Author
Thermo Fisher Scientific
Thermo Fisher Scientific
A unique company dedicated to serving science helping our customers drive productivity and accelerate results. Through our premier brands – Thermo Scientific, Applied Biosystems, Invitrogen, Fisher Scientific and Unity Lab Services.
Tags
Article
Academy
Product
LinkedIn Logo

Collision-Based Ion-activation and Dissociation

Tu, 23.9.2025
| Original article from: Thermo Fisher Scientific / Amanda Lee
Collision-based ion activation, developed nearly 80 years ago, remains the most widely used fragmentation method. Learn about three key techniques on Thermo Scientific MS.
<p><strong>Thermo Fisher Scientific: </strong>Collision-Based Ion-activation and Dissociation</p>

Thermo Fisher Scientific: Collision-Based Ion-activation and Dissociation

Introduced nearly 80 years ago, collision-based ion-activation and dissociation is the most commonly used type of fragmentation commercially available today. In principle, this technique is quite straightforward: apply energy to accelerate the particles of interest until they collide with an inert gas to induce fragmentation. Within these parameters, there are many different sub-categories of the technique. Here we will discuss three collision-based techniques that occur on Thermo Scientific mass spectrometers.

Thermo Fisher Scientific Freedom to Fragment IntroductionThermo Fisher Scientific Freedom to Fragment Introduction

Higher-Energy Collisional Dissociation (HCD)

  1. Process: HCD is a beam-based technique that occurs in the HCD cell or ion routing multipole, where the ion beam is directed into a potential well. Excited molecules collide with collision gas, causing fragmentation. This gas is typically nitrogen and is fast and efficient, but indiscriminate in technique. To ensure a higher energy collision, the potential energy is dropped (typically from the quadrupole into the collision cell) by many 10s of eV. This ensures the ions have high energy as they enter the cell with nitrogen and collide with enough energy for the fragmentation to occur. In peptides and proteins, HCD favors b- and y-ion formation. [Figure 1]
  2. Applications: HCD is commonly used for general proteomics and metabolomics.
  3. Considerations: When using Thermo Scientific mass spectrometers for HCD fragmentation, we generally use normalized collisional energy (NCE) instead of directly putting in the eV of potential well “drop”. The calibration routine determines the optimal voltage for a series of ions in the calibration solution.  Then the routine scales these absolute voltages to a “normalized collision energy” value of 30. As such, an NCE value of 30 should work well across a range of m/z values and charge states. To see the actual applied value, one can check the scan header of the spectrum easily in our Thermo Scientific Freestyle software.

Thermo Fisher Scientific: Collision-Based Ion-activation and Dissociation: Figure 1: Location of peptide/protein backbone cleavage and its nomenclature. Representative here is a 4-amino acid peptide with side chains labeled R1-4.Thermo Fisher Scientific: Collision-Based Ion-activation and Dissociation: Figure 1: Location of peptide/protein backbone cleavage and its nomenclature. Representative here is a 4-amino acid peptide with side chains labeled R1-4.

Collision-Induced Dissociation (CID) (ion trap specific version)

  1. Process: CID, as it occurs in Thermo Fisher Scientific’s dual linear ion trap technology, is a resonance-based technique where molecules are excited by a specific frequency.  These excited molecules will start running into He molecules at a faster rate.  These collisions slowly heat the resonant molecules until they fragment. Once fragmentation occurs, the product ions at new m/z values are no longer excited by the frequency and will cool back down to the center of the trap. In peptides and proteins, CID favors b- and y-ion formation.
  2. Applications: CID is great for small molecules and peptides and helps fragment oligonucleotides, especially when the q-value is adjusted.  It also fragments with higher efficiency, which makes it well-suited for MS3 and higher workflows.
  3. Considerations: Because CID is resonance-based, there is a compromise outlined in the Matthieu equation. With certain voltages and frequencies applied, only certain m/z values can be trapped. Thus, we have something called the 1/3rd rule. This rule approximates the equation and states that any fragment created that is < 1/3 the m/z of the precursor will not be trapped. This rule makes CID less desirable as a primary fragmentation technique, thus more used as the pinch hitter when needed for alternate fragmentation after HCD is shown inefficient for a specific molecule. Furthermore, CID is a slow heating process. As such, activation/fragmentation takes a few milliseconds, whereas HCD is effectively instantaneous. 

Recent developments in oligonucleotide analysis have shown great strides in using CID for fragmentation of many types, including tRNA, snRNA and digested mRNA. To gain the most impact from these experiments, scientists can adjust the q-value applied within the CID resonance. This is usually set to 0.25; however, for oligonucleotides, lowering that value to 0.2-0.15 can help prevent unwanted fragmentation from occurring and specify backbone cleavage that can be more easily characterized.

Source Induced Dissociation (SID)

  • Process: SID is a collisional-induced activation (and if desired, dissociation) technique that is induced in the source region of mass spectrometers. By applying more voltage in the source region, ions are drawn into the mass spectrometer with more energy. Within the source region, there are still atmospheric inert gases that have been drawn in with the analytes of interest. With the additional energy, collisions with these inert gases can occur. Since SID occurs before any mass analyzer, it applies to all ions that enter the mass spectrometer.
  • Applications: SID is most commonly used as a desolvation technique for large native molecules. By applying SID energy in the source, molecules generate vibrational energy, and non-covalent, low-energy bonds will be broken only.
  • Analogy: SID could be compared to shaking off an umbrella after coming in from the rain. The jostle doesn’t cause the whole umbrella to fall apart, but it does cause the water droplets to fall off. This is a little like proteins getting desolvated from SID.
  • Considerations: SID can be a destructive, dissociative process as well, so if that is not the goal, it is important to monitor the energy applied to the analyte. For example, it has been common practice in top-down proteomics to use a small (10-15 eV) setting for SID to aid in desolvation. However, I have successfully taken pure protein (or well chromatographically separated protein) and applied much higher SID (>50 V) to induce dissociation. The resulting fragmentation pattern can yield surprisingly good results. However, this is not common practice, especially not for coeluting species, as the fragmentation would occur on all proteins eluting.

Another consideration for SID is its use in native MS. This is often vital for good detection of native proteins like antibodies. Often, we will employ a stronger SID in these cases, anywhere from 60-120 V, depending on the size and complexity of the protein/protein complex, to ensure proper desolvation and thus detection.

Learn more about collisional induced dissociation (CID) and higher energy collisional induced dissociation (HCD)

Visit us on LinkedIn: #CID #HCD #fragmentation

Amanda Lee

Amanda Lee PhD, Product Marketing Manager

Thermo Fisher Scientific
LinkedIn Logo
 

Related content

Oligonucleotide Analysis Using the Agilent InfinityLab Pro iQ and Altura Oligo HPH-C18 Column

Applications
| 2026 | Agilent Technologies
Instrumentation
LC/MS, LC/SQ, Consumables, LC columns
Manufacturer
Agilent Technologies
Industries
Pharma & Biopharma

Simultaneous Quantitation and Discovery analysis: Combining targeted and untargeted metabolomics on Orbitrap mass spectrometers

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Metabolomics

Analysis of Per- and Polyfluoroalkyl Substances (PFAS) in Wastewater

Applications
| 2026 | Agilent Technologies
Instrumentation
LC/MS, LC/MS/MS, LC/QQQ
Manufacturer
Agilent Technologies
Industries
Environmental

Non-Targeted Screening of Biosolids with the Xevo™ MRT Mass Spectrometer Reveals New Isoforms of PFAS

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/TOF, LC/HRMS
Manufacturer
Waters
Industries
Environmental

High Resolution Characterization of Lipid Nanoparticles Using the Xevo™ Charge Detection Mass Spectrometry (CDMS) Instrument - Single Particle Mass Analysis of Intact LNP-mRNA Formulations

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/HRMS, Particle characterization, LC/IT
Manufacturer
Waters
Industries
Lipidomics
 

Related articles

News from LabRulezLCMS Library - Week 46, 2024
Article | Application

News from LabRulezLCMS Library - Week 46, 2024

This week we bring you applications and other documents by SIELC, Thermo Fisher Scientific, Shimadzu, Waters Corporation, and Agilent Technologies!
LabRulez
tag
share
more
Why You Need a Verification-Class Mass Spectrometer: Solving the Top 5 Pain Points of Translational Metabolomics Researchers
Article | Product

Why You Need a Verification-Class Mass Spectrometer: Solving the Top 5 Pain Points of Translational Metabolomics Researchers

This is the third in a multi-part blog series highlighting the key attributes and impressive benefits of the Stellar mass spectrometer—the first-of-its-kind instrument.
Thermo Fisher Scientific
tag
share
more
General Screening and Multiple Dissociation Methods for Complementary LC–MS Analysis of Pesticides in Beverages: Potential and Pitfalls
Scientific article | Science and research

General Screening and Multiple Dissociation Methods for Complementary LC–MS Analysis of Pesticides in Beverages: Potential and Pitfalls

A ZenoTOF 7600 modified for CID, EAD, and UVPD allowed LC–MS screening of pesticides in juices and wines. About 30 were detected at pg–ng/mL levels, confirmed by multiple dissociation methods.
LabRulez
tag
share
more
Characterization of Asian lacquers by atmospheric solids analysis probe high resolution tandem mass spectrometry coupled with cyclic ion mobility separation
Scientific article | Science and research

Characterization of Asian lacquers by atmospheric solids analysis probe high resolution tandem mass spectrometry coupled with cyclic ion mobility separation

This study uses ASAP-IMS-HRMS to characterize and classify Asian lacquers by origin, detecting phenols, pigments, and resins in historical samples.
LabRulez
tag
share
more
 

Related content

Oligonucleotide Analysis Using the Agilent InfinityLab Pro iQ and Altura Oligo HPH-C18 Column

Applications
| 2026 | Agilent Technologies
Instrumentation
LC/MS, LC/SQ, Consumables, LC columns
Manufacturer
Agilent Technologies
Industries
Pharma & Biopharma

Simultaneous Quantitation and Discovery analysis: Combining targeted and untargeted metabolomics on Orbitrap mass spectrometers

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Metabolomics

Analysis of Per- and Polyfluoroalkyl Substances (PFAS) in Wastewater

Applications
| 2026 | Agilent Technologies
Instrumentation
LC/MS, LC/MS/MS, LC/QQQ
Manufacturer
Agilent Technologies
Industries
Environmental

Non-Targeted Screening of Biosolids with the Xevo™ MRT Mass Spectrometer Reveals New Isoforms of PFAS

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/TOF, LC/HRMS
Manufacturer
Waters
Industries
Environmental

High Resolution Characterization of Lipid Nanoparticles Using the Xevo™ Charge Detection Mass Spectrometry (CDMS) Instrument - Single Particle Mass Analysis of Intact LNP-mRNA Formulations

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/HRMS, Particle characterization, LC/IT
Manufacturer
Waters
Industries
Lipidomics
 

Related articles

News from LabRulezLCMS Library - Week 46, 2024
Article | Application

News from LabRulezLCMS Library - Week 46, 2024

This week we bring you applications and other documents by SIELC, Thermo Fisher Scientific, Shimadzu, Waters Corporation, and Agilent Technologies!
LabRulez
tag
share
more
Why You Need a Verification-Class Mass Spectrometer: Solving the Top 5 Pain Points of Translational Metabolomics Researchers
Article | Product

Why You Need a Verification-Class Mass Spectrometer: Solving the Top 5 Pain Points of Translational Metabolomics Researchers

This is the third in a multi-part blog series highlighting the key attributes and impressive benefits of the Stellar mass spectrometer—the first-of-its-kind instrument.
Thermo Fisher Scientific
tag
share
more
General Screening and Multiple Dissociation Methods for Complementary LC–MS Analysis of Pesticides in Beverages: Potential and Pitfalls
Scientific article | Science and research

General Screening and Multiple Dissociation Methods for Complementary LC–MS Analysis of Pesticides in Beverages: Potential and Pitfalls

A ZenoTOF 7600 modified for CID, EAD, and UVPD allowed LC–MS screening of pesticides in juices and wines. About 30 were detected at pg–ng/mL levels, confirmed by multiple dissociation methods.
LabRulez
tag
share
more
Characterization of Asian lacquers by atmospheric solids analysis probe high resolution tandem mass spectrometry coupled with cyclic ion mobility separation
Scientific article | Science and research

Characterization of Asian lacquers by atmospheric solids analysis probe high resolution tandem mass spectrometry coupled with cyclic ion mobility separation

This study uses ASAP-IMS-HRMS to characterize and classify Asian lacquers by origin, detecting phenols, pigments, and resins in historical samples.
LabRulez
tag
share
more
 

Related content

Oligonucleotide Analysis Using the Agilent InfinityLab Pro iQ and Altura Oligo HPH-C18 Column

Applications
| 2026 | Agilent Technologies
Instrumentation
LC/MS, LC/SQ, Consumables, LC columns
Manufacturer
Agilent Technologies
Industries
Pharma & Biopharma

Simultaneous Quantitation and Discovery analysis: Combining targeted and untargeted metabolomics on Orbitrap mass spectrometers

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Metabolomics

Analysis of Per- and Polyfluoroalkyl Substances (PFAS) in Wastewater

Applications
| 2026 | Agilent Technologies
Instrumentation
LC/MS, LC/MS/MS, LC/QQQ
Manufacturer
Agilent Technologies
Industries
Environmental

Non-Targeted Screening of Biosolids with the Xevo™ MRT Mass Spectrometer Reveals New Isoforms of PFAS

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/TOF, LC/HRMS
Manufacturer
Waters
Industries
Environmental

High Resolution Characterization of Lipid Nanoparticles Using the Xevo™ Charge Detection Mass Spectrometry (CDMS) Instrument - Single Particle Mass Analysis of Intact LNP-mRNA Formulations

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/HRMS, Particle characterization, LC/IT
Manufacturer
Waters
Industries
Lipidomics
 

