Identification of a Panel of 20 Indicator Compounds in Wastewater Effluent Using Rapid, Direct Injection LC/MS/MS on the Agilent 6490 Triple Quadrupole LC/MS
Applications | 2015 | Agilent TechnologiesInstrumentation
Wastewater effluents frequently carry trace organic compounds (TOrCs) such as pharmaceuticals, personal care products, industrial chemicals, and pesticides. These contaminants pose risks to ecosystems and human health, especially as treated wastewater is increasingly reused for drinking water supplies. Reliable indicators are needed to detect and quantify the presence of wastewater-derived TOrCs and to assess treatment efficacy.
This study developed and validated a rapid LC/MS/MS method using direct aqueous injection to identify a panel of 20 indicator compounds in wastewater effluent. The selection combined a literature-based priority scoring system with a year-long monitoring campaign across four wastewater treatment plants to confirm the suitability of the chosen indicators.
The method employs 80 µL direct injection of filtered effluent into an Agilent 1260 Infinity HPLC coupled to an Agilent 6490 Triple Quadrupole mass spectrometer. Chromatography is performed on a Pursuit XRs C-8 column (100 × 2.0 mm, 3 µm) at 0.4 mL/min with a 13-minute cycle. Electrospray ionization operates in simultaneous positive and negative modes with dynamic multiple reaction monitoring and fast polarity switching. Isotopically labeled surrogate standards assess matrix effects and recovery.
The direct-injection approach eliminates laborious extraction, reduces sample volume and labeled-standard use, accelerates throughput, and enhances reproducibility. The 20-compound panel provides a robust tool for detecting wastewater intrusion, evaluating treatment performance, and tracking seasonal and geographic variations in TOrC profiles.
Future work could expand the indicator panel to emerging contaminants, integrate automated sampling and real-time data analytics, and tailor lists for specific reuse scenarios and risk assessments. Advances in high-throughput instrumentation and data processing will further improve water quality monitoring.
A streamlined LC/MS/MS method based on direct injection effectively quantified a targeted panel of 20 wastewater indicator compounds with low detection limits, high precision, and manageable matrix effects. This workflow supports comprehensive monitoring of trace organic contaminants in water reuse and source protection applications.
LC/MS, LC/MS/MS, LC/QQQ
IndustriesEnvironmental
ManufacturerAgilent Technologies
Summary
Importance of the Topic
Wastewater effluents frequently carry trace organic compounds (TOrCs) such as pharmaceuticals, personal care products, industrial chemicals, and pesticides. These contaminants pose risks to ecosystems and human health, especially as treated wastewater is increasingly reused for drinking water supplies. Reliable indicators are needed to detect and quantify the presence of wastewater-derived TOrCs and to assess treatment efficacy.
Objectives and Study Overview
This study developed and validated a rapid LC/MS/MS method using direct aqueous injection to identify a panel of 20 indicator compounds in wastewater effluent. The selection combined a literature-based priority scoring system with a year-long monitoring campaign across four wastewater treatment plants to confirm the suitability of the chosen indicators.
Methodology and Instrumentation
The method employs 80 µL direct injection of filtered effluent into an Agilent 1260 Infinity HPLC coupled to an Agilent 6490 Triple Quadrupole mass spectrometer. Chromatography is performed on a Pursuit XRs C-8 column (100 × 2.0 mm, 3 µm) at 0.4 mL/min with a 13-minute cycle. Electrospray ionization operates in simultaneous positive and negative modes with dynamic multiple reaction monitoring and fast polarity switching. Isotopically labeled surrogate standards assess matrix effects and recovery.
Used Instrumentation
- Agilent 1260 Infinity HPLC with 100 µL syringe loop
- Agilent Pursuit XRs C-8 column (100 × 2.0 mm, 3 µm)
- Agilent 6490 Triple Quadrupole LC/MS system
- 0.2 µm PES syringe filters for sample preparation
Main Results and Discussion
- Indicator Selection: From 48 candidates, 20 compounds scoring >6 on detection frequency, mean concentration, and number of studies were chosen.
- Detection Limits: Method reporting limits ranged from 3–37 ng/L for 19 compounds; sucralose exhibited an MRL of 302 ng/L. Instrument LOQs spanned 2–100 ng/L.
- Recovery and Precision: Spike recoveries (1,000 and 5,000 ng/L) were 83–138% for all but sucralose and iohexol in one matrix. Intra-day RSDs were <5% (except 12% for sucralose); inter-day RSDs were <10% for 18 compounds.
- Matrix Effects: Most analytes showed <30% signal change; iodinated contrast agents experienced up to 60% suppression. Acesulfame and sucralose exhibited signal enhancement.
- Environmental Monitoring: All 20 indicators were detected in at least three of four treatment plants over four sampling events. Some compounds (e.g., acesulfame, sucralose, TCPP, iodinated media) were consistently present at levels detectable even after dilution into drinking water sources.
Benefits and Practical Applications
The direct-injection approach eliminates laborious extraction, reduces sample volume and labeled-standard use, accelerates throughput, and enhances reproducibility. The 20-compound panel provides a robust tool for detecting wastewater intrusion, evaluating treatment performance, and tracking seasonal and geographic variations in TOrC profiles.
Future Trends and Opportunities
Future work could expand the indicator panel to emerging contaminants, integrate automated sampling and real-time data analytics, and tailor lists for specific reuse scenarios and risk assessments. Advances in high-throughput instrumentation and data processing will further improve water quality monitoring.
Conclusion
A streamlined LC/MS/MS method based on direct injection effectively quantified a targeted panel of 20 wastewater indicator compounds with low detection limits, high precision, and manageable matrix effects. This workflow supports comprehensive monitoring of trace organic contaminants in water reuse and source protection applications.
Reference
- Anumol T., et al. Rapid direct injection LC-MS/MS method for analysis of prioritized indicator compounds in wastewater effluent. Environ Sci: Water Res Technol. DOI: 10.1039/c5ew00080g (2015).
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Automated Online SPE for LC/MS/MS Analysis of Trace Organic Contaminants in Water Using the Agilent 1290 Infinity Flexible Cube Module
2015|Agilent Technologies|Applications
Automated Online SPE for LC/MS/MS Analysis of Trace Organic Contaminants in Water Using the Agilent 1290 Infinity Flexible Cube Module Application Note Environmental Authors Abstract Shane Snyder The Agilent 1290 Infinity Flexible Cube module has enabled a fully automated online…
Key words
positive, positivenegative, negativepharmaceutical, pharmaceuticalpersonal, personalcare, caretorcs, torcscube, cubenorgestrel, norgestrelonline, onlinespe, spebenzotriazole, benzotriazoleclofibric, clofibricpropylparaben, propylparabenhydrochlorothiazide, hydrochlorothiazidetriclocarban
AUTOMATED ONLINE SPE/TANDEM MS ANALYSIS OF TRACE ORGANIC CONTAMINANTS IN DRINKING WATER - Agilent Ultivo Triple Quadrupole LC/MS System
2017|Agilent Technologies|Applications
AUTOMATED ONLINE SPE/TANDEM MS ANALYSIS OF TRACE ORGANIC CONTAMINANTS IN DRINKING WATER Agilent Ultivo Triple Quadrupole LC/MS System Introduction Figure 1A. Agilent Ultivo Triple Quadrupole LC/MS. A variety of organic molecules originating from pesticides, pharmaceuticals, personal care products, and industrial…
Key words
counts, countsmin, minacquisition, acquisitionconcentration, concentrationhydrocortisone, hydrocortisonecube, cubetime, timespe, speultivo, ultivonorgestrel, norgestrelcartridge, cartridgepropylparaben, propylparabenonline, onlinegemfibrozil, gemfibrozildeet
Method for Direct Analysis of Contaminants in Surface Waters With High Accuracy and Precision Using an Agilent 6470A Triple Quadrupole LC/MS System
2016|Agilent Technologies|Applications
Method for Direct Analysis of Contaminants in Surface Waters With High Accuracy and Precision Using an Agilent 6470A Triple Quadrupole LC/MS System Application Note Environmental, Surface Water, Personal Care Products (PPCPs) Authors Abstract Jian-Zhong Li and Craig Marvin Using the…
Key words
bupropion, bupropioncotinine, cotininecontaminants, contaminantsdextrorphan, dextrorphansurface, surfaceclarithromycin, clarithromycingemfibrozil, gemfibroziltriclopyr, triclopyrvenlafaxine, venlafaxinetriclosan, triclosanlamotrigine, lamotriginegabapentin, gabapentinsucralose, sucralosemrm, mrmdiltiazem
Towards Single System for Total Water Analysis. LC-MS/MS screening of 325 PPCP Contaminants in Tap and Surface Water
2020|Shimadzu|Posters
ThP082 Towards Single System for Total Water Analysis. LC-MS/MS screening of 325 PPCP Contaminants in Tap and Surface Water. Aurore JAFFUEL1, Mikael LEVI1, Jun WATANABE1. 1 SHIMADZU Corporation, MS Business Unit, Kyoto, Japan. 1. Overview 4.1. Best Low Limits of…
Key words
acid, acidppcp, ppcpatenolol, atenololcarbamazepine, carbamazepineacetylsulfamethoxazole, acetylsulfamethoxazoleazelnipidine, azelnipidinecabozantinib, cabozantinibcapmatinib, capmatinibcerdulatinib, cerdulatinibdacomitinib, dacomitinibdesmethylnaproxene, desmethylnaproxenedibucain, dibucainfosamprenavir, fosamprenaviribrutinib, ibrutinibiobitridol