Why are my recoveries greater than 100%?

We, 21.1.2026
| Original article from: Phenomenex
Although 100% recovery is expected in SPE, values above 100% are sometimes observed. This article explains the most common causes and provides practical guidance on how to identify and resolve them.
<p><strong>Phenomenex: </strong>Why are my recoveries greater than 100%?</p>

Phenomenex: Why are my recoveries greater than 100%?

When performing analyte extraction with SPE we expect 100% recovery. If recovery is lower, this would suggest there is an issue with retrieval of the analyte from the sorbent. But what are some of the reasons you may see a recovery greater than what you were expecting (or over 100%)?

Common problems

  • Co-eluting interferences
  • Contaminants from the solvent or sorbents selected
  • Internal standard is compromised making your reference point inaccurate
  • Recovery calculations are inaccurate

Let's examine each of these common problems.

Co-eluting interferences

These are typically present in the matrix and the goal of the SPE step is to remove them. If you are experiencing recoveries greater than 100%, the first step would be to check the eluent using an orthogonal analytical technique to identify that the increased recovery is a result of a co-elution. 

Where a co-elution is identified, there are a few possible solutions.

  1. Modify the wash steps of the SPE method to selectively remove the impurity during washing
  2. Select an alternative SPE sorbent which will better remove these impurities. This is easier if the analyte / impurity can be selectively ionized, as an ion-exchange sorbent will very often afford cleaner extracts than either reversed or normal phase solutions
  3. Adjust your analytical method to resolve the 2 peaks analytically, negating the necessity to remove this impurity during the SPE step.
  4. Run a blank plasma sample; this will allow you to ascertain the impurities present and use this blank in the calculations to subtract from your recovery

Contaminants from the solvent or sorbent

These will typically be impurities present in the starting materials which are similar in nature to your compound of interest, and as a result co-elute with it during SPE and subsequent analytical analysis. You can follow the steps above to remove co-eluting impurities, or alternatively as this is not an impurity present in your matrix, you can identify the level present in the sorbent using blank runs and subtract this from your main peak during the calculation. Prior to addition of your sample if you equilibrate your sorbent with the elution solvent this should flush out potential contaminants from the sorbent. For impurities introduced from the solvent, typically switching the solvent should resolve this.

Internal standard is compromised

If your internal standard is not showing 100% recovery, your subsequent calculations will be inaccurate. It is always good practice to use an internal standard similar in nature to your compound of interest as this will give the more consistent reference point for your method. 

If there is concern your internal standard is not showing good recovery from the SPE material using your method, you can use an external standard which is added to the elution solvent to calculate absolute recovery.

Recovery calculations are inaccurate

It is important to follow this step wise to ascertain where the errors are appearing.

  1. Calculation Errors - Check all calculations and account for all dilution factors
  2. Integration Errors - Reintegrate all peaks and use manual calculations to confirm
  3. Analytical method errors – Ensure the analytical method used allows correct integration and calculation. Ensure no strong solvents are used for injection and run a calibration across a number of dilutions to ensure linearity. Troubleshoot your analytical method to determine any factors which may lead to poor reproducibility
  4. Ensure detections selected are appropriate and working correctly.

We hope this tip was useful! Check out our resources on solid phase extraction. See you next month for another tip!

Phenomenex
LinkedIn Logo
 

Related content

Determination of Authenticity of Manuka Honey Using the MALDI-8030 Benchtop Linear MALDI-TOF Mass Spectrometer

Applications
| 2026 | Shimadzu
Instrumentation
MALDI, LC/MS, LC/TOF
Manufacturer
Shimadzu
Industries
Food & Agriculture

Analysis of Inorganic Anions in Drinking Water According to EPA Method 300.1 Using Nexera IC -Part A

Applications
| 2026 | Shimadzu
Instrumentation
Ion chromatography
Manufacturer
Shimadzu
Industries
Food & Agriculture

Quality Evaluation of mRNA by Electrophoresis

Applications
| 2026 | Shimadzu
Instrumentation
Capillary electrophoresis
Manufacturer
Shimadzu
Industries
Pharma & Biopharma

LC/MS dMRM Method Refinement Expands Targeted Lipidomics Studies from Plasma to Cells and Tissues

Applications
| 2026 | Agilent Technologies
Instrumentation
LC/MS, LC/MS/MS, LC/QQQ
Manufacturer
Agilent Technologies
Industries
Lipidomics, Clinical Research

Sensitive Detection of Aflatoxins by the Agilent 1290 Infinity III Fluorescence Detector

Applications
| 2026 | Agilent Technologies
Instrumentation
HPLC
Manufacturer
Agilent Technologies
Industries
Food & Agriculture
 

Related articles

Introducing the new Thermo Scientific TSQ Certis Triple Quadrupole Mass Spectrometer
Article | Product

Introducing the new Thermo Scientific TSQ Certis Triple Quadrupole Mass Spectrometer

Thermo Scientific TSQ Certis Triple Quadrupole MS delivers rugged uptime, next-gen quantitative performance, and high-throughput LC-MS/MS for routine, clinical, PFAS, and biopharma labs.
Thermo Fisher Scientific
tag
share
more
ISC 2026 Deadline for abstract submission
Article | Events

ISC 2026 Deadline for abstract submission

Only 8 days left to submit your abstract for an oral presentation at ISC 2026 in Prague. Secure your slot—and your chance to present in Tutorials & Young Speakers.
International Symposium on Chromatography - (ISC)
tag
share
more
Oligos Made Easy - Part 5: SEC
Article | Academy

Oligos Made Easy - Part 5: SEC

Learn how SEC streamlines oligonucleotide purification. From pore size selection to flow optimization, discover practical tips for desalting, buffer exchange, and clean size-based separation.
KNAUER
tag
share
more
News from LabRulezLCMS Library - Week 10, 2026
Article | Application

News from LabRulezLCMS Library - Week 10, 2026

This week we bring you application notes by KNAUER, Shimadzu, Thermo Fisher Scientific and Waters Corporation and technical note by Agilent Technologies!
LabRulez
tag
share
more
 

Related content

Determination of Authenticity of Manuka Honey Using the MALDI-8030 Benchtop Linear MALDI-TOF Mass Spectrometer

Applications
| 2026 | Shimadzu
Instrumentation
MALDI, LC/MS, LC/TOF
Manufacturer
Shimadzu
Industries
Food & Agriculture

Analysis of Inorganic Anions in Drinking Water According to EPA Method 300.1 Using Nexera IC -Part A

Applications
| 2026 | Shimadzu
Instrumentation
Ion chromatography
Manufacturer
Shimadzu
Industries
Food & Agriculture

Quality Evaluation of mRNA by Electrophoresis

Applications
| 2026 | Shimadzu
Instrumentation
Capillary electrophoresis
Manufacturer
Shimadzu
Industries
Pharma & Biopharma

LC/MS dMRM Method Refinement Expands Targeted Lipidomics Studies from Plasma to Cells and Tissues

Applications
| 2026 | Agilent Technologies
Instrumentation
LC/MS, LC/MS/MS, LC/QQQ
Manufacturer
Agilent Technologies
Industries
Lipidomics, Clinical Research

Sensitive Detection of Aflatoxins by the Agilent 1290 Infinity III Fluorescence Detector

Applications
| 2026 | Agilent Technologies
Instrumentation
HPLC
Manufacturer
Agilent Technologies
Industries
Food & Agriculture
 

Related articles

Introducing the new Thermo Scientific TSQ Certis Triple Quadrupole Mass Spectrometer
Article | Product

Introducing the new Thermo Scientific TSQ Certis Triple Quadrupole Mass Spectrometer

Thermo Scientific TSQ Certis Triple Quadrupole MS delivers rugged uptime, next-gen quantitative performance, and high-throughput LC-MS/MS for routine, clinical, PFAS, and biopharma labs.
Thermo Fisher Scientific
tag
share
more
ISC 2026 Deadline for abstract submission
Article | Events

ISC 2026 Deadline for abstract submission

Only 8 days left to submit your abstract for an oral presentation at ISC 2026 in Prague. Secure your slot—and your chance to present in Tutorials & Young Speakers.
International Symposium on Chromatography - (ISC)
tag
share
more
Oligos Made Easy - Part 5: SEC
Article | Academy

Oligos Made Easy - Part 5: SEC

Learn how SEC streamlines oligonucleotide purification. From pore size selection to flow optimization, discover practical tips for desalting, buffer exchange, and clean size-based separation.
KNAUER
tag
share
more
News from LabRulezLCMS Library - Week 10, 2026
Article | Application

News from LabRulezLCMS Library - Week 10, 2026

This week we bring you application notes by KNAUER, Shimadzu, Thermo Fisher Scientific and Waters Corporation and technical note by Agilent Technologies!
LabRulez
tag
share
more
 

Related content

Determination of Authenticity of Manuka Honey Using the MALDI-8030 Benchtop Linear MALDI-TOF Mass Spectrometer

Applications
| 2026 | Shimadzu
Instrumentation
MALDI, LC/MS, LC/TOF
Manufacturer
Shimadzu
Industries
Food & Agriculture

Analysis of Inorganic Anions in Drinking Water According to EPA Method 300.1 Using Nexera IC -Part A

Applications
| 2026 | Shimadzu
Instrumentation
Ion chromatography
Manufacturer
Shimadzu
Industries
Food & Agriculture

Quality Evaluation of mRNA by Electrophoresis

Applications
| 2026 | Shimadzu
Instrumentation
Capillary electrophoresis
Manufacturer
Shimadzu
Industries
Pharma & Biopharma

LC/MS dMRM Method Refinement Expands Targeted Lipidomics Studies from Plasma to Cells and Tissues

Applications
| 2026 | Agilent Technologies
Instrumentation
LC/MS, LC/MS/MS, LC/QQQ
Manufacturer
Agilent Technologies
Industries
Lipidomics, Clinical Research

Sensitive Detection of Aflatoxins by the Agilent 1290 Infinity III Fluorescence Detector

Applications
| 2026 | Agilent Technologies
Instrumentation
HPLC
Manufacturer
Agilent Technologies
Industries
Food & Agriculture
 

Related articles

Introducing the new Thermo Scientific TSQ Certis Triple Quadrupole Mass Spectrometer
Article | Product

Introducing the new Thermo Scientific TSQ Certis Triple Quadrupole Mass Spectrometer

Thermo Scientific TSQ Certis Triple Quadrupole MS delivers rugged uptime, next-gen quantitative performance, and high-throughput LC-MS/MS for routine, clinical, PFAS, and biopharma labs.
Thermo Fisher Scientific
tag
share
more
ISC 2026 Deadline for abstract submission
Article | Events

ISC 2026 Deadline for abstract submission

Only 8 days left to submit your abstract for an oral presentation at ISC 2026 in Prague. Secure your slot—and your chance to present in Tutorials & Young Speakers.
International Symposium on Chromatography - (ISC)
tag
share
more
Oligos Made Easy - Part 5: SEC
Article | Academy

Oligos Made Easy - Part 5: SEC

Learn how SEC streamlines oligonucleotide purification. From pore size selection to flow optimization, discover practical tips for desalting, buffer exchange, and clean size-based separation.
KNAUER
tag
share
more
News from LabRulezLCMS Library - Week 10, 2026
Article | Application

News from LabRulezLCMS Library - Week 10, 2026

This week we bring you application notes by KNAUER, Shimadzu, Thermo Fisher Scientific and Waters Corporation and technical note by Agilent Technologies!
LabRulez
tag
share
more
 

Related content

Determination of Authenticity of Manuka Honey Using the MALDI-8030 Benchtop Linear MALDI-TOF Mass Spectrometer

Applications
| 2026 | Shimadzu
Instrumentation
MALDI, LC/MS, LC/TOF
Manufacturer
Shimadzu
Industries
Food & Agriculture

Analysis of Inorganic Anions in Drinking Water According to EPA Method 300.1 Using Nexera IC -Part A

Applications
| 2026 | Shimadzu
Instrumentation
Ion chromatography
Manufacturer
Shimadzu
Industries
Food & Agriculture

Quality Evaluation of mRNA by Electrophoresis

Applications
| 2026 | Shimadzu
Instrumentation
Capillary electrophoresis
Manufacturer
Shimadzu
Industries
Pharma & Biopharma

LC/MS dMRM Method Refinement Expands Targeted Lipidomics Studies from Plasma to Cells and Tissues

Applications
| 2026 | Agilent Technologies
Instrumentation
LC/MS, LC/MS/MS, LC/QQQ
Manufacturer
Agilent Technologies
Industries
Lipidomics, Clinical Research

Sensitive Detection of Aflatoxins by the Agilent 1290 Infinity III Fluorescence Detector

Applications
| 2026 | Agilent Technologies
Instrumentation
HPLC
Manufacturer
Agilent Technologies
Industries
Food & Agriculture
 

Related articles

Introducing the new Thermo Scientific TSQ Certis Triple Quadrupole Mass Spectrometer
Article | Product

Introducing the new Thermo Scientific TSQ Certis Triple Quadrupole Mass Spectrometer

Thermo Scientific TSQ Certis Triple Quadrupole MS delivers rugged uptime, next-gen quantitative performance, and high-throughput LC-MS/MS for routine, clinical, PFAS, and biopharma labs.
Thermo Fisher Scientific
tag
share
more
ISC 2026 Deadline for abstract submission
Article | Events

ISC 2026 Deadline for abstract submission

Only 8 days left to submit your abstract for an oral presentation at ISC 2026 in Prague. Secure your slot—and your chance to present in Tutorials & Young Speakers.
International Symposium on Chromatography - (ISC)
tag
share
more
Oligos Made Easy - Part 5: SEC
Article | Academy

Oligos Made Easy - Part 5: SEC

Learn how SEC streamlines oligonucleotide purification. From pore size selection to flow optimization, discover practical tips for desalting, buffer exchange, and clean size-based separation.
KNAUER
tag
share
more
News from LabRulezLCMS Library - Week 10, 2026
Article | Application

News from LabRulezLCMS Library - Week 10, 2026

This week we bring you application notes by KNAUER, Shimadzu, Thermo Fisher Scientific and Waters Corporation and technical note by Agilent Technologies!
LabRulez
tag
share
more
Other projects
GCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike