Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Applications | 2015 | Agilent TechnologiesInstrumentation
LC/MS, LC/MS/MS, LC/QQQ
Industries
Forensics
Manufacturer
Agilent Technologies

Summary

Importance of the Topic


The reliable quantitation of amphetamines and designer stimulants in biological matrices is critical for forensic and clinical toxicology. Traditional GC/MS approaches require lengthy sample preparation and derivatization, while LC-MS/MS offers improved throughput, sensitivity, and specificity. This method addresses the need for rapid, robust analysis to support casework and quality control in forensic laboratories.

Study Objectives and Overview


This study outlines the development and validation of an LC-MS/MS protocol using an Agilent 6430 Triple Quadrupole system for thirteen stimulant targets. Following SWGTOX and Virginia Department of Forensic Science guidelines, the method aims to achieve full quantitative performance (linearity, precision, accuracy, limits of detection and quantitation) within a nine-minute analytical cycle.

Methodology and Instrumentation


Extraction: Liquid-liquid extraction from blood with saturated trisodium phosphate buffer and 1-chlorobutane.
Evaporation: Samples dried under gentle heat after acidification.
Reconstitution: Final dilution in aqueous formic acid prior to injection.
Chromatography and Detection: Gradient elution on an Agilent Poroshell EC-C18 column (2.1×75 mm, 2.7 µm) at 50 °C, using 0.1 % formic acid in water and acetonitrile at 0.5 mL/min. Positive ESI dynamic MRM mode monitored optimized mass transitions in a single nine-minute run.

Used Instrumentation


  • Agilent 6430 Triple Quadrupole Mass Spectrometer
  • Agilent 1260 Infinity LC System
  • Agilent Poroshell 120 EC-C18, 2.1×75 mm, 2.7 µm
  • Zymark TurboVap Evaporator
  • Agilent MassHunter Optimizer Software

Key Results and Discussion


The method provided baseline resolution for all 13 analytes within nine minutes, with sharp peak shapes and minimal tailing. Calibration employed linear (1/x) models for certain cathinones and quadratic (1/x) models for the remaining targets, achieving correlation coefficients above 0.985 over the 0.01–2.00 mg/L range. Representative chromatograms and MRM transitions demonstrated consistent ion ratios and retention times. Validation confirmed acceptable precision (<15% RSD), accuracy (±15%), low carryover, robust extraction recovery, and minimal matrix effects in whole blood.

Benefits and Practical Applications of the Method


  • Significantly reduced sample preparation compared to GC/MS, eliminating derivatization.
  • High throughput with nine-minute analysis supporting large sample batches.
  • Enhanced sensitivity and specificity for routine forensic and clinical toxicology.
  • Ready for implementation under SWGTOX and departmental validation guidelines.

Future Trends and Potential Applications


Expanding the analyte panel to cover emerging synthetic stimulants and other drug classes will further increase the method’s utility. Incorporation of automated sample preparation systems and high-resolution mass spectrometry may streamline workflows and improve detection limits. Advanced data-driven MRM scheduling and quantitation algorithms could enhance sensitivity, dynamic range, and robustness.

Conclusion


The Agilent 6430 LC-MS/MS protocol delivers a fast, robust, and fully validated approach for quantifying amphetamines, phentermine, and designer stimulants in whole blood. Its compliance with forensic validation standards makes it an efficient and reliable tool for routine toxicology laboratories.

Reference


  • SWGTOX Standard Practices for Method Validation in Forensic Toxicology, 2013.
  • Virginia Department of Forensic Science Toxicology Procedures Manual.
  • Hudson J., Hutchings J., Wagner R., Friel P. Quantitative Method for Amphetamines, Phentermine, and Designer Stimulants Using an Agilent 6430 LC/MS/MS, Agilent Technologies Application Note, 2015.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Quantitative Method for Amphetamines, Phentermine, and Designer Stimulants Using an Agilent 6430 LC/MS/MS
Quantitative Method for Amphetamines, Phentermine, and Designer Stimulants Using an Agilent 6430 LC/MS/MS Application Note Forensic Toxicology Authors Abstract Jason Hudson, Ph.D., A method was developed for the quantitation of amphetamines, phentermine, and James Hutchings, Ph.D., and designer stimulants in…
Key words
methedrone, methedronemethcathinone, methcathinonepvp, pvpmethylone, methylonebupropion, bupropionmephedrone, mephedronemdpv, mdpvpseudoephedrine, pseudoephedrinephentermine, phenterminemethamphetamine, methamphetamineamphetamine, amphetaminemdma, mdmamda, mdasuppression, suppressionenhancement
A Triple Quadrupole LC/MS/MS Method for Quantitative Analysis of Methylenedioxypyrovalerone (MDPV) and Mephedrone, Common Components of “Bath Salts” in Urine
A Triple Quadrupole LC/MS/MS Method for Quantitative Analysis of Methylenedioxypyrovalerone (MDPV) and Mephedrone, Common Components of “Bath Salts” in Urine Application Note Forensic Toxicology Authors Abstract Guiping Lu, Ph.D. Due to the emerging use of synthetic cathinones – compounds widely…
Key words
mdpv, mdpvmephedrone, mephedronesynthetic, syntheticacs, acsurine, urinecathinones, cathinonesreagent, reagentgrade, grademethylenedioxypyrovalerone, methylenedioxypyrovaleronemrm, mrmspiked, spikedbath, bathparameters, parameterswere, werespe
Ultrafast Screen for Bath Salts in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System
Ultrafast Screen for Bath Salts in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System Application Note Forensic Toxicology Authors Abstract Nikunj R. Parikh, Michelle Romm, and Vaughn P. Miller Agilent Technologies, Inc. Wakefield, MA USA The need for greater…
Key words
interday, interdayintraday, intradayprod, prodauc, aucrapidfire, rapidfiremethoxymethcathinone, methoxymethcathinonemethylone, methyloneurine, urinemethcathinone, methcathinonesalts, saltsmdpv, mdpvmephedrone, mephedronebath, bathfluoromethcathinone, fluoromethcathinoneratio
Analysis of “Bath Salt” Compounds from Urine for Forensic Toxicology Using μElution Mixed-mode SPE Combined with UPLC/MS/MS Detection
Analysis of “Bath Salt” Compounds from Urine for Forensic Toxicology Using µElution Mixed-mode SPE Combined with UPLC/MS/MS Detection Jonathan P. Danaceau, Erin E. Chambers, and Kenneth J. Fountain Waters Corporation, Milford, MA, USA A P P L I C AT…
Key words
µelution, µelutionbath, bathtoxicology, toxicologyuplc, uplcforensic, forensicurine, urinespe, spesalt, saltcompounds, compoundscathinones, cathinonesmixed, mixedalphapyrrolidinopentiophenone, alphapyrrolidinopentiophenonecombined, combinedmdppp, mdpppmcx
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike