High Sensitivity Analysis of Aldosterone in Human Serum Using LCMS™-8060NX

Applications | 2020 | ShimadzuInstrumentation
LC/MS, LC/MS/MS, LC/QQQ
Industries
Clinical Research
Manufacturer
Shimadzu

Summary

Significance of the Topic


Aldosterone is a key mineralocorticoid regulating electrolyte balance, blood volume and pressure. Accurate measurement of its serum levels is critical in diagnosing conditions such as primary hyperaldosteronism, diabetic nephropathy and interstitial nephritis. Conventional immunoassays may overestimate concentrations due to cross-reactivity with structurally similar steroids. High-selectivity LC-MS/MS methods overcome these limitations and enable reliable quantitation at ultralow levels.

Study Objectives and Overview


This work demonstrates a high-sensitivity assay for aldosterone in human serum using the Shimadzu LCMS-8060NX triple quadrupole mass spectrometer. Calibration standards ranged from 0.5 to 10 000 pg/mL. Key goals included establishing linearity, precision and accuracy at the lower limit of quantification in both neat and serum-spiked samples.

Sample Preparation


Standard solutions were serially diluted in 50 % methanol to yield 0.5-10 000 pg/mL calibration levels. Steroid-free human serum was spiked at 1 pg/mL aldosterone. Each aliquot (100 µL) was combined with internal standard, extracted with hexane:ethyl acetate (9:1), centrifuged for 10 min, the organic layer evaporated and residues reconstituted in 50 % methanol for injection.

Analytical Methodology


  • HPLC: Shim-pack Scepter C18-120 (50 × 2.1 mm, 1.9 µm), mobile phase A = 0.05 mM NH4F in water, B = methanol; gradient elution; flow rate 0.3 mL/min; injection volume 25 µL.
  • MS: ESI positive mode, MRM transitions m/z 361.20→343.30 for aldosterone and m/z 365.25→346.25 for the D4 internal standard; interface voltage 1 kV; ion focus 3.5 kV; nebulizing gas 3 L/min; drying gas 10 L/min; heating gas 5 L/min; DL temp 250 °C; block heater 300 °C; interface 400 °C.

Results and Discussion


The calibration curve exhibited excellent linearity over 0.5-10 000 pg/mL (R2 = 0.9998). The %RSD at the LOQ (0.5 pg/mL) was 3.96 %, with accuracy across calibration points between 95.6 % and 109.8 %. Serum samples spiked at 1 pg/mL yielded a mean recovery of 106 % (n=3). Representative MRM chromatograms confirmed clear peak separation and signal-to-noise ratios suitable for trace quantitation.

Practical Benefits and Applications


This LC-MS/MS assay delivers high sensitivity and selectivity for aldosterone, enabling detection at sub-pg/mL levels and robust performance in complex serum matrices. It is well suited for clinical research, diagnostic support and quality control in pharmaceutical or clinical laboratories.

Future Trends and Applications


  • Integration into multiplex steroid panels for comprehensive endocrine profiling.
  • Automation and high-throughput workflows to support routine clinical diagnostics.
  • Application to biomarker discovery and metabolic studies requiring ultra-trace quantitation.

Conclusion


The Shimadzu LCMS-8060NX method offers a reliable, sensitive and accurate platform for aldosterone quantitation in human serum across a broad dynamic range. Its performance in precision, linearity and recovery underscores its applicability in both research and diagnostic settings.

Instrumentation Used


  • Shimadzu LCMS-8060NX triple quadrupole mass spectrometer
  • Nexera X3 UHPLC system
  • Shim-pack Scepter C18-120 column (50 × 2.1 mm, 1.9 µm)

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
High Sensitivity Analysis of Testosterone in Human Serum Using LCMS™-8060NX
Application News No. C227 LC-MS High Sensitivity Analysis of Testosterone in Human Serum Using LCMS™-8060NX This report introduces an example of high sensitivity analysis of testosterone in serum using LCMS8060NX, a triple quadrupole mass spectrometer. Calibration curves prepared using standard…
Key words
testosterone, testosteroneserum, serumsteroid, steroidmedical, medicalintroduces, introducescalibration, calibrationcaused, causedflow, flowbones, bonesionfocus, ionfocussensitivity, sensitivityrea, reanews, newsreport, reportevaporating
Quantitative Analysis of Steroid Hormones Using Triple Quadrupole LC/MS/MS
LAAN-A-LM-E087 Application News C116 Liquid Chromatography Mass Spectrometry Quantitative Analysis of Steroid Hormones Using Triple Quadrupole LC/MS/MS No. Steroid hormones are heavily involved with the control of metabolism, neurotransmission, and intracellular signaling, playing critical roles in the proper functioning of…
Key words
dheas, dheascortisol, cortisolprogesterone, progesteronealdosterone, aldosteroneandrostenedione, androstenedionecoricosterone, coricosteronedhea, dheatestosterone, testosteroneestrone, estroneratio, ratiosteroid, steroidarea, areacorticosterone, corticosteroneneg, negpos
High-Sensitivity Analysis of Cortisol in Human Blood Serum Using LCMS-8060NX Triple Quadrupole Mass Spectrometer
LCMS™-8060NX High Performance Liquid Chromatograph Mass Spectrometer Application News High-Sensitivity Analysis of Cortisol in Human Blood Serum Using LCMS-8060NX Triple Quadrupole Mass Spectrometer N. Kato User Benefits  Quantitative analysis of cortisol with an extremely low concentration value in blood…
Key words
cortisol, cortisoldesteroidized, desteroidizedserum, serumblood, bloodimmunoassay, immunoassayanalysis, analysiseclia, ecliahuman, humanspiked, spikedpituitary, pituitaryquantitative, quantitativemethod, methodluminescence, luminescenceadrenal, adrenalflow
Simultaneous Analysis of 20 Steroid Hormones and High-Sensitivity Analysis of Estrogens
LCMS-8060 Liquid Chromatograph Mass Spectrometer Simultaneous Analysis of 20 Steroid Hormones and High-Sensitivity Analysis of Estrogens Application News Tomoya Niinae, Yohei Arao, and Yutaka Umakoshi User Benefits  Simultaneous analysis of 20 steroid hormones in 15 minutes  Derivatization of…
Key words
derivatized, derivatizedsteroid, steroidhormones, hormonesestrogen, estrogenestrogens, estrogensserum, serumestriol, estriolestrone, estroneestradiol, estradiolpackage, packagemin, minspiked, spikedhormone, hormonenews, newssimultaneous
Other projects
GCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike