Presentations |  | ShimadzuInstrumentation
LC/MS, LC/MS/MS, LC/QQQ
Industries
Manufacturer
Shimadzu

Summary

Significance of the Topic


High-throughput liquid chromatography coupled to mass spectrometry is increasingly important for rapid and accurate analysis of complex samples with very narrow chromatographic peaks. The ability to generate sharp peaks in UHPLC calls for correspondingly fast MS acquisition to preserve chromatographic fidelity and ensure reliable qualitative and quantitative results.

Objectives and Study Overview


This study evaluates the impact of MS scanning speed and polarity switching rates on the detection performance of UHPLC peaks as narrow as 200 ms. The primary goals are to determine the minimum sampling rate required for accurate peak characterization and to demonstrate the benefits of ultra-fast mass measurements up to 15,000 Da/s with switching times of 15 ms.

Methodology and Instrumentation


The experimental setup comprises a UHPLC system with a 2.1 × 50 mm, 1.8 µm C18 column operated at 50 °C and a binary gradient (0.1% formic acid in water/acetonitrile) at 1.8 mL/min. Detection is performed by a quadrupole-based mass spectrometer with electrospray ionization and:
  • Scan speed up to 15,000 Da/s (≈1 Da/67 µs)
  • Polarity switching time of 15 ms
  • Interscan delay ≈ 5 ms
  • UHPLC-MS/MS MRM capability at 500 transitions/s (1 ms dwell, 1 ms pause)

This configuration allows acquisition of 26–85 data points across a 1 s peak, depending on mass range.

Main Results and Discussion


Results show that slower scan speeds lead to reduced peak height, broadened peak widths, and loss of quantitative reliability when points/peak drop below 10. At 15,000 Da/s, normalized peak heights remain at full sensitivity, and peak widths stay narrow. Key findings include:
  • Maintenance of sensitivity at ultra-high scan rates compared to conventional approaches
  • Adequate sampling (≥ 20 points/peak) for 200–500 Da mass ranges
  • Successful analysis of 226 pesticides in a 2 min UHPLC-MS/MS run using ultrafast MRM and full-scan confirmations
  • Simultaneous positive/negative ion detection without sensitivity loss

The study highlights the importance of balancing scan speed, polarity switching, and sensitivity to achieve reliable UHPLC-MS performance.

Benefits and Practical Applications


High-speed MS detection in UHPLC workflows offers several advantages:
  • Reduced risk of peak misidentification and coelution errors through mass-based separation
  • Improved quantitative accuracy and reproducibility by preserving peak shape
  • Enhanced detection of unexpected or low-abundance compounds
  • Capability to perform rapid multi-residue and pharmaceutical analyses within minutes


Future Trends and Applications


Ongoing developments are expected in:
  • Further increases in scan and polarity switching rates for sub-100 ms peak detection
  • Integration of high-resolution MS and real-time data processing with AI for automated identification
  • Expansion of multiplexed MRM workflows for comprehensive screening in environmental, food, and clinical labs
  • Miniaturized high-throughput platforms combining microflow LC with ultra-fast MS detection


Conclusion


Ultra-fast MS scanning and polarity switching are essential for accurate characterization of sharp UHPLC peaks. By maintaining sensitivity at high acquisition rates, the described approach enhances qualitative and quantitative reliability, enabling fast and robust analyses of complex samples.

References


No external literature references were provided in the source document.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Shimadzu LCMS-2020
Shimadzu LCMS-2020
2021|Shimadzu|Brochures and specifications
C146-E121G Liquid Chromatograph Mass Spectrometer LCMS-2020 LCMS-2020 —Seeing is Believing. Ultra Fast UFswitching™ Rapid 15 msec Polarity Switching Ultra Fast UFsensitivity™ Superior Sensitivity with UHPLC Ultra Fast UFscanning™ 15,000 u/sec Fast Scanning Speed Speed is Power. Greater speed. Greater sensitivity.…
Key words
ufscanning, ufscanningultra, ultraerythromycin, erythromycinufswitching, ufswitchingpositive, positivedymuron, dymuronfast, fastcid, cidbrowser, browserswitching, switchingpurification, purificationnegative, negativeion, iondisplays, displaysdata
Shimadzu Liquid Chromatograph Mass Spectrometer LCMS-8050
Shimadzu Liquid Chromatograph Mass Spectrometer LCMS-8050
2021|Shimadzu|Brochures and specifications
C146-E232G Liquid Chromatograph Mass Spectrometer LCMS-8050 Speed and Sensitivity Beyond Comparison Continuing the evolution of Shimadzu’s UF technology, Shimadzu introduces the LCMS-8050 triple quadrupole mass spectrometer, offering unparalleled measurement speeds and high-sensitivity performance. High-sensitivity quantitation delivered at high speed Multi-component…
Key words
switching, switchingufswitching, ufswitchingmrm, mrmufsensitivity, ufsensitivityspeed, speedufscanning, ufscanninghigh, highmpp, mpppolarity, polarityion, ionufsweeper, ufsweeperquantitative, quantitativesensitivity, sensitivitynegative, negativeeven
Shimadzu Liquid Chromatograph Mass Spectrometer LCMS-8050
Shimadzu Liquid Chromatograph Mass Spectrometer LCMS-8050
2019|Shimadzu|Brochures and specifications
C146-E232F Liquid Chromatograph Mass Spectrometer LCMS-8050 Speed and Sensitivity Beyond Comparison Continuing the evolution of Shimadzu's UF technology, Shimadzu introduces the LCMS-8050 triple quadrupole mass spectrometer, offering unparalleled measurement speeds and high-sensitivity performance. High-sensitivity quantitation delivered at high speed Multi-component…
Key words
switching, switchingufswitching, ufswitchingmrm, mrmufsensitivity, ufsensitivityufscanning, ufscanningspeed, speedhigh, highlabsolutions, labsolutionsufsweeper, ufsweepermpp, mppinsight, insightnexera, nexerapolarity, polarityeven, evenchromatograph
Enhancing the data quality by high-speed analysis on a single quadrupole mass spectrometer
Enhancing the data quality by high-speed analysis on a single quadrupole mass spectrometer Katsuaki KOTERASAWA1, Natsuyo ASANO1, Keisuke ISO1, Kazuo MUKAIBATAKE1, Kyoko WATANABE1, Atsuhiko TOYAMA1 1 Shimadzu Corporation, Kyoto, Japan Condition of LC and MS 1.32 sec 7,500 u/sec Table…
Key words
scan, scanpos, poscid, cidnormal, normalsettling, settlinghalf, halfneg, negwidth, widthheight, heightspectrum, spectrumspeed, speedmin, mintic, ticflow, flownecessitated
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike