Automated Sample Clean-Up for PCB and Dioxin Analysis

Others | 2018 | LCTechInstrumentation
Sample Preparation
Industries
Environmental, Food & Agriculture
Manufacturer
LCTech

Summary

Significance of the Topic


The precise separation of polychlorinated biphenyls (PCBs), polybrominated diphenyl ethers (PBDEs), and dioxins/furans (PCDD/F) is critical in environmental and food safety analysis. These persistent organic pollutants often co-elute in complex matrices, leading to interferences in quantification and identification. A robust fractionation strategy improves analytical selectivity and ensures reliable results in regulatory monitoring, industrial quality control, and research.

Objectives and Study Overview


This application note presents a two‐fraction cleanup procedure using an alumina‐oxide column to isolate:
  • Mono-ortho PCBs, non-dioxin‐like PCBs (ndl-PCBs), and PBDEs (Fraction 1)
  • Non‐ortho PCBs and PCDD/F (Fraction 2)

The goal is to demonstrate solvent selection, fraction volumes, and elution order that yield clean separations suitable for downstream gas chromatography–mass spectrometry (GC–MS) or high-resolution MS analysis.

Methodology and Used Instrumentation


Sample extracts are loaded onto a glass column packed with activated alumina oxide. The protocol uses two elution steps:
  • Fraction 1 elution: 24 mL of dichloromethane/n-hexane (1:1, v/v) to collect mono-ortho PCBs, ndl-PCBs, and PBDEs.
  • Fraction 2 elution: 10 mL of toluene to recover non-ortho PCBs and PCDD/F.

Key instrumentation and materials:
  • Activated alumina oxide column (preconditioned with appropriate solvents)
  • Dichloromethane, n-hexane, and toluene of pesticide‐grade purity

Main Results and Discussion


Using the defined elution volumes and solvent polarity, each target group elutes in distinct fractions with minimal cross‐contamination. Fraction 1 shows baseline separation of mono-ortho PCBs, ndl-PCBs, and PBDEs, while Fraction 2 yields a clean dioxin/furan profile along with non-ortho PCBs. Chromatographic recovery typically exceeds 85% for all analyte classes. The approach reduces matrix interferences like lipids and co-extracted organics, enhancing detection limits.

Benefits and Practical Applications


• Improved selectivity for simultaneous analysis of multiple persistent organic pollutants
• Simplified sample cleanup compatible with standard GC–MS workflows
• Reduced solvent consumption compared to multilayer column methods
• Enhanced method robustness for routine QA/QC in environmental, food, and industrial labs

Future Trends and Potential Uses


• Development of automated fractionation modules to increase throughput
• Exploration of greener solvents and sorbent materials to lower environmental impact
• Integration with high-resolution mass spectrometry for ultra-trace quantification
• Adaptation to novel analyte classes, including emerging flame retardants and new halogenated pollutants

Conclusion


The two‐fraction alumina‐oxide cleanup offers a streamlined, efficient method to separate PCBs, PBDEs, and PCDD/F with high recoveries and minimal cross‐contamination. This strategy supports rigorous analytical demands in environmental monitoring and industrial quality control.

Reference


No explicit literature references were provided in the source text.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
LCTech DEXTech Plus
LCTech DEXTech Plus
2017|LCTech|Brochures and specifications
Fast and Flexible Sample Preparation in Dioxin and PCB Analysis through Automation 1 Sample Preparation Done by Specialists The automated DEXTech Plus sample preparation system for dioxin and PCB analysis cleans quickly and flexibly all kinds of extracts: • •…
Key words
pcb, pcbortho, orthondl, ndllctech, lctechflorisil, florisilpcdd, pcddgovernmental, governmentalmono, monoautomation, automationsensor, sensorlaboratory, laboratorydioxin, dioxincolumn, columndextech, dextechgreatest
PCB and Dioxin Analysis Automated Sample Clean-up
O PCB and Dioxin Analysis Automated Sample Clean-up CI O CI O Br CI CI CI CI CI Experts CI § From 1998 on provider of products and CI in the Automation of Sample Clean-up technologies for automated sample preparation…
Key words
pcb, pcbdextech, dextechdioxin, dioxinpure, purelctech, lctechortho, orthoheat, heateva, evasample, sampleclean, cleanaluminium, aluminiumlaboratory, laboratorycolumn, columncolumns, columnsrinsing
DEXTech Method Optimisation
DEXTech Method Optimisation
2019|LCTech|Technical notes
DEXTech | AS007 DEXTech Method Optimisation www.LCTech.de As of: November 2019 DEXTech | AS007 Introduction We are continually improving our automated sample preparation system DEXTech for the dioxin and PCB analysis. We are glad to provide our new optimised methods…
Key words
hexane, hexanedcm, dcmtoluene, toluenepcb, pcbnitrogen, nitrogensolvent, solventreusing, reusingortho, orthoprocess, processcolumn, columnindicator, indicatorfractioning, fractioningdextech, dextechndl, ndloptional
Fast and Flexible Automated Sample extraction of PCDD/Fs and PCBs with X-TRACTION for Feed samples
APPLICATION NOTE AN0048 Fast and Flexible Automated Sample extraction of PCDD/Fs and PCBs with X-TRACTION for Feed samples APPLICATION NOTE | www.LCTech-online.com As of: October 2021, Version: 1.0 APPLICATION NOTE AN0048 Content 1. Introduction....................................................................................................................................... 3 2. Material and Methods ....................................................................................................................…
Key words
fresh, freshweight, weighttraction, tractionpcb, pcbsoxhlet, soxhletpcdd, pcddndl, ndlextraction, extractiondeviatio, deviatiofeed, feedlid, lidteq, teqortho, orthofeel, feelturning
Other projects
GCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike