
antibiotic, antibioticanthelmintic, anthelminticsulfonamides, sulfonamidesbenzimidazoles, benzimidazolescoccidiostats, coccidiostatslactam, lactaminsecticide, insecticidensaids, nsaidsbeta, betaveterinary, veterinarymatrix, matrixpromoters, promotersmqc, mqcmacrolides, macrolidesquinolones
Performance Characteristics of the Agilent 1290 Infinity II Vialsampler
Performance Characteristics of the Agilent 1290 Infinity II Vialsampler Technical Overview Author Abstract Melanie Metzlaff This Technical Overview describes the features and performance of the Agilent Technologies, Inc. Agilent 1290 Infinity II Vialsampler. With a power range up to 1,300…
Key words
rsd, rsdinjection, injectionchlorhexidine, chlorhexidinearea, areaoctanophenone, octanophenoneheptanophenone, heptanophenonepropiophenone, propiophenonebutyrophenone, butyrophenonehexanophenone, hexanophenoneacetanilide, acetanilidevalerophenone, valerophenonemau, mauacetophenone, acetophenonevialsampler, vialsamplerphthalate
Performance Characteristics of the Agilent 1260 Infinity II Online Sample Manager
Technical Overview Performance Characteristics of the Agilent 1260 Infinity II Online Sample Manager Authors Haiko Herschbach, Daniel Kutscher, Edgar Naegele, Thomas Ortmann, Irina Spuling, and Sam Wouters Agilent Technologies, Inc. Introduction In the pharmaceutical and biopharmaceutical industry, it is of…
Key words
feed, feedreactor, reactorinjection, injectionchlorhexidine, chlorhexidinearea, areaspeed, speeddraw, drawonline, onlinevial, vialfrom, fromflow, flowmanager, managermode, modemau, mauvolume
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike