Sensitive Detection of Low-Level Byproducts in a Small Molecule Reaction by the Agilent 1260 Infinity II Prime Online LC/MS System

Applications | 2021 | Agilent TechnologiesInstrumentation
HPLC, LC/MS, LC/SQ
Industries
Pharma & Biopharma
Manufacturer
Agilent Technologies

Summary

Importance of Topic


In small-molecule pharmaceutical synthesis, real-time monitoring of reactants, products and trace byproducts is essential to ensure process control, product quality and to minimize costly downstream purification steps. Sensitive detection of low-level impurities can prevent contamination of the active pharmaceutical ingredient and improve overall process efficiency.

Objectives and Study Overview


This application note describes the integration of the Agilent 1260 Infinity II Prime Online LC/MS System with an MSD iQ single quadrupole detector and the Agilent Online LC Monitoring Software to perform automated, inline sampling and analysis of an aldol condensation reaction between p-anisaldehyde and acetone. The aim is to demonstrate the system’s capability to:
  • Draw and dilute samples directly from the reaction vessel at defined intervals
  • Quantify reactant consumption and product formation in near real time
  • Detect and trend low-level byproducts
  • Generate alerts when reactant levels drop below user-defined limits

Methodology and Used Instrumentation


  • Reaction Conditions: p-Anisaldehyde in acetone/water (2:1 v/v), initiated with NaOH, stirred at room temperature
  • Sample Handling: Agilent 1260 Infinity II Online Sample Manager with automated dilution/quenching into deep-well plates
  • Chromatography: Agilent InfinityLab Poroshell 120 EC-C18 column (2.1×30 mm, 1.9 µm), gradient from 40% to 90% acetonitrile + 0.1% formic acid at 1.3 mL/min, 45 °C
  • Detection: Agilent MSD iQ single quadrupole in positive ESI, SIM mode at m/z 133 (p-anisaldehyde), 177 (z-anisylidene acetone) and 295 (double aldol byproduct); DAD at 290 nm
  • Software Control: Agilent Online LC Monitoring Software scheduling sample draws, dilutions, LC/MS runs and live trending

Main Results and Discussion


Relative quantification was achieved by a one-point calibration on the initial sample to normalize SIM peak areas. The p-anisaldehyde trace declined steadily, while the main product trace increased, as shown in trending plots. A lower-limit alarm was set at 5% residual p-anisaldehyde to signal reaction completion. The byproduct at m/z 295 was first detected after nine minutes and trended upward thereafter. Peak area and height data confirmed conversion from educt to product and allowed estimation of byproduct levels without manual sampling or offline analysis.

Benefits and Practical Applications of the Method


  • Improved sensitivity and selectivity over UV-only monitoring
  • Automated sampling and dilution reduce operator intervention and error
  • Real-time feedback enables dynamic process control and faster decision making
  • Integrated software provides a single interface for experiment setup, control and data visualization
  • Archived samples in deep-well plates enable secondary confirmation by other analytical techniques

Future Trends and Opportunities


As inline LC/MS monitoring becomes more widespread, future developments may include:
  • Integration with continuous flow reactors for full process automation
  • Machine learning algorithms for pattern recognition and predictive control
  • Multiplexed SIM channels for simultaneous monitoring of multiple byproducts
  • Scale-up studies linking lab-scale feedback to pilot and production scales
  • Combining high-resolution MS detectors for detailed impurity profiling

Conclusion


The Agilent 1260 Infinity II Prime Online LC/MS system with MSD iQ demonstrates robust, sensitive and automated monitoring of a small-molecule aldol reaction. It provides reliable quantification of reactant depletion, product formation and trace byproducts, enabling tighter process control, reduced purification burden and streamlined data management in pharmaceutical R&D and manufacturing.

References


1. Viviano M. et al. A Scalable Two-Step Continuous Flow Synthesis of Nabumetone and Related 4-Aryl-2-butanones. Org. Process Res. Dev. 2011, 15, 858–870.
2. Automated Reaction Monitoring by the Agilent 1260 Infinity II Prime Online LC system. Agilent Technologies application note, publication number 5994-3980EN.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Automated “Online“ LC-MS Small Molecule Reaction Monitoring with a Single Quadrupole MS
Poster Reprint ASMS 2024 Poster number WP 483 Automated “Online“ LC-MS Small Molecule Reaction Monitoring with a Single Quadrupole MS Sebastian Weinig-Berger1 , Russel Burge2, Edgar Neagele1, Andreas Mielcarek1 1Agilent Technologies Deutschland GmbH, Germany Waldbronn, BW, 2Agilent Technologies, Inc., Santa…
Key words
educt, eductappearing, appearingbyproduct, byproductreaction, reactionaldol, aldoldilution, dilutionsampling, samplingejection, ejectionsim, simagilent, agilentcondensation, condensationreactor, reactorundershot, undershotcourse, courserequisition
Monitoring of Trace Impurities Using the Agilent 1260 Infinity II Prime Online LC System with High Dynamic Range-DAD
Application Note Small Molecule Pharma Monitoring of Trace Impurities Using the Agilent 1260 Infinity II Prime Online LC System with High Dynamic Range-DAD ×103 4.0 3.6 3.2 2.8 2.4 2.0 1.6 1.2 0.8 0.4 0 0.360 0.260 0.837 0.873 0.1…
Key words
anisaldehyde, anisaldehydeanisylidene, anisylidenebyproduct, byproducteduct, eductdad, dadhdr, hdronline, onlinedynamic, dynamicreaction, reactionacetone, acetonemolecule, moleculeconcentrated, concentrateddiode, diodehigh, highsmall
Automated Reaction Monitoring by the Agilent 1260 Infinity II Prime Online LC
Application Note Biopharma and Pharma Automated Reaction Monitoring by the Agilent 1260 Infinity II Prime Online LC Sampling from the reactor, dilution, quenching, and analysis Authors Christian Lautz F. Hoffmann-La Roche Ltd. Basel, Switzerland Edgar Naegele, Daniel Kutscher, and Sam…
Key words
online, onlineeduct, eductreaction, reactionanisylidene, anisylideneanisaldehyde, anisaldehydeacetone, acetonemanager, managermonitoring, monitoringinfinity, infinitydilution, dilutiondeclining, decliningsampling, samplingmolecule, moleculequantification, quantificationsample
Email Notifications from Your Online LC
Email Notifications from Your Online LC
2023|Agilent Technologies|Technical notes
Application Note Biopharma/Pharma Email Notifications from Your Online LC Messages about instrumental and experimental status sent by Agilent Online LC Monitoring Software for time-saving decisions and unattended operation Author Abstract Edgar Naegele Agilent Technologies, Inc. This application note demonstrates the…
Key words
notification, notificationreaction, reactionemail, emailnotifications, notificationsonline, onlinetrending, trendingsampling, samplingmonitoring, monitoringband, banddilution, dilutionemails, emailspredefined, predefinededuct, eductreactor, reactorunattended
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike