High-Sensitivity Analysis of Cortisol in Human Blood Serum Using LCMS-8060NX Triple Quadrupole Mass Spectrometer
Applications | 2021 | ShimadzuInstrumentation
Cortisol quantification in human serum is critical for diagnosing adrenal and pituitary function and monitoring metabolic and inflammatory conditions.
This study demonstrates a high-sensitivity LC/MS/MS method using Shimadzu LCMS-8060NX for quantifying cortisol at picogram-per-milliliter levels in human blood serum.
Linearity was excellent (R2 > 0.999) across 2.5–100000 pg/mL with accuracy within 80–110%. The limit of quantitation reached 2.5 pg/mL with clear MRM peaks. Spiked desteroidized serum yielded 92 pg/mL at a 100 pg/mL spike, confirming method accuracy.
Integration with automated sample preparation, expansion to multiplex steroid profiling, and adaptation to point-of-care platforms may further enhance clinical relevance and throughput.
The LCMS-8060NX method enables reliable quantification of cortisol down to 2.5 pg/mL in human serum with high selectivity and robustness, representing a valuable tool for endocrinology and biomarker research.
LC/MS, LC/MS/MS, LC/QQQ
IndustriesClinical Research
ManufacturerShimadzu
Summary
Significance of the Topic
Cortisol quantification in human serum is critical for diagnosing adrenal and pituitary function and monitoring metabolic and inflammatory conditions.
Aims and Study Overview
This study demonstrates a high-sensitivity LC/MS/MS method using Shimadzu LCMS-8060NX for quantifying cortisol at picogram-per-milliliter levels in human blood serum.
Methodology and Sample Preparation
- Reference standard serial dilutions (2.5–100000 pg/mL) in 50% methanol.
- SPE cleanup using EVOLUTE EXPRESS ABN plates with acidified serum loading, multi-step washing, and elution.
- Reconstitution in 50% methanol.
Used Instrumentation
- Shimadzu Nexera X3 HPLC with Shim-pack Scepter C18 column (50×2.1 mm, 3 μm).
- LCMS-8060NX triple quadrupole with IonFocus™ interface, ESI positive mode.
Analysis Conditions
- Mobile phases: 0.2 mM NH4F in water and methanol with gradient from 50% to 95% B.
- Flow rate: 0.3 mL/min; injection volume: 2 μL.
- MRM transitions: cortisol (363.0→121.1) and internal standard D4 (367.1→121.1).
Main Results and Discussion
Linearity was excellent (R2 > 0.999) across 2.5–100000 pg/mL with accuracy within 80–110%. The limit of quantitation reached 2.5 pg/mL with clear MRM peaks. Spiked desteroidized serum yielded 92 pg/mL at a 100 pg/mL spike, confirming method accuracy.
Benefits and Practical Applications
- Ultra-sensitive cortisol detection for clinical and research settings.
- Reduced MS contamination via IonFocus, enhancing instrument robustness.
- Comprehensive workflow including sample preparation and analysis.
Future Trends and Potential Applications
Integration with automated sample preparation, expansion to multiplex steroid profiling, and adaptation to point-of-care platforms may further enhance clinical relevance and throughput.
Conclusion
The LCMS-8060NX method enables reliable quantification of cortisol down to 2.5 pg/mL in human serum with high selectivity and robustness, representing a valuable tool for endocrinology and biomarker research.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
High Sensitivity Analysis of Aldosterone in Human Serum Using LCMS™-8060NX
2020|Shimadzu|Applications
Application News No. C228 LC-MS High Sensitivity Analysis of Aldosterone in Human Serum Using LCMS™-8060NX This report introduces an example of high sensitivity analysis of aldosterone in serum using LCMS-8060NX, a triple quadrupole mass spectrometer. Calibration curves prepared using standard…
Key words
aldosterone, aldosteroneserum, serumsteroid, steroidmedical, medicalintroduces, introducesradioimmunoassay, radioimmunoassayeia, eiaria, riaflow, flowionfocus, ionfocussensitivity, sensitivityarea, areanews, newsreport, reportshowed
High Sensitivity Analysis of Testosterone in Human Serum Using LCMS™-8060NX
2020|Shimadzu|Applications
Application News No. C227 LC-MS High Sensitivity Analysis of Testosterone in Human Serum Using LCMS™-8060NX This report introduces an example of high sensitivity analysis of testosterone in serum using LCMS8060NX, a triple quadrupole mass spectrometer. Calibration curves prepared using standard…
Key words
testosterone, testosteroneserum, serumsteroid, steroidmedical, medicalintroduces, introducescalibration, calibrationcaused, causedflow, flowbones, bonesionfocus, ionfocusrea, reasensitivity, sensitivitynews, newsreport, reportevaporating
Quantitation of Endogenous Steroids in Serum Using Dried Blood Spot Serum Separator Card and Triple Quadrupole Mass Spectrometry
2023|Shimadzu|Posters
Quantitation of Endogenous Steroids in Serum Using Dried Blood Spot Serum Separator Card and Triple Quadrupole Mass Spectrometry Vikki Johnson1; Nicholas Chestara2; Yoshiyuki Okamura1 1Shimadzu Scientific Instruments, Columbia, MD 2DPX Technologies, La Jolla, CA 1. Overview This poster utilized LC-MS/MS…
Key words
dheas, dheascortsol, cortsolprogesterone, progesteronetestosterone, testosteronehighqc, highqclowqc, lowqcmidqc, midqcsteroids, steroidssample, samplecortisol, cortisolendogenous, endogenousdried, driedcards, cardsblood, bloodarea
Quantitation of endogenous steroids in serum using plasma separation cards and triple quadruple mass spectrometry
2024|Shimadzu|Posters
WP 193 Quantitation of endogenous steroids in serum using plasma separation cards and triple quadruple mass spectrometry Vikki Johnson1, Nicholas Chestara2, MayaRae Mugosa3, Jefferson Kinney3, Yoshiyuki Okamura1 (1) Shimadzu Scientific Instruments, Columbia, MD; (2) DPX Technologies, La Jolla, CA; (3)…
Key words
steroids, steroidstelimmune, telimmuneplasma, plasmatestosterone, testosteronecorrelation, correlationdpx, dpxcards, cardsimmunoassay, immunoassaywere, werearea, areaelisa, elisaincubated, incubatedquantitation, quantitationratio, ratiocard