1220 Infinity/Infinity II Mobile LC - Site Preparation Checklist
Manuals | 2016 | Agilent TechnologiesInstrumentation
Proper site preparation is crucial for deploying mobile liquid chromatography (LC) systems safely and ensuring their long-term reliability. By defining clear requirements for space, utilities and mounting, laboratories and mobile facilities can avoid operational delays and potential damage to sensitive instrumentation.
This checklist aims to guide customers through the necessary preparations before installation of the Agilent 1220 Infinity Mobile-LC Upgrade Kit. It covers customer responsibilities, spatial constraints, mounting procedures and environmental considerations to guarantee a smooth and timely setup.
The following components and setup steps are key for the mobile LC installation:
Customers are responsible for selecting appropriate fastening hardware (screws or bolts) capable of withstanding up to 2000 N per attachment point under 1 G acceleration. The fastener heads must not protrude more than 15 mm above the unit’s base.
Key installation and environmental criteria include:
Adherence to these guidelines delivers:
Advances in compact LC systems and mobile laboratory integration point toward:
Thorough site preparation is the foundation of successful mobile LC deployment. By following the outlined space, mounting and environmental requirements, users can ensure consistent instrument performance, operator safety and swift commissioning of analytics in mobile environments.
HPLC
IndustriesManufacturerAgilent Technologies
Summary
Importance of the topic
Proper site preparation is crucial for deploying mobile liquid chromatography (LC) systems safely and ensuring their long-term reliability. By defining clear requirements for space, utilities and mounting, laboratories and mobile facilities can avoid operational delays and potential damage to sensitive instrumentation.
Objectives and overview
This checklist aims to guide customers through the necessary preparations before installation of the Agilent 1220 Infinity Mobile-LC Upgrade Kit. It covers customer responsibilities, spatial constraints, mounting procedures and environmental considerations to guarantee a smooth and timely setup.
Methodology and instrumentation
The following components and setup steps are key for the mobile LC installation:
- Agilent 1220 Infinity Mobile-LC Upgrade Kit (G4292A)
- Agilent 1220 Infinity LC modules:
G4290B/C (1220 LC with VWD sensor)
G4294B (1220 LC with DAD detector) - Attenuation unit featuring four 8.5 mm diameter lower-plate holes and corresponding 20 mm upper-plate holes for secure fastening
Customers are responsible for selecting appropriate fastening hardware (screws or bolts) capable of withstanding up to 2000 N per attachment point under 1 G acceleration. The fastener heads must not protrude more than 15 mm above the unit’s base.
Main requirements and discussion
Key installation and environmental criteria include:
- Bench space: Minimum clear area of 560 mm (W) × 665 mm (D) × 970 mm (H), with an extra 200 mm above solvent bottles for handling
- Workbench strength and material: Must support combined weight of LC module (30–43 kg), mobile kit (14.2 kg) and solvent containers
- Mounting solution: Four secure fastening points aligned with unit holes, using customer-provided hardware suitable for bench material
- Environmental conditions: Operating temperature range per instrument specifications; adequate ventilation to exhaust solvent vapors; protection from direct sunlight, drafts and vibration; gradual temperature equilibration to prevent internal condensation
Benefits and practical applications
Adherence to these guidelines delivers:
- Reliable chromatographic performance and detector stability during vehicle motion
- Enhanced user safety through proper solvent containment and ventilation
- Reduced installation times and minimized risk of instrument damage
- Flexibility for on-site analyses in field laboratories, environmental monitoring, forensic testing and mobile QC operations
Future trends and possibilities
Advances in compact LC systems and mobile laboratory integration point toward:
- Miniaturized, high-throughput LC modules for rapid field testing
- Wireless data connectivity and remote monitoring of system health
- Automated sample preparation combined with on-board analysis platforms
- Regulatory-compliant mobile units for pharmaceutical and environmental screening in remote areas
Conclusion
Thorough site preparation is the foundation of successful mobile LC deployment. By following the outlined space, mounting and environmental requirements, users can ensure consistent instrument performance, operator safety and swift commissioning of analytics in mobile environments.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
1220 Infinity/Infinity II LC - Site Preparation Checklist
2016|Agilent Technologies|Manuals
1220 Infinity/Infinity II LC Site Preparation Checklist 1220Preparation Site Infinity/Infinity Checklist II LC Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of…
Key words
site, sitechecklist, checklistoperating, operatingtechnologies, technologiescustomer, customerweight, weightpreparation, preparationambient, ambientagilent, agilentconsumption, consumptionheight, heightinstallation, installationinfinity, infinityhumidity, humiditypower
Agilent 1220 Infinity/Infinity II LC - Site Preparation Checklist
2021|Agilent Technologies|Manuals
Agilent CrossLab Start Up Services Agilent 1220 Infinity/Infinity II LC Site Preparation Checklist Thank you for purchasing an instrument from Agilent Technologies. CrossLab Start Up is focused on helping customers shorten the time it takes to start realizing the full…
Key words
special, specialsite, sitecustomer, customeroperating, operatingnotes, notesweight, weightinstrument, instrumentservice, serviceinstallation, installationyour, yourrequirements, requirementsambient, ambientpreparation, preparationdescription, descriptionconsumption
Agilent 1290 Infinity LC/1260 Infinity LC - Site Preparation Checklist
2020|Agilent Technologies|Manuals
1290 Infinity LC/1260 Infinity LC Site Preparation Checklist 1290Preparation Site Infinity LC/1260 Checklist Infinity LC Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or…
Key words
checklist, checklistsite, sitetechnologies, technologiesagilent, agilentpreparation, preparationfrpsduwphqw, frpsduwphqwxwrvdpsohu, xwrvdpsohuhwhfwru, hwhfwruroxpq, roxpqhjdvvhu, hjdvvhurequirements, requirementsdescription, descriptionrrohu, rrohucustomer, customerconsumption
Agilent 6500 Series Q-TOF LC/MS Systems – Site Preparation Checklist
2017|Agilent Technologies|Manuals
6500 Series Q-TOF LC/MS Systems – Site Preparation Checklist Thank you for purchasing an Agilent system. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation…
Key words
checklist, checklistsite, sitetechnologies, technologiesrough, roughley, leyagilent, agilentpreparation, preparationrobert, robertsource, sourceengineering, engineeringspecial, specialcondensing, condensingpumps, pumpspump, pumpsupport