HPLC Column Comparison Screening Study for Reversed Phase Columns
Presentations | | MerckInstrumentation
The choice of reversed phase HPLC column strongly influences separation efficiency, resolution, analysis time and method robustness. A systematic screening of columns streamlines method development across applications in pharmaceuticals, environmental and industrial analysis. By understanding retention and selectivity differences early, analysts can reduce trial and error and ensure reliable, transferable methods.
This study established a structured protocol to compare selectivity and retention for multiple reversed phase stationary phases. Representative mixtures of neutrals, acidic and basic compounds were tested under controlled mobile phase conditions. Results provide a reference for analysts to select an appropriate phase as a starting point in method development.
Compound mixes included parabens, alkyl and functionalized benzenes, simple and pharmaceutical bases, and acids. Mobile phases comprised acetonitrile water for nonionizable compounds and 25 mM phosphate buffers at pH 2 and pH 7 for ionizable species. Acetonitrile concentration was adjusted to achieve k between 1 and 5. Columns tested were C8, C18, C16, Cyano, RP-Amide and fluoro phenyl HS F5. Analyses were performed on an HPLC system equipped with an automated switching valve and UV detection at 220 nm. Column dimensions were 15 cm by 4.6 mm with 5 μm particle size.
Retention factors and USP tailing factors were measured across phases. High correlation of log k values between C8 and C18 (R2 ~ 0.97) indicates similar selectivity, while lower correlation with HS F5 (R2 ~ 0.73) highlights distinct interactions. Screening data reveal phase similarities and guide decisions based on desired selectivity shifts.
The screening workflow accelerates identification of suitable stationary phases, reducing time spent on column selection. A practical example with codeine and phenacetin demonstrates how to use k versus acetonitrile and pH data to choose conditions that optimize elution order, runtime or peak spacing. This approach enhances method flexibility when sample matrices or formulations change.
Future developments may integrate high throughput screening with chemometric models and machine learning to predict optimal phases for novel analytes. Emerging stationary phase chemistries and online column switching strategies will further streamline method development and increase analytical versatility.
A systematic column screening strategy provides clear insights into retention and selectivity differences among reversed phase stationary phases. This approach enables rapid, informed column selection and supports robust method development adaptable to changing analytical requirements.
Consumables, LC columns
IndustriesManufacturerMerck
Summary
Significance of the Topic
The choice of reversed phase HPLC column strongly influences separation efficiency, resolution, analysis time and method robustness. A systematic screening of columns streamlines method development across applications in pharmaceuticals, environmental and industrial analysis. By understanding retention and selectivity differences early, analysts can reduce trial and error and ensure reliable, transferable methods.
Objectives and Overview of the Study
This study established a structured protocol to compare selectivity and retention for multiple reversed phase stationary phases. Representative mixtures of neutrals, acidic and basic compounds were tested under controlled mobile phase conditions. Results provide a reference for analysts to select an appropriate phase as a starting point in method development.
Methodology and Instrumentation
Compound mixes included parabens, alkyl and functionalized benzenes, simple and pharmaceutical bases, and acids. Mobile phases comprised acetonitrile water for nonionizable compounds and 25 mM phosphate buffers at pH 2 and pH 7 for ionizable species. Acetonitrile concentration was adjusted to achieve k between 1 and 5. Columns tested were C8, C18, C16, Cyano, RP-Amide and fluoro phenyl HS F5. Analyses were performed on an HPLC system equipped with an automated switching valve and UV detection at 220 nm. Column dimensions were 15 cm by 4.6 mm with 5 μm particle size.
Main Results and Discussion
Retention factors and USP tailing factors were measured across phases. High correlation of log k values between C8 and C18 (R2 ~ 0.97) indicates similar selectivity, while lower correlation with HS F5 (R2 ~ 0.73) highlights distinct interactions. Screening data reveal phase similarities and guide decisions based on desired selectivity shifts.