Related articles

News from LabRulezLCMS Library - Week 46, 2024
Article | Application

News from LabRulezLCMS Library - Week 46, 2024

This week we bring you applications and other documents by SIELC, Thermo Fisher Scientific, Shimadzu, Waters Corporation, and Agilent Technologies!
LabRulez
tag
share
more
Why You Need a Verification-Class Mass Spectrometer: Solving the Top 5 Pain Points of Translational Metabolomics Researchers
Article | Product

Why You Need a Verification-Class Mass Spectrometer: Solving the Top 5 Pain Points of Translational Metabolomics Researchers

This is the third in a multi-part blog series highlighting the key attributes and impressive benefits of the Stellar mass spectrometer—the first-of-its-kind instrument.
Thermo Fisher Scientific
tag
share
more
General Screening and Multiple Dissociation Methods for Complementary LC–MS Analysis of Pesticides in Beverages: Potential and Pitfalls
Scientific article | Science and research

General Screening and Multiple Dissociation Methods for Complementary LC–MS Analysis of Pesticides in Beverages: Potential and Pitfalls

A ZenoTOF 7600 modified for CID, EAD, and UVPD allowed LC–MS screening of pesticides in juices and wines. About 30 were detected at pg–ng/mL levels, confirmed by multiple dissociation methods.
LabRulez
tag
share
more
Characterization of Asian lacquers by atmospheric solids analysis probe high resolution tandem mass spectrometry coupled with cyclic ion mobility separation
Scientific article | Science and research

Characterization of Asian lacquers by atmospheric solids analysis probe high resolution tandem mass spectrometry coupled with cyclic ion mobility separation

This study uses ASAP-IMS-HRMS to characterize and classify Asian lacquers by origin, detecting phenols, pigments, and resins in historical samples.
LabRulez
tag
share
more
 

Related content

Oligonucleotide Analysis Using the Agilent InfinityLab Pro iQ and Altura Oligo HPH-C18 Column

Applications
| 2026 | Agilent Technologies
Instrumentation
LC/MS, LC/SQ, Consumables, LC columns
Manufacturer
Agilent Technologies
Industries
Pharma & Biopharma

Simultaneous Quantitation and Discovery analysis: Combining targeted and untargeted metabolomics on Orbitrap mass spectrometers

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Metabolomics

Analysis of Per- and Polyfluoroalkyl Substances (PFAS) in Wastewater

Applications
| 2026 | Agilent Technologies
Instrumentation
LC/MS, LC/MS/MS, LC/QQQ
Manufacturer
Agilent Technologies
Industries
Environmental

Non-Targeted Screening of Biosolids with the Xevo™ MRT Mass Spectrometer Reveals New Isoforms of PFAS

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/TOF, LC/HRMS
Manufacturer
Waters
Industries
Environmental

High Resolution Characterization of Lipid Nanoparticles Using the Xevo™ Charge Detection Mass Spectrometry (CDMS) Instrument - Single Particle Mass Analysis of Intact LNP-mRNA Formulations

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/HRMS, Particle characterization, LC/IT
Manufacturer
Waters
Industries
Lipidomics
 

Related articles

News from LabRulezLCMS Library - Week 46, 2024
Article | Application

News from LabRulezLCMS Library - Week 46, 2024

This week we bring you applications and other documents by SIELC, Thermo Fisher Scientific, Shimadzu, Waters Corporation, and Agilent Technologies!
LabRulez
tag
share
more
Why You Need a Verification-Class Mass Spectrometer: Solving the Top 5 Pain Points of Translational Metabolomics Researchers
Article | Product

Why You Need a Verification-Class Mass Spectrometer: Solving the Top 5 Pain Points of Translational Metabolomics Researchers

This is the third in a multi-part blog series highlighting the key attributes and impressive benefits of the Stellar mass spectrometer—the first-of-its-kind instrument.
Thermo Fisher Scientific
tag
share
more
General Screening and Multiple Dissociation Methods for Complementary LC–MS Analysis of Pesticides in Beverages: Potential and Pitfalls
Scientific article | Science and research

General Screening and Multiple Dissociation Methods for Complementary LC–MS Analysis of Pesticides in Beverages: Potential and Pitfalls

A ZenoTOF 7600 modified for CID, EAD, and UVPD allowed LC–MS screening of pesticides in juices and wines. About 30 were detected at pg–ng/mL levels, confirmed by multiple dissociation methods.
LabRulez
tag
share
more
Characterization of Asian lacquers by atmospheric solids analysis probe high resolution tandem mass spectrometry coupled with cyclic ion mobility separation
Scientific article | Science and research

Characterization of Asian lacquers by atmospheric solids analysis probe high resolution tandem mass spectrometry coupled with cyclic ion mobility separation

This study uses ASAP-IMS-HRMS to characterize and classify Asian lacquers by origin, detecting phenols, pigments, and resins in historical samples.
LabRulez
tag
share
more
Other projects
GCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike