High-Resolution Cation-Exchange Alternative to Peptide Mapping for Protein ID and QA/QC
Applications | 2016 | Thermo Fisher ScientificInstrumentation
Consumables, HPLC, LC columns
IndustriesProteomics
ManufacturerThermo Fisher Scientific
Summary
Importance of the Topic
Peptide mapping is a cornerstone technique for confirming protein identity and ensuring quality control of biopharmaceuticals. Traditional reversed-phase separations excel in many applications, but high-resolution cation-exchange chromatography offers complementary selectivity based on peptide charge and can reveal additional proteoforms or impurities.Study Objectives and Overview
This work aimed to establish a robust, high-resolution cation-exchange peptide mapping method as an alternative to reversed-phase approaches. Using Thermo Scientific ProPac SCX-10 (10 µm) and the next-generation MabPac SCX-10 (3 µm) columns, the study evaluated separation performance for a model myoglobin tryptic digest and a synthetic peptide with byproducts. Comparisons were made against reversed-phase columns under various mobile phase conditions to assess resolution, analysis time, and peak capacity.Methodology and Used Instrumentation
A systematic workflow was applied:- Sample preparation: reduction with DTT, alkylation with iodoacetamide, dialysis, and overnight tryptic digestion of equine myoglobin.
- Synthetic peptide: Acetyl-YNIQKESTLPLVLRLRGG-amide prepared with byproducts, calculated pI 11.4.
- Mobile phases: 20 mM triethylamine phosphate (TEAP) at pH 2.0 or 3.9 with 50% acetonitrile (A) and 100 mM sodium perchlorate in phase A (B).
- Gradient programs: linear increases of phase B at flow rates from 0.7 to 1.4 mL/min and shortened run times via higher flow rates.
- Thermo Scientific Dionex UltiMate 3000 ×2 Biocompatible Analytical LC system (DGP-3600BM pump, SRD-3600 degasser, WPS-3000TBFC autosampler with fraction collection, TCC-3000SD column compartment, DAD-3000 detector).
- Thermo Scientific Dionex Chromeleon CDS v6.80 or higher.
- Thermo Scientific Orion 2-Star Benchtop pH meter.
Main Results and Discussion
- High-resolution separation of myoglobin peptides was achieved across pH and flow-rate conditions, with accelerated gradients maintaining peak integrity (Figure 1).
- Synthetic peptide separations demonstrated that higher pH (3.9) improved resolution for basic sequences compared to pH 2.0 (Figure 2).
- Comparative analyses against reversed-phase columns (Acclaim PA2 and Acclaim 300 C18) showed that ProPac SCX-10 delivers comparable or superior separation of closely eluting peptides when using formic acid or TFA as modifiers (Figures 3 and 4).
- Peak capacity calculations indicated that ion-exchange columns reach values similar to reversed-phase C18 columns (Table 1), and the MabPac SCX-10 (3 µm) further increased efficiency and resolution while reducing analysis time below 10 minutes (Figures 5 and 6).
Benefits and Practical Applications
- Orthogonal separation mechanism to reversed-phase chromatography enhances detection of charge variants and hydrophilic or hydrophobic differences.
- Fast method development by adjusting pH, salt concentration, and flow rate without sacrificing resolution.
- Applicable in QA/QC workflows for biotherapeutics, peptide impurity profiling, and orthogonal method validation in regulated environments.
Future Trends and Potential Applications
- Coupling high-resolution cation-exchange peptide mapping with mass spectrometry for 2D-LC workflows in proteomics.
- Integration of next-generation particle technologies to further boost throughput and sensitivity.
- Automation and high-throughput screening of peptide libraries or therapeutic antibody digests.
- Extension to other post-translational modification analyses and complex mixture profiling.
Conclusion
High-resolution cation-exchange chromatography on ProPac and MabPac SCX-10 columns offers a powerful alternative or complement to traditional reversed-phase peptide mapping. The methods deliver excellent resolution, flexible method speed, and robust peak capacity, making them well suited for protein identification, impurity profiling, and advanced QC applications.References
- Dionex Application Note 521: Automated 2D LC Coupled to ESI-MS/MS for the Analysis of Complex Peptide Samples. Sunnyvale, CA, 2002.
- Dionex Application Note 126: Determination of Hemoglobin Variants by Cation-Exchange Chromatography. Sunnyvale, CA, 2007.
- Dionex Application Update 183: Separation of Peptides from Enzymatic Digestion of Different Acclaim Columns: A Comparative Study. Sunnyvale, CA, 2011.
- Dionex Technical Note 74: High Peak Capacity Nano LC Peptide Separations Using Long Packed Columns. Sunnyvale, CA, 2009.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
2019/2021 Chromatography Consumables Catalog - Bio LC columns and accessories
2019|Thermo Fisher Scientific|Others
Connected chromatography solutions 2019/2021 Chromatography Consumables Catalog The collective power of chromatography Expect reproducible results with sample prep, columns and vials Maximizing your chromatography productivity and achieving reproducible results requires optimizing the whole workflow from sample to knowledge. By choosing…
Key words
columns, columnsmabpac, mabpacsize, sizecolumn, columnlength, lengthparticle, particlebiolc, biolcproswift, proswiftprotein, proteinpropac, propacvariants, variantsexchange, exchangeresolution, resolutionbiobasic, biobasicmonolith
Thermo Bio LC Columns - See your protein therapeutics in high resolution
2020|Thermo Fisher Scientific|Brochures and specifications
See your protein therapeutics in high resolution Bio LC Columns Contents 2 Introduction to monoclonal antibody (mAb) characterization 3 Affinity titer determination 4 Aggregate separation 5 Charge variant analysis 6 Oxidation variant analysis 11 Antibody drug conjugates (ADCs) analysis 12…
Key words
mabpac, mabpacmab, mabcolumn, columncolumns, columnshplc, hplcformat, formatsize, sizescfc, scfcmau, maumin, minparticle, particlemmae, mmaecharge, chargeantibody, antibodypropac
BioLC columns and accessories (Chromatography consumables catalog)
2022|Thermo Fisher Scientific|Brochures and specifications
Chromatography consumables Connected chromatography solutions Chromatography consumables catalog Comprehensive products to support your chromatography workflows Sample preparation solutions Sample handling solutions Low-flow LC columns and accessories BioLC columns and accessories LC columns and accessories GC columns and accessories • EASY-Spray™…
Key words
columns, columnsbiolc, biolcmabpac, mabpacaccesories, accesoriessize, sizephase, phasemobile, mobilereading, readingcolumn, columnbuffer, buffervariant, variantpore, porehplc, hplcmab, mablength
BioLC columns and accessories
2024|Thermo Fisher Scientific|Brochures and specifications
BioLC columns Connected chromatography solutions BioLC columns and accessories Introduction Your complete tool kit Thermo Fisher Scientific has innovative Thermo Scientific™ BioLC columns for each step of your therapeutic protein characterization, no matter how challenging your separation. Here is just…
Key words