Waters BioResolve SEC mAb Guard and Columns - CARE AND USE MANUAL

Manuals | 2020 | WatersInstrumentation
Consumables, LC columns
Industries
Manufacturer
Waters

Summary

Importance of the Topic


Monoclonal antibodies are essential biotherapeutics whose safety and efficacy depend on tight control of high molecular weight aggregates and low molecular weight fragments. Native size-exclusion chromatography (SEC) remains the gold standard for assessing non-covalent aggregation and fragmentation, enabling robust quality control during development and manufacturing.

Objectives and Study Overview


This manual introduces Waters BioResolve SEC mAb Columns and Guards, designed under cGMP and ISO 9001 protocols for reproducible quantitation of mAb monomers (~150 kDa), aggregates (≥300 kDa), and fragments (≤100 kDa). Key goals include optimizing column selection based on system dispersion, standardizing performance with a mAb Size Variant Standard, and extending column lifetime.

Methodology and Instrumentation


  • System Dispersion Measurement: Use the 5-Sigma bandspread method with a caffeine test (0.16 mg/mL, 0.5 µL injection) to determine extra-column volume and select the appropriate column ID and length.
  • Chromatographic Conditions: Mobile phase of 50 mM sodium phosphate pH 7.0 with 200 mM KCl; flow rates of 0.2 mL/min for 4.6 mm ID and 0.575 mL/min for 7.8 mm ID; column temperature 35 °C; UV detection at 280 nm.
  • Instrumentation: ACQUITY UPLC H-Class Bio, ACQUITY UPLC Arc Bio, and Alliance HPLC systems; BioResolve SEC mAb Columns (200 Å, 2.5 µm) in 4.6×150, 4.6×300, 7.8×150, and 7.8×300 mm formats; matching BioResolve SEC mAb Guards; TUV and 2489 UV/Vis detectors; cooled sample managers.
  • Sample Standard Preparation: The Waters mAb Size Variant Standard contains lyophilized NISTmAb RM 8671 plus non-reduced IdeS fragments. Reconstitute to 1–2 mg/mL, vortex upright and inverted, and inject 1.8–10 µL depending on column format.
  • Equilibration and Performance Verification: Flush with at least 10 column volumes of buffer at incremental flow-rate increases (0.1 mL/min); benchmark using uracil or mAb standard and compare to Column Test Report metrics (retention, 5-Sigma efficiency, USP tailing).
  • Maintenance: Prepare buffers with high-purity water and 0.22 µm filtration; avoid sinker filters; clean columns with low-pH or organic flushes; store in shipping solvent (10% acetonitrile in 25 mM phosphate pH 7.0 + 100 mM KCl) at 4 °C for short-term or –20 °C for long-term.

Main Results and Discussion


  • Resolution strongly correlates with system dispersion and column dimensions. High-dispersion systems (≈49 µL) require 7.8×300 mm columns for USP resolution >2 and start p/v >3 between monomer and 100 kDa fragment.
  • Low-dispersion UPLC systems (≈10 µL) achieve adequate separation on shorter columns (4.6×150 mm, Rs~1.9) but benefit from longer IDs for higher efficiency (up to 18,000 plates on 7.8×300 mm).
  • BioResolve SEC mAb Guards capture particulates and excipients, extending column life with only minor impact on separation (Rs reduction from 3.2 to 2.9).
  • Column specifications (pH 2.5–8.0, 4–60 °C, salt 100–500 mM, max backpressure 4500/6500 psi) support broad method flexibility.

Benefits and Practical Applications


  • Enables precise quantitation of mAb aggregates and fragments critical for regulatory compliance and formulation screening.
  • Offers column formats tailored to diverse LC platforms from legacy HPLC to modern UPLC.
  • Standardized mAb Size Variant Standard ensures batch-to-batch column consistency and facilitates troubleshooting.
  • Guard columns reduce downtime and replacement costs by trapping contaminants upstream of the analytical column.

Future Trends and Potential Applications


Advances in reducing extra-column dispersion, integration with multi-detector and mass spectrometry platforms, and development of phase-modified stationary supports will drive higher throughput and deeper characterization of complex biotherapeutics such as antibody–drug conjugates and multispecific antibodies.

Conclusion


Waters BioResolve SEC mAb Columns and Guards deliver a robust, reproducible SEC solution for biotherapeutic analysis. Careful evaluation of LC system dispersion, methodical column selection, rigorous performance verification, and diligent maintenance are essential to achieving optimal resolution and extending column life.

References


Waters Application Note Impact of LC System Dispersion on the Size-Exclusion Chromatography Analysis of Monoclonal IgG Antibody Aggregates and Fragments: Selecting the Optimal Column for Your Method, p/n: 720006336EN
Waters Application Note Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates, p/n: 720004076EN
Waters Size-Exclusion Chromatography Optimization Guide, p/n: 720006067EN
Waters Care and Use Manual for mAb Size Variant Standard, p/n: 720006811EN

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
High Resolution Size-Exclusion Chromatography Separations of mAb Aggregates, Monomers, and Fragments using BioResolve SEC mAb Columns on UPLC, UHPLC, and HPLC Chromatography Systems
Application Note High Resolution Size-Exclusion Chromatography Separations of mAb Aggregates, Monomers, and Fragments using BioResolve SEC mAb Columns on UPLC, UHPLC, and HPLC Chromatography Systems Pamela C. Iraneta, Stephan M. Koza Waters Corporation For research use only. Not for use…
Key words
mab, mabdispersion, dispersionvariant, variantbioresolve, bioresolvesec, secsize, sizebio, bioacquity, acquitysystem, systemwaters, watersuplc, uplccolumn, columnresolution, resolutionseparations, separationscolumns
Waters XBridge Premier Protein SEC 250 Å, 2.5 μm Columns - CARE AND USE MANUAL
[ CARE AND USE MANUAL ] XBridge Premier Protein SEC 250 Å, 2.5 µm Columns CONTENTS I. INTRODUCTION Thank you for choosing a Waters™ SEC Column. The XBridge™ I. INTRODUCTION II. SEC ANALYSIS OF PROTEIN AGGREGATES, MONOMERS, AND FRAGMENTS…
Key words
column, columnmab, mabsec, seccare, carevariant, variantuse, useprotein, proteinmanual, manualpremier, premierionhance, ionhancemonomer, monomerhmws, hmwssize, sizecan, canrecommended
Waters ACQUITY Premier Protein SEC 250 Å, 1.7 μm Columns - CARE AND USE MANUAL
[ CARE AND USE MANUAL ] ACQUITY Premier Protein SEC 250 Å, 1.7 µm Columns CONTENTS I. INTRODUCTION Thank you for choosing a Waters™ SEC Column. The ACQUITY™ I. INTRODUCTION II. SEC ANALYSIS OF PROTEIN AGGREGATES, MONOMERS, AND FRAGMENTS…
Key words
column, columnmab, mabcare, carevariant, variantsec, secuse, usemanual, manualpremier, premierprotein, proteinionhance, ionhancemonomer, monomersize, sizehmws, hmwswaters, watersrecommended
SIZE-EXCLUSION CHROMATOGRAPHY (SEC) OF PEPTIDES, PROTEINS, AND AAVs
[ APPLICATION NOTEBOOK ] SIZE-EXCLUSION CHROMATOGRAPHY (SEC) OF PEPTIDES, PROTEINS, AND AAVs [ SIZE-EXCLUSION CHROMATOGRAPHY ] [ SIZE-EXCLUSION CHROMATOGRAPHY ] Introduction Gel-permeation chromatography (GPC), size-exclusion chromatography (SEC), and gel-filtration chromatography (GFC) are commonly used terms to describe a liquid chromatographic…
Key words
sec, secexclusion, exclusionsize, sizeuplc, uplcacquity, acquitychromatography, chromatographyprotein, proteinaggregates, aggregateshplc, hplcdispersion, dispersioncolumn, columnmonomer, monomerigg, iggantibody, antibodybio
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike