Separation of Urea Herbicides Using a Core Enhanced Technology Accucore HPLC Column
Applications | 2011 | Thermo Fisher ScientificInstrumentation
Consumables, HPLC, LC columns
IndustriesFood & Agriculture
ManufacturerThermo Fisher Scientific
Summary
Importance of the Topic
The analysis of urea herbicides plays a vital role in environmental monitoring and public health protection. Urea-based herbicides are widely applied in agricultural and non-agricultural settings to control weeds by inhibiting photosynthesis. Due to their persistence in the environment, trace levels can enter drinking water sources, posing risks of acute and chronic toxicity in humans and wildlife.Study Objectives and Overview
This work illustrates the rapid and efficient separation of six common urea herbicides using a core–shell particle HPLC column. The goal was to achieve baseline resolution of monuron, metoxuron, linuron, diuron, terbuthiuron, and chlortoluron within a short analysis time, enhancing throughput for routine quality control and environmental screening.Methodology and Instrumentation
The method employed a Thermo Scientific Accucore RP-MS column (2.6 µm, 100 × 2.1 mm) featuring Core Enhanced Technology. Sample preparation involved diluting a 1 mg/mL standard mixture to 20 µg/mL in water after initial dissolution in 1:1 acetonitrile/water. Chromatographic conditions included 240 nm UV detection, 25 °C column temperature, a 0.43 mL/min flow rate, and a gradient from 35 % to 60 % acetonitrile over 3.2 minutes.Used Instrumentation
- Accucore RP-MS 2.6 µm, 100 × 2.1 mm column
- Accela HPLC/UHPLC system
- UV detector set at 240 nm
- Standard glassware and certified HPLC solvents
Key Results and Discussion
The optimized method achieved complete separation of the six urea herbicides in under 3.5 minutes with sharp, symmetrical peaks and excellent reproducibility. Backpressure remained low thanks to the 2.6 µm core–shell particles. Peak elution order was monuron, metoxuron, linuron, diuron, terbuthiuron, and chlortoluron with retention times ranging from 1.12 to 3.36 minutes.Benefits and Practical Applications
This rapid HPLC approach offers significant advantages for routine analysis in environmental laboratories and agricultural quality control. High throughput, robust performance, and minimal solvent usage enhance cost-effectiveness. The reliable retention time reproducibility supports compliance with regulatory monitoring requirements.Future Trends and Opportunities
Advancements in core–shell and sub-2 µm particle technologies will likely further reduce analysis times and improve separation efficiency. Coupling with high-resolution mass spectrometry can enable ultra-trace analysis for emerging contaminants. Automation and online sample preparation will expand applicability in high-volume testing environments.Conclusion
The described HPLC method using a core-shell Accucore RP-MS column delivers rapid, efficient, and reliable separation of urea herbicides. Its performance makes it an excellent choice for environmental monitoring, food safety, and industrial quality control applications.Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Analysis of Triazine Pesticides Using a Core Enhanced Technology Accucore HPLC Column
2011|Thermo Fisher Scientific|Applications
Application Note: ANCCSCETPEST Analysis of Triazine Pesticides Using a Core Enhanced Technology Accucore HPLC Column Anila Khan, Thermo Fisher Scientific, Runcorn, Cheshire, UK Abstract Key Words • Accucore C18 • Superficially porous • Core Enhanced Technology • Fused core •…
Key words
triazines, triazinesscientific, scientificfisher, fishercore, corethermo, thermohplc, hplcnorth, northtriazine, triazineamerica, americadomestic, domesticfast, fastpesticides, pesticidesltd, ltdanccscetpest, anccscetpestenhanced
HPLC-UV Method for the Determination of Telmisartan Using a Core Enhanced Technology Accucore Column
2011|Thermo Fisher Scientific|Applications
Application Note: ANCCSCETTELMIS HPLC-UV Method for the Determination of Telmisartan Using a Core Enhanced Technology Accucore Column Eilidh MacRitchie, Thermo Fisher Scientific, Runcorn, Cheshire, UK Abstract Key Words • Telmisartan • Accucore RP-MS • Hypertension • Solid Core • Core…
Key words
telmisartan, telmisartanaccucore, accucorescientific, scientificfisher, fishercore, corethermo, thermonorth, northhplc, hplcamerica, americaasymmetry, asymmetrydomestic, domesticgrade, gradeltd, ltdanccscettelmis, anccscettelmisenhanced
Separation of Eighteen Pesticides Using Accucore RP-MS HPLC Column
2012|Thermo Fisher Scientific|Applications
Application Note: ANCCSCETHERBS Separation of Eighteen Pesticides Using Accucore RP-MS HPLC Column Derek Hillbeck, Thermo Fisher Scientific, Runcorn, Cheshire, UK Abstract Key Words • Pesticides This application shows the use of the Thermo Scientific Accucore RP-MS column for the separation…
Key words
accucore, accucorescientific, scientificisoproturon, isoproturonfisher, fisherpesticides, pesticidessebuthylazin, sebuthylazindiuron, diuronmetolachlor, metolachlordesethylatrazine, desethylatrazineminutes, minuteseighteen, eighteenthermo, thermonorth, northhplc, hplcamerica
Analysis of Steroids Using a Core Enhanced Technology Accucore HPLC Column
2011|Thermo Fisher Scientific|Applications
Application Note: ANCCSCETSTEROID Analysis of Steroids Using a Core Enhanced Technology Accucore HPLC Column Joanne Gartland, Thermo Fisher Scientific, Runcorn, Cheshire, UK Abstract Key Words • Progestogens • Accucore RP-MS • Fused core • Superficially porous • Core Enhanced Technology…
Key words
accucore, accucorescientific, scientificfisher, fisherprogestogens, progestogenscore, corethermo, thermonorth, northhplc, hplcamerica, americaprogesterone, progesteronedomestic, domesticsteroid, steroidsteroids, steroidsltd, ltdanccscetsteroid