Benefits and Practical Applications of the Method
The screening workflow accelerates identification of suitable stationary phases, reducing time spent on column selection. A practical example with codeine and phenacetin demonstrates how to use k versus acetonitrile and pH data to choose conditions that optimize elution order, runtime or peak spacing. This approach enhances method flexibility when sample matrices or formulations change.
Future Trends and Potential Applications
Future developments may integrate high throughput screening with chemometric models and machine learning to predict optimal phases for novel analytes. Emerging stationary phase chemistries and online column switching strategies will further streamline method development and increase analytical versatility.
Conclusion
A systematic column screening strategy provides clear insights into retention and selectivity differences among reversed phase stationary phases. This approach enables rapid, informed column selection and supports robust method development adaptable to changing analytical requirements.
References
- Carmen T. Santasania, David S. Bell and Tracy L. Ascah-Ross, HPLC Column Comparison Screening Study for Reversed Phase Columns, Supelco T403157GIB2
- Re Discover Method Development Guide T402075A, Supelco
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
HPLC Column Comparison Screening Study for Reversed Phase Columns
|Merck|Presentations
HPLC Column Comparison Screening Study for Reversed Phase Columns Carmen T. Santasania, Eric M. Snyder, Matilal Sarker, and Tracy L. Ascah Supelco, 595 North Harrison Road, Bellefonte, PA 16823, USA T403058 2 Introduction When developing a new HPLC method, selecting…
Key words
phenacetin, phenacetincodeine, codeineselectivity, selectivityscreening, screeningstudy, studycyano, cyanophases, phasespredicted, predicteddifferences, differencesalpha, alphacolumn, columnapplying, applyingdiscovery, discoveryretention, retentioncompounds
Use Discovery to Reduce HPLC Method Development Time
2003|Merck|Technical notes
595 North Harrison Road Bellefonte, PA 16823-0048 USA Telephone 800-247-6628 • 814-359-3441 Fax 800-447-3044 • 814-359-3044 email: supelco@s ial.com sigma-aldrich.com/supelco Bulletin 937 Use Discovery to Reduce HPLC Method Development Time Following the Discovery method development protocol and using the Column…
Key words
buffer, bufferphenacetin, phenacetinnitrobenzoic, nitrobenzoicparaben, parabendiscovery, discoverycodeine, codeineacid, acidbenzene, benzenecyano, cyanofunctionalized, functionalizedphase, phasepropyl, propylnitrosobenzene, nitrosobenzenebenzyl, benzylhexafluorobenzene
Multi-Dimensional LC/MS Using Orthogonal Reversed-Phase Stationary Phases
|Merck|Presentations
Multi-Dimensional LC/MS Using Orthogonal Reversed-Phase Stationary Phases David S. Bell and Shawn R. Wyatt Supelco 595 North Harrison Road, Bellefonte, PA 16823 T403167 GIL Outline • Introduction - What is multi-dimensional chromatography? - Molecular interactions • Methods using multi-dimensional chromatography…
Key words
interactions, interactionsscan, scancoupling, couplingphases, phasesstationary, stationarydimensional, dimensionalphase, phasecomplementary, complementarydistinctly, distinctlymulti, multiionic, ionicpolar, polartic, ticsingle, singlegalactosidase
Troubleshooting Liquid Chromatographic Separations: Avoiding Common Issues through Informed Method Development Practices
2014|Merck|Presentations
Troubleshooting Liquid Chromatographic Separations: Avoiding Common Issues through Informed Method Development Practices Dave S. Bell Supelco, Div. of Sigma-Aldrich Bellefonte, PA 16823 USA sigma-aldrich.com/analytical T414096 © 2014 Sigma-Aldrich Co. All rights reserved. Objective •Most of the difficult problems encountered are…
Key words
rightsall, rightsallaldrich, aldrichsigma, sigmainteractions, interactionsphases, phasesselectivity, selectivitypolar, polarstationary, stationaryavailability, availabilitysilanol, silanolstrong, strongepg, epgmoderate, moderateruggedness, ruggednessphase